version 1.4, 2003/06/19 16:19:41
|
version 1.5, 2004/09/17 04:06:56
|
|
|
\title{How to Set the HMS Magnets} | \title{How to Set the HMS Magnets} |
\howtotype{user} % ``expert'', ``user'', ``reference'' | \howtotype{user} % ``expert'', ``user'', ``reference'' |
%\experiment{Name of experiment} % Optional | %\experiment{Name of experiment} % Optional |
\author{Dave Gaskell} |
\author{Mike Fowler and Dave Gaskell} |
\category{magnet} % Subject area of this document | \category{magnet} % Subject area of this document |
%\maintainer{Name of person maintaining document} % Optional | %\maintainer{Name of person maintaining document} % Optional |
\date{\today} % Can use \today as the argument | \date{\today} % Can use \today as the argument |
|
|
The HMS magnet settings are calculated using the program ``field03''. This | The HMS magnet settings are calculated using the program ``field03''. This |
program will provide the needed set--currents for Q1, Q2, and Q3, and the | program will provide the needed set--currents for Q1, Q2, and Q3, and the |
needed field for the dipole. To | needed field for the dipole. To |
access this program, log into cdaqs1 or cdaqs2 as cdaq. Go to the |
access this program, log into cdaqs1 or cdaqs3 as cdaq. Go to the |
\texttt{FIELD} directory (\texttt{cd FIELD}) and run \texttt{field03}. An | \texttt{FIELD} directory (\texttt{cd FIELD}) and run \texttt{field03}. An |
example follows:\\ | example follows:\\ |
\vspace*{0.2in}\\ | \vspace*{0.2in}\\ |
|
|
| |
\subsection{Set the dipole}\label{setdipole} | \subsection{Set the dipole}\label{setdipole} |
| |
The dipole can be set in three modes: current control, manual NMR field |
The dipole can be set in three modes: Current Control, Manual NMR Field |
control, and automatic NMR controlled ramp. The default method is at the |
Control, and Automatic NMR Controlled Ramp. The default method is at the |
discretion of the experiment -- consult your run coordinator. In this | discretion of the experiment -- consult your run coordinator. In this |
section, setting the dipole using manual NMR field control and automatic NMR |
section, setting the dipole using automatic NMR controlled ramp will be discussed. To switch |
controlled ramp will be discussed. |
to another operation mode, the dipole current must be zero. |
| |
\subsubsection{Manual NMR Field Control} |
|
| |
\begin{itemize} |
\subsubsection{Automatic NMR Controlled Ramp (Field Regulation Mode) - Starting from Zero} |
| |
|
\begin{itemize} |
\item{Select the dipole screen on the magnet controls console.} | \item{Select the dipole screen on the magnet controls console.} |
|
\item{Press F1 to go to the dipole main menu screen.} |
|
\item{Check the interlock status in the upper right corner of the main menu. If the interlock |
|
status field is flashing red:} |
| |
\item{Press F1 to go to the dipole main menu screen. Use the arrow keys to |
\begin{itemize} |
move the cursor to the \texttt{NMR manual op.} field - then type \texttt{exe}.} |
\item{Using the arrow keys, position the cursor over the \texttt{INTERLOCK} menu item. |
|
Type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{RESET INTERLOCK} menu item. Type |
|
\texttt{exe} and press ENTER.} |
|
\item{Press F1 and return to main menu.} |
|
\end{itemize} |
| |
\item{Move the cursor to the \texttt{Field target values} field - |
\item{Position the cursor over the \texttt{FIELD REGULATION} menu item - then type |
type \texttt{exe}.} |
{\texttt{exe}} and press ENTER.} |
| |
\item{Select the \texttt{High} field (type \texttt{exe}) and enter the first 4 |
\item{Position the cursor over the \texttt{DEFAULT mode} menu item. Type \texttt{exe} |
digits of the magnetic field and hit enter. For the example field from |
and press ENTER.} |
field03 shown in Section~\ref{field} (0.5497165 T), the first 4 digits are |
|
0549 for 0.549.} |
|
| |
\item{Select the \texttt{Low} field and enter the last 4 digits of the magnetic |
\item{Position the cursor over the \texttt{NMR start, PWR on} menu item, type \texttt{exe} and |
field.} |
press ENTER. After approximately one minute the NMR unit is activated. The NMR is on when the status |
|
field is white starts on a green background.} |
| |
\item{Select the \texttt{Write field target values} field - type \texttt{exe}.} |
\item{Position the cursor over the \texttt{Set DEFAULT values} menu item, type \texttt{exe} and press |
|
ENTER.} |
| |
\item{The magnet should start ramping - ramping is complete when the NMR |
\item{Position the cursor over the \texttt{high} value menu item, type \texttt{exe} and press ENTER.} |
status bit 5 is 1.} |
|
| |
\item{Make sure that the \texttt{Auto.measurement} field is on. This is |
\item{Type in the first 4 digits of the desired magnetic field and hit ENTER. For the example field from |
indicated by green highlighted asterisks \texttt{**}. } |
