Next: Cryogenic and Oxygen Deficiency
Up: Description of Hall C
Previous: Fire Hazards
Subsections
Hall C utilizes a large number of electro-magnets, both resistive and
superconducting. The HMS is a QQQD spectrometer.
All HMS magnets use superconducting coils with the capability to bend
particles with momenta up to 7.3 GeV/c.
The SOS is a normal-conducting QD
spectrometer with a maximum central
momentum of 1.7 GeV/c. In addition, Hall C contains a number of magnets
along the beam line, notably the Møller polarimeter and the Raster Magnets.
In Table 1, we give an overview of possible voltages and currents for the
main Hall C magnets.
Table 1:
Voltages and Currents for the Hall C Magnets
| Magnet |
Current [A] |
Voltage [V] |
| |
|
|
| SOS Quad |
1000 |
160 |
| SOS D1 |
1000 |
250 |
| SOS D2 |
1000 |
160 |
| HMS Q1,Q2,Q3 |
1250 |
10 |
| HMS Dipole |
3000 |
10 |
| Møller Small Quad |
300 |
30 |
| Møller Large Quad |
1500 |
150 |
| Møller Solenoid |
120 |
10 |
| Raster Magnets |
100 |
300-600 (AC) |
The magnets in the experimental areas are typically energized by remote
control. During major down times the magnets are powered down for personal safety
reasons as well as to reduce electrical power consumption.
During short interruptions of beam delivery,
with hall personnel entering the hall in the controlled access mode, the
magnets are typically left energized. The main reason is that the time
constants of large size magnets are long (of the order of hours), and
frequent ramping or cycling will lead to inefficient operation. Also, every
ramp of a large superconducting magnet involves some risk of permanent
damage to the magnet coil.
The principal hazards associated with the magnets are:
- Electrical
- All the spectrometer magnets, and some of the
beam line magnets, are high current devices
(
3000 Amps). The power supplies that provide this
current are potentially lethal. The most exposed and
hence most dangerous places are inside the supplies themselves and at the magnet
power leads.
- Magnetic
- The spectrometer magnets produce large fields
(
10 kG). The magnets all have return yokes
and thus the external fields (near the magnets) are not nearly as large
as the central fields but they may still be significant (up to about
1 kG).
The raster magnets do not have a return yoke and may have external fields
of a few hundred Gauss. Personnel working in the proximity of energized
magnets are exposed to the following magnetic hazards:
- danger of metal tools coming into contact with exposed leads, shorting
out the leads, depositing a large amount of power in the tool, vaporizing
the metal, and creating an arc.
- danger of metal objects being attracted by the magnet fringe field,
and becoming airborne, possibly pinching body parts.
- danger of cardiac pacemakers or other electronic medical devices
no longer functioning properly in the presence of magnetic fields.
- danger of metallic medical implants (non-electronic) being adversely affected
by magnetic fields.
- lose of information from magnetic data storage driver such as
tapes, disks, credit cards.
- LCW
- The magnet coils in the case of the SOS and the magnet
power supplies for both spectrometers as well as the current leads of the
HMS are all cooled by the Low Conductivity Water, LCW, system. This
is a high pressure system, P
240 PSI, and an unconfined stream
of water at this pressure could cause injury.
- Fire
- There also exists a potential fire problem associated
with the high current power supplies. A short in one of the coils,
or insufficient water cooling may lead to a fire in the power lines or the
coils.
Two different modes of operation need to be distinguished:
(1) routine operation involving work in the vicinity of the magnets,
but not in close proximity to the electrical connections, and not
involving any work that could result in purposely getting into contact with the
coils or the leads, and
(2) non-routine operation involving work on or near the exposed
current conductors or connections
(typically requiring removal of the shield) or any work that could result in
contact, intentional or otherwise, with the coils or the leads.
The following measures shall be taken by the cognizant hall engineer (or his designee)
to mitigate the hazards during routine operation:
- The current carrying conductors must be protected against accidental
contact or mechanical impact by appropriate measures (e.g. run cables in grounded
metal conduits or cable trays).
