(file) Return to drift_gas_system.tex CVS log (file) (dir) Up to [HallC] / Documents / Howtos

  1 saw   1.1 \documentclass{chowto}
  2           
  3           \title{Drift Chamber Gas System Operation}
  4           \howtotype{expert} % ``expert'', ``user'', ``reference''
  5           %\experiment{Name of experiment} % Optional
  6           \author{H. Fenker}
  7           \category{general} % Subject area of this document
  8           
  9 saw   1.5 \maintainer{H. Fenker} % Optional
 10           \date{December 18, 2007} % Can use \today as the argument
 11 saw   1.1 
 12 saw   1.5 %To attempt to make pretty-looking pdf files...(not standard howto)
 13 saw   1.1 
 14           \begin{document}
 15           \providecommand{\degg}{\ensuremath{^{\circ}\ }}
 16           
 17           \begin{abstract}
 18 saw   1.5 This document provides detailed setup information for the drift
 19           chamber gas mixing system, as well as the correct procedure for
 20           refilling the alcohol supply and changing gas bottles. This
 21           information is intended for use by {\bf gas system experts only.}  For
 22           day-to-day shift worker instructions, refer to the corresponding {\it
 23           user} howto document.
 24           
 25           {\bf Please Note:} The Hall-C Drift Chamber Gas System was
 26           significantly changed in early 2007. It is now a {\it pressure-driven
 27           system} whereas it had been flow-controlled. Long-time Hall-C staff
 28           and users will find that the system operates quite differently now.
 29 saw   1.1 \end{abstract}
 30           
 31 saw   1.5 %=====================================================================
 32           %=====================================================================
 33 saw   1.1 \section{Overview}
 34           
 35 saw   1.5 The drift chamber gas is composed of 50\% Argon and 50\% Ethane (by
 36           volume), bubbled through isopropanol maintained at a temperature such
 37           that the gas mixture contains approximately 1\% alcohol vapor. The
 38           mixing system that produces this gas is housed in the Hall-C gas shed
 39           (Bldg. 96c). The bottles supplying the gas to the mixing system are
 40           attached to two two-bottle manifolds outside the gas shed, within the
 41           fenced-in gas bottle yard.
 42           
 43           A single mixing system provides the gas for both the SOS and the HMS
 44           detectors. Gas is delivered to the hall at a pressure of about 9 psi
 45           (500~Torr) above atmospheric pressure.  Each detector's flow is
 46           controlled by its own individual needle-valve with flowmeter, located
 47           in the appropriate shield house, near the detector. The job of the
 48           mixing system is to simultaneously maintain the delivery pressure and
 49           mixing ratio by providing whatever total flow rate of gas (between
 50           zero and 5.28 standard liters per minute [SLM]) is demanded by the
 51           detectors.  This is accomplished by controlling two flow valves (argon
 52           and ethane) so that they each flow the same volume of gas while
 53           keeping the output pressure nearly constant. The mixing system flow
 54           diagram is shown in Fig. \ref{fig:gas_mixer_diagram}.
 55           %=====================================================================
 56 saw   1.5 %=====================================================================
 57 saw   1.2 \section{Gas Interlock System}
 58 saw   1.5 The flow of gas from the supply bottles may be automatically shut off
 59           by normally-closed solenoid valves installed in the primary argon and
 60           ethane manifolds. Several conditions such as over-temperature, fan
 61           failure, gas leak, and fire alarm must all be in the non-alarm state
 62           before these valves will open. Alarm conditions are indicated on the
 63           gas system alarm panel on the lower-left side of the center
 64           counting-house console.
 65           
 66           When any of the required conditions is not satisfied the sounder on
 67           the panel in the counting room will make an annoying noise and both
 68           solenoid valves will close. The audible alarm may be silenced by a
 69           toggle switch on the panel. Be certain to return it to the ``on''
 70           position as soon as the fault is cleared.
 71           
 72           The most confusing, but most common alarm condition is ``Low
 73           Pressure''.  The solenoid valves will not remain open unless there is
 74           already ample pressure on the output side of both valves. This
 75           prevents, for example, the flow of pure ethane to the drift chambers
 76           when the argon bottle is empty. The way to clear this condition is to
 77           make sure there are no other faults and that both argon and ethane
 78           manifolds are properly pressurized and fitted with non-empty bottles;
 79 saw   1.5 then press and hold the ``override'' button on the alarm panel for
 80           several minutes. This button forces the solenoid valves to open even
 81           if there are fault conditions present. If all is well, gas will flow
 82           through the valves and clear the ``low pressure'' condition so that
 83           the button may be released.