field03 shown in Section~\ref{field} (0.5497165 T), the first 4 digits are 0549 for 0.549.} |
If it is not on, you can activate it by typing \texttt{exe} in that field. |
|
| |
\item{Type \texttt{exe} in the \texttt{Field-ctrl ON} field.} |
\item{Position the cursor over the \texttt{low} value menu item, type \texttt{exe} and press ENTER.} |
| |
\item{Watch the field for a bit to make sure it settles. You may have turn |
\item{Type in the last 4 digits of the desired magnetic field and hit ENTER. Example: for 0.5497165 T, type |
\texttt{Auto.measurement} on again a few times.} |
7165.} |
| |
\end{itemize} |
\item{Position the cursor over the \texttt{CLOSE} menu item, type \texttt{exe} and press ENTER.} |
| |
NOTE: Sometimes, when the magnet is already set and regulating and you wish |
\item{Position the cursor over the \texttt{Start Excitation} menu item, type \texttt{exe} and press ENTER.} |
to change the field, the dipole will not start a new excitation when you |
|
\texttt{Write field target values}. If this happens, you can alleviate the |
|
problem by briefly turning the NMR off and then on again (the NMR is rather |
|
slow to respond when turned off, so be patient when turning it on again). |
|
| |
|
\item{The dipole will now follow the excitation curve and ramp to the desired field.} |
| |
|
\end{itemize} |
| |
\subsubsection{Automatic NMR Controlled Ramp} |
|
|
\subsubsection{Changing the HMS Dipole Field in Regulation Mode} |
| |
\begin{itemize} | \begin{itemize} |
\item{Select the dipole screen on the magnet controls console.} |
|
| |
\item{Press F1 to go to the dipole main menu screen. Use the arrow keys to |
\item{Position the cursor over the \texttt{Set DEFAULT values} menu item, type \texttt{exe} and |
move the cursor to the \texttt{FIELD REGULATION} field - then type |
press ENTER.} |
\texttt{exe}.} |
|
|
\item{Position the cursor over the \texttt{high} menu item, type \texttt{exe} and |
|
press ENTER.} |
|
|
|
\item{Type in the first 4 digits of the desired magnetic field and hit ENTER. For the example field from |
|
field03 shown in Section~\ref{field} (0.5497165 T), the first 4 digits are 0549 for 0.549.} |
|
|
|
\item{Position the cursor over the \texttt{low} value menu item, type \texttt{exe} and press ENTER.} |
| |
\item{Make sure the program is set for \texttt{DEFAULT mode}. Then, move the |
\item{Type in the last 4 digits of the desired magnetic field and hit ENTER. Example: for 0.5497165 T, type |
cursor to the \texttt{Set DEFAULTvalues} field - type \texttt{exe}.} |
7165.} |
| |
\item{Select the \texttt{High} field (type \texttt{exe}) and enter the first 4 |
\item{Position the cursor over the \texttt{CLOSE} menu item, type \texttt{exe} and press ENTER.} |
digits of the magnetic field and hit enter. For the example field from |
|
field03 shown in Section~\ref{field} (0.5497165 T), the first 4 digits are |
|
0549 for 0.549.} |
|
| |
\item{Select the \texttt{Low} field and enter the last 4 digits of the magnetic |
\item{Position the cursor over the \texttt{Start Excitation} menu item, type \texttt{exe} and press ENTER.} |
field.} |
|
| |
\item{Type \texttt{exe} in the \texttt{Start excitation} field - the magnet |
\item{The dipole will now follow the excitation curve and ramp to the desired field.} |
will start ramping.} |
\end{itemize} |
|
|
|
\subsubsection{HMS Dipole Ramp up-err/ Field crtl-err} |
|
|
|
Occasionally the NMR unit terminates the field ramp up/down or terminates the field control. |
|
When this happens, the \texttt{Ramp up-err} or \texttt{Field crtl-err} will change to red. |
|
You should always record any fault indications in the hclog, then perform the following: |
| |
|
\begin{itemize} |
|
\item{Position the cursor over the \texttt{RESET} menu item next to the error indicator, type |
|
\texttt{exe} and press ENTER. The NMR unit will continue the ramp up/down or restart |
|
field control.} |
\end{itemize} | \end{itemize} |
| |
\section{More information} |
\section{More information about the Quadrupoles} |
| |
\subsection{Changing the quadrupole polarity}\label{quadpol} | \subsection{Changing the quadrupole polarity}\label{quadpol} |
| |
|
|
\item{When the magnet reaches zero current, type \texttt{psu:pol:+} or | \item{When the magnet reaches zero current, type \texttt{psu:pol:+} or |
\texttt{psu:pol:-} for positive or negative polarity.} | \texttt{psu:pol:-} for positive or negative polarity.} |
| |
\item{At its new polarity, ramp the magnet to 950 Amps: |
\item{At its new polarity, ramp the magnet to its maximum current (Q1=950 Amps, Q2=900 Amps, Q3=850 Amps): |
\texttt{psu:setcur:950.0}.} |
\texttt{psu:setcur:xxx.x}.} |
| |
\item{You can now ramp the magnet down to the desired current: | \item{You can now ramp the magnet down to the desired current: |
\texttt{psu:setcur:xxx.xx}, \texttt{xxx.xx} is the absolute value of the | \texttt{psu:setcur:xxx.xx}, \texttt{xxx.xx} is the absolute value of the |
