1 saw 1.1 \documentclass{chowto}
2
3 \title{Monitoring HMS ``Utilities'' Systems}
4 \howtotype{user} % ``expert'', ``user'', ``reference''
5 %\experiment{Name of experiment} % Optional
6 \author{H. Fenker}
7 \category{hms}
8
9 %\maintainer{Name of person maintaining document} % Optional
10 \date{April 4, 2003} % Can use \today as the argument
11
12 \begin{document}
13 \providecommand{\degg}{\ensuremath{^{\circ}\ }}
14
15 \begin{abstract}
16 Herein is descirbed how to verify the status of the services
17 required by the detectors in the HMS:
18 high and low-voltage electrical supplies, gas, and environmental
19 conditions. Basically, these are the {\it utilities} that the
20 detectors need in order to function. They should be continuously
21 monitored or, at a minimum, manually checked at least one time
22 saw 1.1 during each eight-hour shift while the detectors are to be
23 kept ready for data.
24
25 Note that users must also check the {\it quality} of the data
26 produced by the detectors. Maintenance of the {\it utilities}
27 described here is necessary but not necessarily sufficient
28 to maintain good detector performance.
29
30 \end{abstract}
31
32
33 \section{Room Temperature}
34
35 The ambient temperature inside the HMS shield house is indicated
36 by one of the two temperture gauges mounted on the wall of the
37 Counting Room above the far left of the control console. The
38 reading is in degrees Fahrenheit, and should be no higher than
39 about $75\degg F$.
40
41 \section{Drift Chambers}
42
43 saw 1.1 The HMS includes two identical drift chambers, each of which
44 contains six planes of sense wires.
45
46 \subsection{High Voltage}
47
48 Each sense plane requires
49 three high-voltage supplies (denoted ``triangle'', ``square'',
50 and ``circle''). Further, each chamber uses an additional
51 high-voltage channel to supply the ``guard'' wires. The
52 nominal voltages are shown in \cite{howto:hms_nominal_settings}.
53 Check and modify the voltages using the High Voltage GUI
54 \cite{howto:caen_hv_gui}. Readback voltages should be within
55 about 20~V of the setpoint voltages. Normally the GUI
56 monitors the voltages and sounds an alarm if a channel is out of tolerance.
57 In this case, the user must simply verify that the alarm has
58 not been turned off. Typical currents in the wire chambers are less
59 than 10~$\mu$A.
60
61
62 \subsection {Low Voltage}
63
64 saw 1.1 The drift chamber preamplifiers, located on the side and top of
65 the chambers, require three supply voltages: +5V, -5V, and
66 V$_{threshold}$. The $\pm 5V$ supplies are located in the HMS
67 shield house and cannot be monitored remotely. (If one of them
68 dies or a fuse blows, the entire effected chamber will have
69 no hits.) The threshold-voltage supplies, one for each chamber,
70 are located in rack
71 CH03B10 in the Counting house Electronics Room. The nominal
72 V$_{threshold}$ is noted in \cite{howto:hms_nominals}.
73 It is only to be altered
74 by a staff expert, who will post the new nominal value at the
75 power supply if it differs from the value mentioned here.
76
77 These threshold voltages should be checked using {\bf only}
78 the digital voltmeter built in to the power supply.
79
80 \subsection {Gas}
81
82 The drift chambers must be continuously flushed with fresh
83 operating gas. The gas is supplied from the gas shed. It is
84 composed of approximately 49.5\% argon, 49.5\% ethane, and
85 saw 1.1 1\% {\it 2-propanol} (isopropyl alcohol). This mixture is automatically
86 controlled by the gas mixing system \cite{howto:drift_gas_system}.
87 Users must simply verify that the readings in the gas shed
88 are correct and that the gas supply does not run out. The
89 amount of argon left in a bottle can be determined from the
90 bottle pressure, while the ethane content may only be measured
91 by observing the drop in gross bottle weight. Physicists
92 should understand why this is true.
93
94 Nominal settings are given in \cite{howto:hms_nominal_settings}.
95
96 \section{Trigger Scintillators}
97
98 The trigger scintillators consist of two sets of crossed
99 scintillator paddles (H1X, H1Y) and (H2X, H2Y). Each paddle
100 has two phototubes (one at each end). The $X$ planes have
101 sixteen paddles and the $Y$ planes have ten paddles each.
102
103 A hodoscope system expert will determine the appropriate high
104 voltage for each of the 104 phototubes during the commissioning
105 period of each experiment. Check the logbook for a record of what
106 saw 1.1 the present values are -- they should also be saved by the HV GUI.
107 The settings listed in \cite{howto:hms_nominal_settings}
108 are typical values only.
109
110 All of the HMS hodoscope tube bases have a load resistance of $1M\Omega$
111 and should, therefore, draw in the neighborhood of 2.5~mA for a
112 supply voltage of 2500~V.
113
114 \section{Aerogel \v{C}erenkov Counter}
115
116 The aerogel \v{C}erenkov counter needs high voltage for each of
117 its phototubes. System experts will update the settings in the
118 HV GUI and make a log entry. High voltages are monitored by the
119 GUI, which sounds an alarm in case of error (if the GUI is running).
120 Typical voltage settings are given in \cite{howto:hms_nominal_settings}.
121
122 \section{Gas Cerenkov Counter}
123
124 The HMS Gas \v{C}erenkov needs high voltage for each of its
125 two photomultiplier tubes and needs to have constant pressure
126 and temperature.
127 saw 1.1
128 \subsection{High Voltage}
129 The high voltages are supplied by the HMS
130 high voltage GUI, of course. The PMT bases for the gas \v{C}erenkov
131 use {\it positive} high voltage. Note that all the other PMT bases
132 in the HMS are designed for use with {\it negative} HV.
133 They operate at between 2000 and 2700 Volts. Nominal voltages
134 are given in \cite{howto:hms_nominal_settings}.
135
136
137 \subsection{Pressure and Temperature}
138 The pressure and temperature are
139 displayed by digital indicators located underneath the shield
140 house, near the primary magnet control screen. The displays
141 are viewed by a TV camera feeding a monitor in the Counting
142 Room. The two numbers shown are the temperature in $C\degg$
143 and the pressure in psia. See \cite{howto:hms_nominal_settings}.
144 The nominal pressure varies by experiment; the desired value
145 should be noted in the logbook during experiment commissioning.
146
147 \section{Lead-glass Calorimeter}
148 saw 1.1
149 The calorimeter consists of four layers of thirteen glass blocks
150 each. Layers A and B have two phototubes per block. layers $C$ and
151 $D$ have one tube per block. The HV GUI will sound an alarm if
152 a channel's voltage differs significantly from the setpoint.
153 Typical voltages are shown in \cite{howto:hms_nominal_settings}.
154
155 \section{Related {\it Howtos}}
156 \begin{itemize}
157 \item High Voltage GUI
158 \item HMS Nominal Settings \cite{howto:hms_nominal_settings}.
159 \end{itemize}
160
161 \end{document}
162
163 % Revision history:
164 % $Log$
|