 84 saw   1.2 
 85 jones 1.3 \begin{figure}[hbt]
 86 saw   1.5 \psfig{figure=drift_gas_system-gasmixer2007.eps,width=\textwidth,bbllx=12,bblly=12,bburx=750,bbury=590}
 87 saw   1.1 \caption{Diagram of Hall~C Gas Mixing System\label{fig:gas_mixer_diagram}}
 88 jones 1.3 \vspace{0.5cm}
 89 saw   1.1 \end{figure}
 90 saw   1.5 %=====================================================================
 91           %=====================================================================
 92 jones 1.3 \section{Operating the Mass Flow Controller.}
 93 saw   1.1 
 94 jones 1.3 The gas flow is controlled by a MKS~647 controller and mass flow
 95 saw   1.5 control valves.  Delivery pressure is sensed at the mixer output by a
 96           {\it Baratron} pressure transducer.  The 647 is menu driven from a
 97           display on the front panel using a keypad with numeric and cursor
 98           controls for input. It features a non-volatile memory so that most of
 99           its settings are retained even if the unit is unpowered.  The
100           pressure-regulation parameters do need to be reset if power is lost,
101           however, so the controller is powered through an uninterruptible power
102           supply (UPS).
103           
104           The temperature of the control
105           system must be maintained within the
106           operating range of the MKS~647 (15\degg C - 40\degg C), the pressure
107           transducer (0\degg C - 50\degg C), and the mass-flow meters and valves
108           (0\degg C - 50\degg C).  Therefore,
109           the heating and cooling systems in the gas shed must be maintained in
110           working order. The temperature is indicated by the
111           red LED display labeled \emph{Controller Temperature},
112           located in the right-hand rack of the gas shed.
113 jones 1.3 
114 saw   1.5 %=====================================================================
115           %=====================================================================
116 jones 1.3 \subsection{Settings for Normal Operation}
117           
118 saw   1.5 A summary of all of the settings required to make the controller
119           function properly is given in Table~\ref{tab:mixer_nominals}. The
120           table also shows which screen contains each parameter. Instructions
121           for setting parameters are given below. Detailed instructions for
122           configuring and operating the MKS~647 can be found in the
123           manufacturer's instruction manual\cite{647C_EN_0504A1}.
124           
125 jones 1.3 \begin{table}[hbt]
126           \begin{minipage}[h!]{\textwidth}
127           {\scriptsize
128 saw   1.5 \begin{center}
129 jones 1.3 \begin{tabular}{|l|c|l|}
130           \hline
131           Parameter   & Set To   & {\it Controller Screen}/Comments\\ \hline
132           \multicolumn{2}{|c|}{\bf Manual Valves  } &  Valves are labeled        \\
133           Valves 1, 2, 4, 8 & OPEN                  &         \\
134           Valves 3, 5, 6, 7 & CLOSED                &         \\ \hline
135 saw   1.5 \multicolumn{2}{|c|}{\bf Pressure PID Loop Settings   } & {\it Pressure Control} (Fig. \ref{fig:pressure_control})  \\
136 jones 1.3 Pressure    & 500 Torr &                   \\
137           PID Mode    & AUTO     &                   \\
138           PID GAIN    & 4.0      &                   \\
139           PID INTEG   & 10.0     &                   \\
140           PID LEAD    & 0.000    &                   \\ \hline
141 saw   1.5 \bf Mixture     & 1        & {\it User} (Fig. \ref{fig:user_display})/ Lower-right corner\\ \hline
142           \multicolumn{2}{|c|}{\bf Gas Composition for Mixture 1} & {\it Gas Composition} (Fig. \ref{fig:gas_composition})\\
143           Channel 1   & 1.000    &  (Argon)                 \\
144           Channel 2   & 1.000    &  (Ethane)                 \\
145 jones 1.3 Channel 3   & 0.000    &                   \\
146           Channel 4   & 0.000    &                   \\ \hline
147 saw   1.5 \multicolumn{2}{|c|}{\bf MFC Valve Size } & {\it Range Selection} (Fig. \ref{fig:range_selection})  \\
148           \multicolumn{2}{|c|}{\bf  / Gas}             & {\it Gas Selection} (Fig. \ref{fig:gas_selection})  \\
149           Channel 1   & 2.0 SLM / Ar         & provides 2.78 SLM Argon\\
150           Channel 2   & 5.0 SLM / C$_2$H$_6$ & provides 2.50 SLM Ethane\\