|
|
\end{itemize} | \end{itemize} |
| |
| |
\subsection{Changing the dipole polarity} |
\subsection{HMS Quadrupole Interlock Recovery} |
|
|
DRAFT - This section is still under development!!! |
|
|
|
\begin{itemize} |
|
|
|
\item{To change the dipole, first let the dipole ramp to zero current - this |
|
can be accomplished by:} |
|
|
|
\begin{itemize} |
|
\item{Typing \texttt{exe} in the \texttt{Stop Excitation} field if you |
|
are using the Automatic NMR ramping mode or} |
|
\item{Typing \texttt{exe} in the \texttt{NMR OFF} field if you are |
|
in NMR Manual operation mode.} |
|
\end{itemize} |
|
|
|
\item{You can monitor the dipole current by going to the \texttt{INTERLOCK} |
|
page.} |
|
|
|
\item{Once the dipole reaches zero current, type \texttt{exe} in the |
|
\texttt{Change polarity} box. NOTE: It may be necessary to turn the NMR back |
|
on before doing this.} |
|
|
|
\item{Once the polarity has changed, set the dipole using the instructions in |
|
Section \ref{setdipole}.} |
|
|
|
\end{itemize} |
|
|
|
\subsection{Resetting a HMS Quadrupole Power Supply} |
|
| |
Occasionally, an HMS quad power supply will trip off because a real or | Occasionally, an HMS quad power supply will trip off because a real or |
annoyance power supply fault is sensed. You should always record any fault |
annoyance power supply fault is sensed. Prudent checks of the system temperature, |
indications in the hclog. Once you have determined that the fault is not real |
pressure, and liquid level are wise. You should always record any fault |
|
indications in the hclog! Once you have determined that the fault is not real |
or that it has been corrected, and that there is no indication of a quench, | or that it has been corrected, and that there is no indication of a quench, |
the power supply may be reset by executing the following commands: | the power supply may be reset by executing the following commands: |
| |
\begin{itemize} | \begin{itemize} |
\item{Type \texttt{psu:setcur:0.0}.} |
\item{Type \texttt{psu:setcur:0.0} in the Command Line box.} |
\item{WAIT for the magnet to finish ramping to zero current/field (if it | \item{WAIT for the magnet to finish ramping to zero current/field (if it |
hasn't already).} | hasn't already).} |
\item{Type \texttt{reset}.} | \item{Type \texttt{reset}.} |
|
|
Section~\ref{quad_set}.} | Section~\ref{quad_set}.} |
\end{itemize} | \end{itemize} |
| |
\subsection{Quadrupole Quench/Dump Switch} |
\subsection{Quadrupole Quench/Dump Switch Open Interlock} |
| |
Recovery from a quadrupole quench can now be done remotely. There is little |
Recovery from a quadrupole quench can be done remotely. There is little |
or no possibility of a ``real" quench occurring in the main coils so one | or no possibility of a ``real" quench occurring in the main coils so one |
should always reset and continue. Prudent checks of the system temperature, | should always reset and continue. Prudent checks of the system temperature, |
pressure and liquid level are wise. There have been real quenches of the trim |
pressure and liquid level are wise. You should record any fault indications |
|
in the hclog! There have been real quenches of the trim |
coils that subsequently cause enough noise to trip the main coil off. There | coils that subsequently cause enough noise to trip the main coil off. There |
are other causes of fast discharges but these have all been the result of | are other causes of fast discharges but these have all been the result of |
electrical noise or a detection level set too low. There have been no real | electrical noise or a detection level set too low. There have been no real |
|
|
dump switch open interlock and main coil quench interlock are shown on the | dump switch open interlock and main coil quench interlock are shown on the |
ILK screen of the quadrupole display screen. | ILK screen of the quadrupole display screen. |
| |
To remotely reset the dump switch type the following commands in the proper |
To remotely reset the dump switch execute the following steps in the proper |
order: | order: |
| |
\begin{itemize} | \begin{itemize} |
\item{\texttt{qreset}} |
\item{Position the cursor over the {\texttt{ILK}} box and press Enter.} |
\item{\texttt{reset}} |
\item{In the Command Line box type \texttt{qreset}. The Main Coil Quench and/or |
\item{(wait seven minutes)} |
the Current Lead Overvoltage interlock will change to green.} |
\item{\texttt{reset}} |
%\item{\texttt{reset}} |
\item{\texttt{psu:reset}} |
\item{Wait 7 minutes for the Dump Switch Open interlock to change to green.} |
\item{\texttt{psu:on}} |
\item{Position the cursor over the {\texttt{PSU}} box and press Enter.} |
|
\item{Type \texttt{psu:reset}. The MPS OVERTEMP indicator will turn green.} |
|
\item{Type \texttt{psu:on} and set the magnet as normal.} |
\end{itemize} | \end{itemize} |
| |
