Return to hms_detector_monitoring.tex CVS log | Up to [HallC] / Documents / Howtos |
File: [HallC] / Documents / Howtos / hms_detector_monitoring.tex
(download)
/
(as text)
Revision: 1.1, Fri Apr 4 16:02:23 2003 UTC (21 years, 5 months ago) by saw Branch: MAIN CVS Tags: mar2005, HEAD Branch point for: hks05 Initial Checkin |
\documentclass{chowto} \title{Monitoring HMS ``Utilities'' Systems} \howtotype{user} % ``expert'', ``user'', ``reference'' %\experiment{Name of experiment} % Optional \author{H. Fenker} \category{hms} %\maintainer{Name of person maintaining document} % Optional \date{April 4, 2003} % Can use \today as the argument \begin{document} \providecommand{\degg}{\ensuremath{^{\circ}\ }} \begin{abstract} Herein is descirbed how to verify the status of the services required by the detectors in the HMS: high and low-voltage electrical supplies, gas, and environmental conditions. Basically, these are the {\it utilities} that the detectors need in order to function. They should be continuously monitored or, at a minimum, manually checked at least one time during each eight-hour shift while the detectors are to be kept ready for data. Note that users must also check the {\it quality} of the data produced by the detectors. Maintenance of the {\it utilities} described here is necessary but not necessarily sufficient to maintain good detector performance. \end{abstract} \section{Room Temperature} The ambient temperature inside the HMS shield house is indicated by one of the two temperture gauges mounted on the wall of the Counting Room above the far left of the control console. The reading is in degrees Fahrenheit, and should be no higher than about $75\degg F$. \section{Drift Chambers} The HMS includes two identical drift chambers, each of which contains six planes of sense wires. \subsection{High Voltage} Each sense plane requires three high-voltage supplies (denoted ``triangle'', ``square'', and ``circle''). Further, each chamber uses an additional high-voltage channel to supply the ``guard'' wires. The nominal voltages are shown in \cite{howto:hms_nominal_settings}. Check and modify the voltages using the High Voltage GUI \cite{howto:caen_hv_gui}. Readback voltages should be within about 20~V of the setpoint voltages. Normally the GUI monitors the voltages and sounds an alarm if a channel is out of tolerance. In this case, the user must simply verify that the alarm has not been turned off. Typical currents in the wire chambers are less than 10~$\mu$A. \subsection {Low Voltage} The drift chamber preamplifiers, located on the side and top of the chambers, require three supply voltages: +5V, -5V, and V$_{threshold}$. The $\pm 5V$ supplies are located in the HMS shield house and cannot be monitored remotely. (If one of them dies or a fuse blows, the entire effected chamber will have no hits.) The threshold-voltage supplies, one for each chamber, are located in rack CH03B10 in the Counting house Electronics Room. The nominal V$_{threshold}$ is noted in \cite{howto:hms_nominals}. It is only to be altered by a staff expert, who will post the new nominal value at the power supply if it differs from the value mentioned here. These threshold voltages should be checked using {\bf only} the digital voltmeter built in to the power supply. \subsection {Gas} The drift chambers must be continuously flushed with fresh operating gas. The gas is supplied from the gas shed. It is composed of approximately 49.5\% argon, 49.5\% ethane, and 1\% {\it 2-propanol} (isopropyl alcohol). This mixture is automatically controlled by the gas mixing system \cite{howto:drift_gas_system}. Users must simply verify that the readings in the gas shed are correct and that the gas supply does not run out. The amount of argon left in a bottle can be determined from the bottle pressure, while the ethane content may only be measured by observing the drop in gross bottle weight. Physicists should understand why this is true. Nominal settings are given in \cite{howto:hms_nominal_settings}. \section{Trigger Scintillators} The trigger scintillators consist of two sets of crossed scintillator paddles (H1X, H1Y) and (H2X, H2Y). Each paddle has two phototubes (one at each end). The $X$ planes have sixteen paddles and the $Y$ planes have ten paddles each. A hodoscope system expert will determine the appropriate high voltage for each of the 104 phototubes during the commissioning period of each experiment. Check the logbook for a record of what the present values are -- they should also be saved by the HV GUI. The settings listed in \cite{howto:hms_nominal_settings} are typical values only. All of the HMS hodoscope tube bases have a load resistance of $1M\Omega$ and should, therefore, draw in the neighborhood of 2.5~mA for a supply voltage of 2500~V. \section{Aerogel \v{C}erenkov Counter} The aerogel \v{C}erenkov counter needs high voltage for each of its phototubes. System experts will update the settings in the HV GUI and make a log entry. High voltages are monitored by the GUI, which sounds an alarm in case of error (if the GUI is running). Typical voltage settings are given in \cite{howto:hms_nominal_settings}. \section{Gas Cerenkov Counter} The HMS Gas \v{C}erenkov needs high voltage for each of its two photomultiplier tubes and needs to have constant pressure and temperature. \subsection{High Voltage} The high voltages are supplied by the HMS high voltage GUI, of course. The PMT bases for the gas \v{C}erenkov use {\it positive} high voltage. Note that all the other PMT bases in the HMS are designed for use with {\it negative} HV. They operate at between 2000 and 2700 Volts. Nominal voltages are given in \cite{howto:hms_nominal_settings}. \subsection{Pressure and Temperature} The pressure and temperature are displayed by digital indicators located underneath the shield house, near the primary magnet control screen. The displays are viewed by a TV camera feeding a monitor in the Counting Room. The two numbers shown are the temperature in $C\degg$ and the pressure in psia. See \cite{howto:hms_nominal_settings}. The nominal pressure varies by experiment; the desired value should be noted in the logbook during experiment commissioning. \section{Lead-glass Calorimeter} The calorimeter consists of four layers of thirteen glass blocks each. Layers A and B have two phototubes per block. layers $C$ and $D$ have one tube per block. The HV GUI will sound an alarm if a channel's voltage differs significantly from the setpoint. Typical voltages are shown in \cite{howto:hms_nominal_settings}. \section{Related {\it Howtos}} \begin{itemize} \item High Voltage GUI \item HMS Nominal Settings \cite{howto:hms_nominal_settings}. \end{itemize} \end{document} % Revision history: % $Log: hms_detector_monitoring.tex,v $ % Revision 1.1 2003/04/04 16:02:23 saw % Initial Checkin %
Analyzer/Replay: Mark Jones, Documents: Stephen Wood |
Powered by ViewCVS 0.9.2-cvsgraph-1.4.0 |