151 jones 1.3 Channel 3   & \it unused&                  \\
152           Channel 4   & \it unused&                  \\ \hline
153 saw   1.5 \multicolumn{2}{|c|}{\bf Channels ON/OFF Settings} & {\it Extended Display} (Fig. \ref{fig:extended_display})   \\
154 jones 1.3 Channel 1   & ON       &                   Press ``ON 1'' (\em Argon)  \\
155           Channel 2   & ON       &                   Press ``ON 2'' (\em Ethane) \\
156           Channel 3   & OFF      &                   Press ``OFF 3'' (\em not in use) \\
157           Channel 4   & OFF      &                   Press ``OFF 4'' (\em not in use) \\ \hline
158 saw   1.5 \multicolumn{2}{|c|}{\bf Pressure Transducer} & {\it Pressure Setup}  \\
159           Controller  & STD      &                   \\ 
160           Range F.S.  & 1000 Torr&                   \\ \hline
161           \multicolumn{2}{|c|}{\bf MFC Valve Controls     } & {\it Mode Selection} (Fig. \ref{fig:mode_selection}) \\
162 jones 1.3 Channel 1   & PID      &                                   \\
163           Channel 2   & SLAVE / 1&                                   \\
164           Channel 3   & INDEP    &                                   \\
165           Channel 4   & INDEP    &                                   \\ \hline
166           \bf Alcohol Temp. & $2^\circ $ C & Electronic Temperature  \\
167                             &              & Control Box             \\ \hline
168           \hline
169           \end{tabular}
170 saw   1.5 \end{center}
171           }%end of \scriptsize
172 jones 1.3 \end{minipage}
173           \caption{Normal Valve and Parameter Settings for the Gas Mixing System.
174           \label{tab:mixer_nominals}}
175           \end{table}
176 saw   1.5 %=====================================================================
177           \subsection{General Operation of the Mass Flow Controller}
178           If the controller screen is dark, press {\bf ESC} to awaken the
179           display.  Many screens merely provide a menu of other screens you may
180           access: simply press the item number you desire. To go up one level in
181           the menu hierarchy, press {\bf ESC}. The \emph{Menu Tree} for the 647C
182           controller is shown in Fig. \ref{fig:command_tree}.
183           
184           In general, to change a parameter displayed on the controller screen
185           use the {\bf left/right} arrow keys to move the cursor to the item you
186           wish to change. Then either use the number keys to enter the value
187           desired for that item (numeric parameter) or use the {\bf ENTER} or
188           {\bf up/down} keys to cycle a parameter through its available settings
189           (configuration parameter). Numeric parameters may be incrementally
190           modified by using the {\bf up/down} arrow keys. To make certain that a
191           new parameter becomes active, move the cursor off of the parameter
192           after you have entered the new value.
193           
194           The initial menu upon startup is the {\bf Main Menu}
195           (Fig.~\ref{fig:main_menu}).  For normal operation use the {\bf User
196           Display} menu (Fig.~\ref{fig:user_display}). It shows the amount of
197 saw   1.5 each gas currently flowing, the total gas flow, and the current
198           delivery pressure. This display also shows which of several possible
199           pre-defined gas mixtures is selected. These mixtures are configured 
200           on the {\bf Gas Composition} screen, Fig. \ref{fig:gas_composition}.
201           For normal operation, we use
202           only mixture {\bf \#1}, (number shown on the lower-right of the
203           display). {\bf Only on this screen can this parameter be changed.}
204           Mixture {\bf \#2} is usually configured to provide 100\% argon for
205           purging flammable gas out of the chambers.
206           
207           The {\bf Extended Display} menu (Fig.~\ref{fig:extended_display})
208           shows actual flow, flow set point, units, valve full-scale range, gas
209           calibration factor, whether that channel is enabled, and whether each
210           channel is operating in master, slave, PID, or independent mode. This
211           display is most useful to a system expert wishing to verify the system
212           parameter settings.  Most parameters cannot be modified from this
213           screen, however.