To manually reset the dump switch at the control rack in the hall (if you have | To manually reset the dump switch at the control rack in the hall (if you have |
|
|
\item{Type 'psu:on'. } | \item{Type 'psu:on'. } |
\end{itemize} | \end{itemize} |
| |
\subsection{Dipole Quench Recovery} |
|
| |
The recovery from a dipole quench ``real or imaginary" can be completely |
\subsection{Rebooting the HMS quadrupole computers} |
remote. It is wise to check the coil voltage traces and system temperature, |
|
pressure and liquid level before restarting. A restart requires the following |
Occasionally, one or more of the magnet control computers will become hung, |
steps. |
stop responding, or cease to function. If this is the case, check the hall |
|
quadrupole hall probes and dipole NMR readback values on the TV monitors. If |
|
they are still at their nominal values, it is NOT necessary to reboot the |
|
computer right away. You may wait until some opportune moment, i.e. when |
|
changing kinematics or when beam may go away for an extended period. |
|
|
|
When you do decide to reboot the affected PC, first check the Reboot How--to |
|
\cite{reboot}. The steps to follow for the quadrupoles are are summarized below. |
| |
\begin{itemize} | \begin{itemize} |
\item{Clear interlocks.} |
\item{Rebooting the quad computer will crash the magnet so make sure |
\item{Select MPS manual operations.} |
you are really ready to reboot!} |
\item{Set desired current or field as before.} |
\item{Go to the screen of the affected computer (NumLock-minus-NumLock, etc.).} |
\item{Select ``write set data".} |
\item{Go to the reboot panel (on cdaqs1) and reboot that PC.} |
|
\item{The PC will come up with two ``sections''. On startup, the cursor will |
|
be in the bottom (blue?) section. Get to the top section by hitting the |
|
\texttt{Esc} key. NOTE: If the display remains blank after the reboot and you do not |
|
see the startup screen, go to one of the other magnet screens, then back to the screen |
|
of the affected computer.} |
|
\item{Now you can move the cursor to the "PSU" box - hit ENTER to get to the PSU screen.} |
|
\item{The screen will show an "MPS OVERTEMP" fault. This is OK.} |
|
\item{If the setcurrent readback value is not 0.0, type \texttt{psu:setcur:0.0} in the Command Line box. |
|
The MPS not ready will change from ``Not Ready'' to ``Ready'' when the current is zero.} |
|
\item{Type \texttt{psu:reset}. MPS OVERTEMP will change to green.} |
|
\item{Wait 7 minutes or until the MPS OVERTEMP changes to red.} |
|
\item{Type \texttt{psu:reset}. The MPS OVERTEMP should change to green. If not, repeat \texttt{psu:reset}.} |
|
\item{Type \texttt{psu:on} and set the magnet current as you would normally.} |
\end{itemize} | \end{itemize} |
| |
|
\section{More information about the Dipole} |
| |
The HMS dipole has never had a quench. Even the failure of a current lead in |
\subsection{Turning the dipole power supply on} |
October '93 did not quench the coil, according to the Hall C engineering |
|
staff. There have been numerous fast/slow discharges and all were due to |
|
electrical noise, low trip settings, deliberate operator initiated trips, or |
|
interlock dissatisfaction. |
|
| |
\subsection{Setting the dipole in current control mode} |
\begin{itemize} |
|
\item{In Hall C, unlock the and turn on the dipole main breaker.} |
|
\item{Turn on the dipole power supply switch located on top of the power supply.} |
|
\item{Open the center door and press the diode reset button. Close the door.} |
|
\item{Press RESET to clear the interlocks on the power supply. NOTE: One interlock will remain red.} |
|
\item{Turn the NMR on.} |
|
\end{itemize} |
| |
Perform the following steps to operate the dipole in current control mode. |
\subsection{Changing the dipole polarity} |
| |
\begin{itemize} | \begin{itemize} |
\item{Unlock the main breaker.} |
|
\item{Turn on the Dipole MPS breaker.} |
\item{To change the dipole, first let the dipole ramp to zero current - the current can be forced to |
\item{Switch to MPS local.} |
ramp down quickly by:} |
\item{Clear all interlock on MPS.} |
|
\item{Switch back to MPS remote.} |
\begin{itemize} |
\item{Select MPS manual operation from main menu.} |
\item{Position the cursor over the \texttt{Interlock} menu item. Type \texttt{exe} and press |
\item{Clear any interlocks.} |
ENTER.} |
\item{Select NMR stop and wait $\sim$ 1 min.} |
\item{Position the cursor over the \texttt{Fast Discharge by user} menu item. Type \texttt{exe} and |
\item{Select NMR start and wait $\sim$ 1 min.} |
press ENTER.} |
\item{If MPS interlock appears, try resetting.} |
|
\item{If not clear, then restart NMR and clear.} |
|
\item{Enter hi(gh) bit current 0 to 3000 for integer part |
|
of the current in Amps.} |
|
\item{Enter low bit current 0 to 999 for 0.000 to 0.999 amps.} |
|
\item{Change settling time if necessary.} |
|
\item{Select ``write data to MPS".} |