- All exposed current leads and terminations shall be covered by
non-conductive or grounded shields in such a manner as to
make it impossible for personnel to accidentally touch exposed leads with
either their body or with a tool (the electrical connections at the SOS
quadrupole, ``easily" accessible at the Hall C pivot point
have been covered with a plexiglass box).
Personnel shall be instructed not to
reach inside the shields. Warning signs shall to be placed on the shields;
the signs shall read:
DANGER
THIS GUARD MAY ONLY
BE REMOVED BY
AUTHORIZED PERSONNEL
UTILIZING JLAB
LOCKOUT - TAGOUT
PROCEDURES
- Whenever a magnet is energized, a flashing light on the magnet or on the
magnet support structure
must be activated to notify and warn personnel of the associated electrical
and magnetic field hazards.
- Administrative measures shall be implemented, as appropriate for the
situation, to reduce the danger of metal objects being attracted by the magnet
fringe field and becoming airborne.
(Note that for most magnets strong magnetic fields are only encountered within
non-accessible areas inside the magnet.) Areas where these measures are in
effect shall be clearly marked.
- To reduce the danger of magnetic fields to people using pacemakers
or other medical implants,
warning signs shall be prominently displayed at the entrance to each hall. The
sign shall read:
DANGER
SAFETY HAZARDS MAY EXIST FROM
THE MECHANICAL FORCES EXERTED
BY THE MAGNETIC FIELDS UPON
MEDICAL IMPLANTS
NO PACEMAKERS
are established according to hazard class and
mode of work.
The mode of work is determined by the nature of the work:
- de-energized
- energized with reduced safety and restricted manipulative operations
- energized with manipulative operations
The hazard class is determined by the type work (electrical or electronic)
and the combination of voltage and current.
- Anyone working on the HMS or SOS power supplies must comply with
the Standard Operating Procedure (S.O.P.). They must be trained and qualified
and obey the new arc flash and shock hazard procedure.
- All maintenance shall be performed in strict accordance
with the Jefferson Lab EH
S Manual, (6220 and 6230) and with
Lock Tag and Try (LTT).
The following references should be consulted before power supply
maintenance or operations are attempted: 1) the operating procedure
provided by the manufacturer (Inverpower for the SOS or Danfysik for the HMS
[4]) 2) the simplified magnet power supply maintenance procedure.
- Removal of any protective shield or cover for an electrical conductor
shall be performed using administrative lockout procedures. The lockout
shall be performed by the cognizant hall engineer (or his designee).
The administrative lock shall not be removed until the protective shield
or cover has been fully re-installed.
There will be no mode 3 (``hot work'') on any power supply in Hall C.
Mode 3 work is defined as manipulative operations that are conducted with equipment
fully energized and with some or all normal protective barriers removed.
There are crash buttons in the counting house which are interlocked
with a key (one for each spectrometer).
When the keys are locked off, in crash mode, the power supplies cannot be
energized. The keys to the crash button are controlled by W. Vulcan (SOS), and
S. Lassiter, M. Fowler, P. Brindza, and W. Vulcan for the HMS. The crash
buttons are certified by Hall C electrical personnel.
The practice of keeping electro-magnets energized in the experimental areas
during short accesses
provides substantial benefits to the quality and effectiveness of the physics
program. The resulting hazards have been mitigated by a combination of
protective shields, personnel training, warning lights and signs, and
administrative procedures.
The LCW water system for the SOS dipole and quadrupoles has been plumbed using
hoses rated for 600 PSI. These hoses have been tested during the magnet
mapping. The power supply cooling water hoses have a similar safety margin.
In case of problems with LCW contact one of the responsible personnel
listed in Section 4.
Next: Cryogenic and Oxygen Deficiency
Up: Description of Hall C
Previous: Fire Hazards
Hall-C Staff
2008-01-17