214           
215           Delivery pressure set-point and pressure \emph{PID-loop} control parameters
216           may be configured from the {\bf Pressure Control} screen
217           (Fig.~\ref{fig:pressure_control}).
218 saw   1.5 %
219           \begin{figure}[tb]
220           \begin{center}
221           \framebox{
222           \begin{minipage}{.65\textwidth}
223           \footnotesize
224           \begin{itemize}
225           \item MAIN MENU (Fig. \ref{fig:main_menu})
226              \begin{enumerate}
227              \item [1] USER DISPLAY (Fig.\ref{fig:user_display})
228              \item [2] EXTENDED DISPLAY (Fig.~\ref{fig:extended_display})
229              \item [3] PRESSURE CONTROL (Fig.~\ref{fig:pressure_control})
230              \item [4] DIAGNOSTICS
231                 \begin{enumerate}
232                 \item [4.1] ERROR LISTING
233                 \item [4.2] SIGNALS
234                 \end{enumerate}
235              \item [5] INSTRUMENT SETUP
236                 \begin{enumerate}
237                 \item [5.1] RANGE SELECTION (Fig.~\ref{fig:range_selection})
238                 \item [5.2] GAS SELECTION (Fig.~\ref{fig:gas_selection})
239 saw   1.5       \item [5.3] MODE SELECTION (Fig.~\ref{fig:mode_selection}) 
240                 \item [5.4] ZERO ADJUST
241                 \item [5.5] TRIP LIMITS
242                 \item [5.6] GAS COMPOSITION (Fig.~\ref{fig:gas_composition})
243                 \end{enumerate}
244              \item [6] SYSTEM SETUP
245              \item [7] PRESSURE SETUP 
246              \item [ ]
247              \item [9] INFORMATION
248              \item [0] POWER OFF
249              \end{enumerate}
250           \end{itemize}
251           \end{minipage}
252           } %%end of \framebox{
253           \caption{Command Tree for the MKS-647C Control Panel}
254           \label{fig:command_tree}
255           \end{center}
256           \end{figure}
257           %
258 jones 1.3 \begin{center}
259           \begin{figure}[hbt]
260           \begin{minipage}{2.7in}
261 saw   1.5 \psfig{figure=drift_gas_system-main_menu.eps,width=2.6in,height=1.8in}
262           \caption{MKS~647 Main Menu\label{fig:main_menu}}
263 jones 1.3 \end{minipage}
264           \begin{minipage}{2.7in}
265 saw   1.5 \psfig{figure=drift_gas_system-user_display.eps,width=2.6in,height=1.8in}
266           \caption{MKS~647 User Display\label{fig:user_display}}
267 jones 1.3 \end{minipage}
268           \end{figure}
269           \end{center}
270 saw   1.5 %
271 jones 1.3 \begin{center}
272           \begin{figure}[hbt]
273           \begin{minipage}{2.7in}
274 saw   1.5 \psfig{figure=drift_gas_system-extended_display.eps,width=2.6in,height=1.8in}
275 jones 1.3 \caption{Extended Display Screen\label{fig:extended_display}}
276           \end{minipage}
277           \begin{minipage}{2.7in}
278 saw   1.5 \psfig{figure=drift_gas_system-pressure_control.eps,width=2.6in,height=1.8in}
279 jones 1.3 \caption{Pressure Control Screen\label{fig:pressure_control}}
280           \end{minipage}
281           \end{figure}
282           \end{center}
283 saw   1.5 %=====================================================================
284 jones 1.3 \subsection{Gas Flow Rates}
285           \label{sec:gas_flow_rates}
286           The flow rates are adjusted automatically by the controller in order
287           to maintain a constant delivery pressure at the output. Only the flow
288 saw   1.5 {\em ratios} should be set by the operator. We use a 1:1 ratio, set on
289           the {\bf Gas Composition} screen (Fig.~\ref{fig:gas_composition}), as
290           indicated in Table~\ref{tab:mixer_nominals}.
291           
292           The average total flow should equal the sum of the flows to all of the
293           detectors in the shield house. (Note that the ball-type flowmeters in
294           the shield house are calibrated for nitrogen. The approximate
295           multiplier to convert these readings for 50/50 Argon-Ethane is 0.9 .)