|
\end{itemize} | \end{itemize} |
|
\item{You can monitor the dipole current in the upper left corner of the \texttt{INTERLOCK} |
|
page. Wait until the current readback is zero.} |
| |
\subsection{Rebooting the magnet control computers} |
\item{Position the cursor over the \texttt{Reset Interlock} menu item. Type \texttt{exe} and press ENTER.} |
| |
Occassionally, one or more of the magnet control computers will become hung, |
\item{Press F1 (function key \# 1) to return to the Main Menu.} |
stop responding, or cease to function. If this is the case, check the hall |
|
quadrupole hall probes and dipole NMR readback values on the TV monitors. If |
|
they are still at their nominal values, it is NOT necessary to reboot the |
|
computer right away. You may wait until some opportune moment, i.e. when |
|
changing kinematics or when beam may go away for an extended peiod. |
|
| |
When you do decide to reboot the affected PC, first check the Reboot How--to |
\item{Position the cursor over the \texttt{Field Regulation, NMR Manual op.}, or the \texttt{MPS Manuel op.} menu item, |
\cite{reboot}. The steps to follow are summarized below. |
depending on which mode is being used. Type \texttt{exe} and press ENTER.} |
|
|
|
\item{Position the cursor over the \texttt{NMR start, PWR on} menu item. Type \texttt{exe} and press ENTER. |
|
After approximately one minute the NMR unit is activated. The NMR is on when the status |
|
field is white starts on a green background.} |
|
|
|
\item{Position the cursor over the \texttt{Change +/-} menu item. Type \texttt{exe} and press ENTER.} |
|
|
|
\item{Once the polarity has changed, set the dipole using the instructions in |
|
Section \ref{setdipole}.} |
|
|
|
\end{itemize} |
| |
\subsubsection{Rebooting HMS quad computer} |
|
| |
DRAFT! STILL IN PROGRESS!!! |
\subsection{Dipole Quench Recovery / Interlock Recovery } |
|
|
|
The recovery from a dipole quench ``real or imaginary" can be completely remote. It is wise to check the |
|
coil voltages, system temperature, pressure, and liquid levels before restarting. You should always record |
|
any fault indications in the hclog! A restart requires the following steps. |
|
|
|
\begin{itemize} |
|
|
|
\item{Position the cursor over the \texttt{Interlock} menu item. Type \texttt{exe} and press ENTER.} |
|
\item{Monitor the current value in the upper left corner. Wait until the current read back is zero.} |
|
\begin{itemize} |
|
\item{The current can be forced to ramp down quickly. Position the cursor over the \texttt{Fast Discharge by |
|
user} menu item. Type \texttt{exe} and press ENTER.} |
|
\end{itemize} |
|
\item{Position the cursor over the \texttt{Reset Interlock menu item}. Type texttt{exe} and press ENTER.} |
|
\item{Go to the reboot panel (on cdaqs1) and reset the Dipole Power Supply Diode} |
|
\item{Press F1 (function key \# 1) to return to the Main Menu.} |
|
\item{Position the cursor over the \texttt{Field Regulation, NMR Manual op.}, or the \texttt{MPS Manual op.} page, |
|
depending on the mode of operation being used.} |
|
\item{Position the cursor over the \texttt{NMR start, PWR on} menu item, type \texttt{exe} and press ENTER. |
|
After approximately one minute the NMR unit is activated. The NMR is on when the status field is white |
|
stars on a green background.} |
|
\item{Set desired current or field.} |
|
\end{itemize} |
|
|
|
The HMS dipole has never had a real quench. Even the failure of a current lead in October '93 did not |
|
quench the coil, according to the Hall C engineering staff. There have been numerous fast/slow discharges |
|
and all were due to electrical noise, low trip settings, deliberate operator initiated trips, or interlock |
|
dissatisfaction. |
|
|
|
\subsection{HMS Dipole NMR-err / MPS-err} |
|
A NMR-err or MPS-err indicates that a communication error has occurred between the control system and |
|
the NMR unit and/or the Power Supply. You should always record any fault indications in the hclog! |
|
|
|
\begin{itemize} |
|
\item{Set the Magnet alarms (near the target computer) to Bypass.} |
|
\item{Position the cursor over the \texttt{Interlock} menu item. Type \texttt{exe} and press ENTER.} |
|
\item{Monitor the current value in the upper left corner. Wait until the current read back is zero.} |
|
\begin{itemize} |
|
\item{The current can be forced to ramp down quickly. Position the cursor over the \texttt{Fast Discharge by |
|
user} menu item. Type \texttt{exe} and press ENTER.} |
|
\end{itemize} |
|
\item{Press F1 (function key \# 1) to go to the Main Menu.} |
|
\item{Position the cursor over the \texttt{Field Regulation, NMR Manual op.}, or the |
|
\texttt{MPS Manual op.} page, depending on the mode of operation being used. Type \texttt{exe} |
|
and press ENTER.} |
|
\item{Position the cursor over the \texttt{RS232 restart} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Press F7 (function key \# 7) to go to the \texttt{Device} page.} |