296 jones 1.3 
297           System configuration parameters specifying the full-scale flow
298           capacity (for nitrogen) of each valve, the types of gases actually
299 saw   1.5 flowing through each valve, and the mode of control for the valves are
300           set in the screens pictured in Figs.~\ref{fig:range_selection},
301           \ref{fig:gas_selection}, and \ref{fig:mode_selection}. These figures
302           show the nominal settings for the Hall-C system.
303 jones 1.3 
304           \begin{center}
305           \begin{figure}[hbt]
306           \begin{minipage}{2.7in}
307 saw   1.5 \psfig{figure=drift_gas_system-gas_composition.eps,width=2.6in,height=1.8in}
308 jones 1.3 \caption{Gas Composition Screen\label{fig:gas_composition}}
309           \end{minipage}
310           \begin{minipage}{2.7in}
311 saw   1.5 \psfig{figure=drift_gas_system-range_selection.eps,width=2.6in,height=1.8in}
312 jones 1.3 \caption{Range Selection Screen\label{fig:range_selection}}
313           \end{minipage}
314           \end{figure}
315           \begin{figure}[hbt]
316           \begin{minipage}{2.7in}
317 saw   1.5 \psfig{figure=drift_gas_system-gas_selection.eps,width=2.6in,height=1.8in}
318 jones 1.3 \caption{Gas Selection Screen\label{fig:gas_selection}}
319           \end{minipage}
320           \begin{minipage}{2.7in}
321 saw   1.5 \psfig{figure=drift_gas_system-mode_selection.eps,width=2.6in,height=1.8in}
322 jones 1.3 \caption{Mode Selection Screen\label{fig:mode_selection}}
323           \end{minipage}
324           \end{figure}
325           \end{center}
326 saw   1.1 
327 saw   1.5 %=====================================================================
328 jones 1.3 \subsection{To set the Delivery Pressure:}
329 saw   1.1 
330 saw   1.5 Navigate to the {\bf Pressure Control} menu. The pressure set-point (in
331           Torr) is indicated at the bottom-center of the screen. This value
332           should be se to 500.0. Note that the system can not respond instantly
333           to a change in requested gas pressure: it has no way to release excess
334           pressure and must wait for the detector systems to consume it
335           \footnote{An \emph{over-pressure relief valve} releases gas through a
336           small oil bubbler in the gas shed if the pressure exceeds about
337           600~Torr.}; it cannot build up pressure any faster than the
338           flow-control valves can supply it. It may take thirty minutes or so
339           for the pressure regulation system to stabilize at a new set-point or
340           stabilize in response to a change in total gas consumption. However,
341           you should be able to observe a change in the gas flow within a few
342           seconds (possibly up to a minute) after a set-point change.
343 saw   1.1 
344 saw   1.5 %=====================================================================
345 saw   1.1 \subsection{To turn gas flow on or off:}
346           
347 saw   1.5 The gas flow can be turned on or off while in any menu.  In the
348           Extended Display menu the bottom line displays ``ON'' or ``OFF'', by
349           channel, to show which mass flow valves are enabled.  ``ON" must be
350           displayed in the bottom row of the Extended Display menu for gas to be
351           flowing in a particular channel.
352 saw   1.1 
353 saw   1.5 For gas to flow, two conditions must be met: 
354           \begin{enumerate}
355           \item Each channel (1 and 2) must be enabled by pressing ``ON" and then
356           the desired channel number.
357           \item The entire system must be enabled by pressing
358           ``ON" and then ``ALL" from the keypad. 
359           \end{enumerate}
360           Thus, each valve can be controlled individually using ``ON/OFF~-~\emph{Channel Number}'',
361           or all flow can be controlled using ``ON/OFF~-~ALL''. If the system
362           is enabled, the status line at the bottom of every screen will
363           indicate ``FLOW ON GAS ON''. {\bf Note:} because Valve-2 (ethane) is 
364           normally slaved to Valve-1 (argon), when Valve-1 is disabled there will
365           be no flow through Valve-2.