|
\item{Wait for all the devices to restart. } |
|
\item{Position the cursor over the \texttt{Reset Interlock} menu item. Type \texttt{exe} and press ENTER.} |
|
\item{Go to the reboot panel (on cdaqs1) and reset the Dipole Power Supply Diode.} |
|
\item{Press F1 (function key \# 1) to return to the Main Menu.} |
|
\item{Position the cursor over the \texttt{Field Regulation, NMR Manual op.}, or the \texttt{MPS Manual op.} |
|
page, depending on the mode of operation being used. Type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{NMR start, PWR on} menu item, type \texttt{exe} and press ENTER. |
|
After approximately one minute the NMR unit is activated. The NMR is on when the status field |
|
is white stars on a green background.} |
|
\item{The field or current may now be set.} |
|
\item{Set the Magnet alarms (near the target computer) to Active.} |
|
\end{itemize} |
|
|
|
\subsection{Restricting the NMR } |
|
The NMR unit is very sensitive to electromagnetic interference, so when the unit is searching for the |
|
magnetic field the unit will try to lock in on a NMR signal without magnetic field. For most cases the NMR |
|
will simply attempt to lock in on the interference for a short period of time, then give up and continue on. |
|
However, when ramping the Dipole to high field levels, (0.50 T and above), the NMR unit will try to lock in |
|
on several interference signals and may never reach the magnet field set point after several Ramp up-err |
|
resets. When this happens it is possible to restrict the number of probes that the NMR will search and |
|
improve the probability of obtaining a good field lock. Perform the following steps to restrict the number of |
|
probes. |
|
|
|
NOTE: The procedure only works when the Dipole is at zero amps. |
|
\begin{itemize} |
|
\item{Press F1 (function key \# 1) to go to the Main Menu.} |
|
\item{Position the cursor over the \texttt{NMR manual op.} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{Multiplexer} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{Number of multiplexer channel} menu item, type |
|
\texttt{exe} and press ENTER.} |
|
\item{Type in \texttt{1}, and press ENTER.} |
|
\item{Position the cursor over the \texttt{Start multiplexer channel} menu item, type \texttt{exe} |
|
and press ENTER.} |
|
\item{Type in the probe number from the chart below, and press ENTER.} |
|
NMR Probes: |
|
\begin{tabular}{cc} |
|
Probe \# & Range (Tesla) \\ |
|
2 & 0.09T - 0.26T \\ |
|
3 & 0.17T - 0.52T \\ |
|
4 & 0.35T - 1.05T \\ |
|
5 & 0.70T - 2.10T |
|
\end{tabular} |
|
|
|
\item{Position the cursor over the \texttt{Write field multiplexer parameter} menu item, type \texttt{exe} and press |
|
ENTER.} |
|
\item{Position the cursor over the \texttt{NMR stop, PWR off} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{NMR start, PWR on} menu item, type \texttt{exe} and press ENTER.} |
|
|
|
\end{itemize} |
|
|
|
\subsection{Setting the HMS Dipole in current control mode } |
|
|
|
Perform the following steps to operate the dipole in current control mode. |
|
|
|
\subsubsection{MPS Manual operation: Starting from zero.} |
| |
\begin{itemize} | \begin{itemize} |
\item{Rebooting the quad computer will crash the magnet so make sure |
\item{Press F1 (function key \# 1) to go to the Main Menu.} |
you are really ready to reboot!} |
\item{Check the Interlock status, upper right corner of the main menu. |
\item{Go to the screen of the affected computer (NumLock-minus-NumLock, etc.).} |
If the interlock status field is flashing red:} |
\item{Go to the reboot panel (on jlabs1) and reboot that PC.} |
|
\item{The PC will come up with two ``sections''. On startup, the cursor will |
\begin{itemize} |
be in the bottom (blue?) section. Get to the top section by hitting the |
\item{Using the arrow keys, position the cursor over the \texttt{Interlock} menu item. |
\texttt{Esc} key.} |
Type \texttt{exe} and press ENTER.} |
\item{Now you can move the cursor to the "PSU" box - type \texttt{exe} to get |
\item{Position the cursor over the \texttt{Reset Interlock} menu item. Type \texttt{exe} and press ENTER.} |
to the PSU screen.} |
\item{Press F1 to return to the Main Menu.} |
\item{The screen will likely show an "MPS OVERTEMP" fault. This is OK.} |
\end{itemize} |
\item{If the setcurrent readback value is not 0.0, type |
|
\texttt{psu:setcur:0.0}.} |
\item{Position the cursor over the MPS Manual op. menu item. Type EXE and press Enter.} |
\item{Wait 5 minutes.} |
\item{Position the cursor over the \texttt{NMR start, PWR on} menu item, type \texttt{exe} and press ENTER. |
\item{Type \texttt{reset} and then \texttt{psu:reset}. The ``MPS OVERTEMP'' |
After approximately one minute the NMR unit is activated. |
fault should go away. If it does not, repeat \texttt{psu:reset}.} |
The NMR is on when the status field is white stars on a green background.. } |
\item{Type \texttt{psu:on} and set the magnet current as you would normally.} |
\item{Position the cursor over the \texttt{Set current high} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Type high current value 0 to 3000 for integer part of the current in Amps, and press ENTER.} |
|
\item{Position the cursor over the \texttt{Set current low} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Enter low bit current value 0 to 999 for 0.000 to 0.999 amps.} |
|
\item{Position the cursor over the \texttt{Automatic measurements} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{Write set data} menu item, type \texttt{exe} and press ENTER.} |
|
\item{The Dipole will now ramp to the desired current.} |
|
\end{itemize} |
|
|
|
\subsubsection{MPS Manual operation: Changing the current value} |
|
\begin{itemize} |
|
\item{Position the cursor over the \texttt{Set current high} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Type high current value 0 to 3000 for integer part of the current in Amps, and press ENTER.} |
|
\item{Position the cursor over the \texttt{Set current low} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Enter low bit current value 0 to 999 for 0.000 to 0.999 amps. } |
|
\item{Position the cursor over the \texttt{Write set data} menu item, type \texttt{exe} and press ENTER.} |
|
\item{The dipole will now ramp to the new current.} |
|
\end{itemize} |
|
|
|
\subsection{Setting the HMS Dipole in NMR Manual operation } |
|
|
|
|
|
\subsubsection{NMR Manual operation: Starting from zero.} |
|
\begin{itemize} |
|
\item{Press F1 (function key \#1) to go to the Main Menu.} |
|
\item{Check the Interlock status, upper right corner of the main menu. |
|
If the interlock status field is flashing red:} |
|
\begin{itemize} |
|
\item{Using the arrow keys, position the cursor over the \texttt{Interlock} menu item. Type \texttt{exe} and press |
|
ENTER.} |
|
\item{Position the cursor over the Reset Interlock menu item. Type \texttt{exe} and press ENTER.} |
|
\item{Press F1 to return to the Main Menu.} |
|
\end{itemize} |
|
\item{Position the cursor over the \texttt{NMR manual op.} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{NMR start, PWR on} menu item, type \texttt{exe} and press ENTER. |
|
After approximately one minute the NMR unit is activated. |
|
The NMR is on when the status field is white stars on a green background.} |
|
\item{Position the cursor over the \texttt{Auto Measurement} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{Field Target Value} menu item, type \texttt{exe} and press ENTER. |
|
The field target value is a 32 bit word split into two 16 bit words, high and low.} |
|
\item{Position the cursor over the \texttt{high value} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Type in the field target value high word, and press ENTER. |
|
(Example: For 1.2345678 T type 1234 )} |
|
\item{Position the cursor over the \texttt{low value} menu item, type \texttt{exe} and press ENTER. |
|
(Example: For 1.2345678 T type 5678 )} |
|
\item{Position the cursor over the Write field target values menu item, type \texttt{exe} and press ENTER.} |
|
\item{Wait (The dipole power supply will ramp the magnet. |
|
After the power supply has ramped the magnet, the Searchmode will automatically switch to on.) |
|
The NMR unit will search for the field, lock on, and calculate the regulation window. When this |
|
calculation is complete, and if successful, the NMR-status 5 will change to 1, and the auto |
|
measurements will switch to manual measurement.) } |
|
\begin{itemize} |
|
\item{If the NMR-status 5 displays a value other than 1 visit the web site \\ |
|
\texttt{http://www.jlab.org/$\sim$fowler/nmrstatusie.htm} \\ and enter the NMR register values to obtain |
|
more details. You should always record any fault indications in the hclog!} |
|
\end{itemize} |
|
\item{Position the cursor over the \texttt{Auto Measurement} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{Field-ctrl ON} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Wait ( The Searchmode will automatically switch on, and the NMR will reacquire field lock. |
|
The NMR will then make adjustments to the power supply, and if successful, the NMR-status 5 |
|
will change to 1, and the auto measurements will switch to manual measurement.)} |
|
\begin{itemize} |
|
\item{If the NMR-status 5 is not set to 1 then an error has occurred.} |
|
\item{Position the cursor over the Field-ctrl OFF menu item, type \texttt{exe} and press ENTER.} |
|
\item{Return to the previous step (turn \texttt{Field-ctrl ON}).} |
|
\end{itemize} |
|
\item{Position the cursor over the \texttt{Auto Measurement} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Wait (It takes approximately three minutes for the NMR to fine tune the field to the set value. |
|