366           
367           %=====================================================================
368           %=====================================================================
369 saw   1.1 \section{To Change a Gas Bottle}
370           
371 saw   1.2 The argon and ethane supply bottles should be replaced by new (full)
372 saw   1.5 bottles when the bottle content drops below about 10\% of its
373           capacity.  For argon, the bottle content is directly indicated by the
374           bottle pressure: a new bottle usually contains 2000 to
375           3000~psig. Argon bottles should be changed whenever the bottle
376           pressure is found to be below about 200~psig. 
377           
378           Ethane bottles, on the
379           other hand, contain liquefied ethane. Thus the bottle pressure is just
380           the vapor pressure of ethane at whatever the current temperature
381           happens to be. At 70\degg F this is about 544~psig. The pressure gauge
382           will not tell you how much ethane is left in the bottle until it reads
383           zero! Instead, we measure the ethane content by observing the weight
384           of the bottle and comparing it to the weight when the bottle was
385           full. A standard B-size cylinder contains about 32~pounds of ethane.
386           The ethane cylinders on the manifold sit on scales which have been 
387           pre-set to indicate the net weight of ethane in the bottle.
388           Numbers in the green portion of the dial indicate ethane remaining.
389           If the indicator points to the red portion of the dial, the bottle
390           is empty.
391           
392           Handling and connecting bottles of compressed gas require special
393 saw   1.5 knowledge.  The high pressure gas stored in the cylinders (bottles)
394           constitutes significant stored energy. Mishandling of a gas bottle can
395           pose a lethal hazard! Refer to the JLab ESH\&Q Manual\cite{bi:jlabehs}
396           for safe handling practices. If you do not already know how to safely
397           manipulate compressed gas hardware, have a knowledgeable person train
398           you.
399           
400           After attaching a new gas bottle to the supply manifold, check the
401           connection for leaks using \emph{Snoop} or a similar leak detector.
402           
403           %=====================================================================
404           %=====================================================================
405           \section{The Alcohol Bubbler}
406 saw   1.1 
407 jones 1.3 To reduce the rate of aging of the wire chambers, the operating gas
408           contains a small quantity of alcohol vapor. The vapor is added by
409           bubbling the argon/ethane mixture through liquid alcohol. The
410           temperature of the alcohol controls the alcohol vapor pressure, which
411           determines the amount of vapor added to the gas. The alcohol content
412           also affects the electron drift velocity in the wire chambers, so it
413           must be held approximately constant.
414           
415 saw   1.5 Gas is bubbled through the liquid alcohol inside the glass
416           dome vessel in the refrigerator. The dome is covered by a
417           perforated steel cylinder as a precaution against breakage. The
418           alcohol level is controlled by a float valve inside the metal cold
419           reservoir, which is also inside the refrigerator. As long as there is
420           alcohol in the warm reservoir (sitting on top of the refrigerator),
421           the liquid levels inside the refrigerator will remain constant. A
422           drain valve (\#7) inside the refrigerator is available for emptying
423           all liquid from the system.  It is for use by experts only and should
424           remain closed during normal operation.
425           
426           {\bf The reservoir should be refilled before it becomes empty to
427           maintain a head of liquid over the float valve.  This will prevent air
428           from entering the system.}
429           %=====================================================================
430           \subsection{To by-pass the alcohol system}
431           
432           Open valve {\bf 3}, then promptly close valves {\bf 1 \& 2}, in that order!
433           
434           \noindent To restore flow through the bubbler, open valves {\bf 1 \& 2} then
435           close valve {\bf 3}.
436 saw   1.5 
437           %=====================================================================
438           \subsection {Alcohol Temperature Control}
439           
440           To keep the alcohol temperature (and thus the vapor pressure) constant,
441           the alcohol bubbler is housed in a refrigerator which is controlled by
442           an electronic temperature regulator having 1~C\degg  sensitivity. The 
443           controller is located on a shelf in the right-hand rack of the gas mixing
444           system. Normally, the actual temperature in the refrigerator is
445           indicated on the front panel of the controller. The controller should
446           be set to maintain a temperature of 2\degg  C.
447           
448           %=====================================================================
449           \pagebreak[4] 
450           \subsection{Step-by-Step Instructions for Refilling the Alcohol Bubbler}
451           {\em These 13 steps must be individually completed in the order listed!}
452           Refer to Fig.~\ref{fig:gas_mixer_diagram}.\\
453           
454           \begin{minipage}{\textwidth}
455 saw   1.1 \begin{enumerate}
456 saw   1.5 \footnotesize{
457 jones 1.3 \item{Remove the three screws securing the cover on the alcohol fill-tank 
458           (the top-most metal tank at the rear of the mixer rack. Carefully remove the lid
459           without allowing dust or dirt to fall into the tank.}
460           \item{Fill this tank to about 75\% full from a bottle of 2-propanol.}
461           \item{Replace the lid and retaining screws.}
462           \item{Remove the small brass cap which seals the port on the lid.}
463           \item{Isolate the warm reservoir by {\bf closing valves 4 and 8.}}
464           \item{Release pressure in the warm reservoir by {\bf slowly opening valve 6 fully.}}
465           \item{{\bf Open valve 5 fully} to begin the flow of alcohol from the fill-tank to the warm reservoir.}
466           \item{{\bf Monitor the level} of alcohol in the warm reservoir sight-tube.}
467           \item{When the level is 2~cm from the top of the sight-tube, {\bf close valve 5.}}
468           \item{Replace the brass cap on the lid of the fill-tank.}
469           \item{{\bf Close valve 6, {\it then} slowly open valve 4.}}
470           \item{{\bf Open valve 8.}}
471 saw   1.5 \item{Record what you did in both the gas logbook and the electronic logbook.}
472           } %end of scriptsize 
473 saw   1.1 \end{enumerate}
474 saw   1.5 \end{minipage}
475 saw   1.1 
476 saw   1.5 %=====================================================================
477           %=====================================================================
478 saw   1.1 \section{Gas Filters Maintenance}
479           
480           There are gas filters on the argon and ethane supply lines just inside
481           the gas shed. These filters should be replaced on a regular schedule.
482           See Bill Vulcan for details.
483           
484 saw   1.5 %=====================================================================
485           %=====================================================================
486 saw   1.1 \section{Secure Pressure Regulators}
487           
488           The gas mixing system is protected from failure or mis-setting of the 
489           primary pressure regulators (the ones mounted on the manifolds on the
490           exterior of the gas shed -- near the bottles) by {\it hidden} regulators
491           mounted just inside the gas shed. It is these regulators which actually
492           set the maximum supply pressure to the mixing valves. These regulators
493           should {\em never} be adjusted by other than a gas system expert! The
494           nominal secondary pressure supplied by both the argon and ethane
495 jones 1.3 secure regulators is 30 psig.
496           
497 saw   1.5 %=====================================================================
498           %=====================================================================
499 jones 1.3 \section{Ethane Flow Restrictor}
500           
501           A calibrated flow-restricting orifice is installed at the outlet of
502           the main ethane high-pressure manifold outside the gas shed. This
503           orifice passively limits the flow rate of ethane into the gas shed
504           even if there is a catastrophic failure of the flow and pressure-controlling
505           devices inside the shed. It is model {\tt IC-DM4-9-SS} manufactured by
506           {\em O'Keefe Controls Co.}, Trumbull, CT, USA. While it may look like
507           simply a stainless-steel union fitting, it is in fact a precision part,
508           and a necessary component of the gas safety system. It limits the flow
509 saw   1.5 of ethane to less than 26.5 SLM, which is a little
510 jones 1.3 less than ten times the maximum capacity of the flow-control system.
511 saw   1.1 
512 saw   1.5 %=====================================================================
513           %=====================================================================
514           
515           \raggedright  %this fixes formatting of bibliography, too.
516 saw   1.1 
517           \section{Related {\it Howtos}}
518           \begin{itemize}
519 jones 1.3 \item MKS~647 Mass Flow Controller Manual \cite{647C_EN_0504A1}
520 saw   1.1 \item Base Equipment Shift Checklist Items \cite{howto:base_equip_checklist}
521           \end{itemize}
522           
523           \end{document}
524           
525           % Revision history:
526           %  1st draft by Howard Fenker 27FEB03 -- taken from existing ops manual.
527 jones 1.3 %  Rev. 1.1 - added notes on bottle changing, ethane bottle pressure.
528 saw   1.6 % $Log: drift_gas_system.tex,v $
529           % Revision 1.5  2008/01/02 16:46:03  saw
530           % Latest update
531           %

Analyzer/Replay: Mark Jones, Documents: Stephen Wood
Powered by
ViewCVS 0.9.2-cvsgraph-1.4.0