When the NMR finished, the field the NMR-status 5 will change to 8, and the auto measurements |
|
will switch to manual measurement.)} |
|
\item{Position the cursor over the \texttt{Auto Measurement} menu item, type \texttt{exe} and press ENTER.} |
|
\item{The Dipole is now ready.} |
|
\end{itemize} |
|
|
|
\subsubsection{NMR Manual operation: Changing the field target value} |
|
\begin{itemize} |
|
\item{Position the cursor over the \texttt{Field-ctrl OFF} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{Field Target Value} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{high value} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Type in the field target value high word, and press ENTER.} |
|
\item{Position the cursor over the \texttt{low value} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Type in the field target value low word, and press ENTER. } |
|
\item{Position the cursor over the \texttt{Write field target values} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{Auto Measurement} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Wait (The dipole power supply will ramp the magnet. |
|
After the power supply has ramped the magnet, the Searchmode will automatically switch to on.) |
|
The NMR unit will search for the field, lock on, and calculate the regulation window. When this |
|
calculation is complete, and if successful, the NMR-status 5 will change to 1, and the auto |
|
measurements will switch to manual measurement.)} |
|
\begin{itemize} |
|
\item{If the NMR-status 5 is not set to 1 visit the web site \\ |
|
\texttt{http://www.jlab.org/$\sim$fowler/nmrstatusie.htm}\\ and enter the NMR register values to obtain |
|
additional details. You should always record any fault indications in the hclog!} |
|
\end{itemize} |
|
\item{Position the cursor over the \texttt{Auto Measurement} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Position the cursor over the \texttt{Field-ctrl ON} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Wait ( The Searchmode will automatically switch on, and the NMR will reacquire field lock. |
|
The NMR will then make adjustments to the power supply, and if successful, the NMR-status 5 |
|
will change to 1, and the auto measurements will switch to manual measurement.)} |
|
\begin{itemize} |
|
\item{If the NMR-status 5 is not set to 1 then an error has occurred.} |
|
\item{Position the cursor over the \texttt{Field-ctrl OFF} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Return to the previous step (turn \texttt{Field-ctrl ON}).} |
|
\end{itemize} |
|
\item{Position the cursor over the \texttt{Auto Measurement} menu item, type \texttt{exe} and press ENTER.} |
|
\item{Wait (It takes approximately three minutes for the NMR to fine tune the field to the set value. |
|
When the NMR is finished, the NMR-status 5 will change to 8, and the auto measurements will |
|
switch to manual measurement.)} |
|
\item{Position the cursor over the \texttt{Auto Measurement} menu item, type \texttt{exe} and press ENTER.} |
|
\item{The Dipole is now ready.} |
\end{itemize} | \end{itemize} |
| |
\subsubsection{Rebooting HMS dipole} |
\subsection{Rebooting the dipole control computers} |
|
|
|
Occasionally, one or more of the magnet control computers will become hung, |
|
stop responding, or cease to function. If this is the case, check the hall |
|
quadrupole hall probes and dipole NMR readback values on the TV monitors. If |
|
they are still at their nominal values, it is NOT necessary to reboot the |
|
computer right away. You may wait until some opportune moment, i.e. when |
|
changing kinematics or when beam may go away for an extended period. |
| |
DRAFT! STILL IN PROGRESS!!! |
When you do decide to reboot the affected PC, first check the Reboot How--to |
|
\cite{reboot}. The steps to follow are summarized below. |
| |
\begin{itemize} | \begin{itemize} |
| |
|
|
| |
\item{Go to the dipole screen (NumLock-minus-NumLock, \texttt{D}, | \item{Go to the dipole screen (NumLock-minus-NumLock, \texttt{D}, |
\texttt{return}.).} | \texttt{return}.).} |
\item{Go to the reboot panel (on jlabs1) and reboot the dipole PC.} |
\item{Go to the reboot panel (on cdaqs1) and reboot the dipole PC.} |
|
\item{If the display remains blank after the reboot, go to a quadrupole screen, then |
|
back to the dipole screen.} |
\item{Once the computer has recovered, type \texttt{hms} at the prompt. This | \item{Once the computer has recovered, type \texttt{hms} at the prompt. This |
should start the magnet control program.} |
will start the magnet control program.} |
| |
\end{itemize} | \end{itemize} |
| |
|
|
| |
% Revision history: | % Revision history: |
% $Log$ | % $Log$ |
|
% Revision 1.5 2004/09/17 04:06:56 saw |
|
% Include text from Mike Fowler |
|
% |
% Revision 1.4 2003/06/19 16:19:41 saw | % Revision 1.4 2003/06/19 16:19:41 saw |
% Add wait of 5 minutes after reboot. | % Add wait of 5 minutes after reboot. |
% | % |