Return to hms_nominals.tex CVS log | Up to [HallC] / Documents / Howtos |
File: [HallC] / Documents / Howtos / hms_nominals.tex
(download)
/
(as text)
Revision: 1.3, Fri Apr 4 15:08:27 2003 UTC (21 years, 5 months ago) by saw Branch: MAIN CVS Tags: mar2005 Branch point for: hks05 Changes since 1.2: +18 -9 lines Fix some citations, add nominal Aerogel voltages |
\documentclass{chowto} \title{HMS Nominal Settings: Voltages, Currents, Flows, etc.} \howtotype{reference} % ``expert'', ``user'', ``reference'' %\experiment{Name of experiment} % Optional \author{H. Fenker} \category{hms} % Subject area of this document %\maintainer{Name of person maintaining document} % Optional \date{March 6, 2003} % Can use \today as the argument \begin{document} \providecommand{\degg}{\ensuremath{^{\circ}\ }} \begin{abstract} This document collects the nominal setpoints for all HMS detector utility supplies. Phototube and wire chamber high voltages may change over time, of course. The values listed here were taken from the control system as found on March 6, 2003. \end{abstract} \section{Related {\it Howtos}} \begin{itemize} \item Monitoring HMS Utilities Systems \cite{howto:hms_detector_monitoring} \item Base Equipment Checklist \cite{howto:base_equip_checklist} \item Monitoring Gas Mixing System and Interpret Alarms \cite{howto:response_to_flam_gas_alarm} \item HV Use and Reconfiguration \item Drift Gas System \cite{howto:drift_gas_system} \end{itemize} \section{Tables of Typical / Nominal Readings} \begin{table}[hp] \centering \begin{tabular}{|c|l|r|r|r|} \hline Chamber & Plane & Triangle & Square & Circle \\ \hline 1 & X1 & 2500 & 2250 & 1800 \\ 1 & X2 & 2500 & 2250 & 1800 \\ 1 & U & 2500 & 2250 & 1800 \\ 1 & V & 2500 & 2250 & 1800 \\ 1 & Y1 & 2500 & 2250 & 1800 \\ 1 & Y2 & 2500 & 2250 & 1800 \\ 1 & \multicolumn{4}{l|}{Guard Plane: 1500} \\ 2 & X1 & 2500 & 2250 & 1800 \\ 2 & X2 & 2500 & 2250 & 1800 \\ 2 & U & 2500 & 2250 & 1800 \\ 2 & V & 2500 & 2250 & 1800 \\ 2 & Y1 & 2500 & 2250 & 1800 \\ 2 & Y2 & 2500 & 2250 & 1800 \\ 2 & \multicolumn{4}{l|}{Guard Plane: 1500} \\ \hline \end{tabular} \caption{Nominal HMS Drift Chamber High Voltage Setpoints (Volts). Normal currents for wire chamber high voltages are less than $10\mu A$. \label{tab:drift_hv_settings}} \end{table} \begin{table}[hp] \centering \small \begin{tabular}{|c|r|r|r|l|} \hline Item & Voltage & Tolerance & Current & Location\\ \hline +5 Preamp Supply & 5.0 & $\pm 0.2$ & 40 A& HMS Hut \\ -5 Preamp Supply & 5.0 & $\pm 0.2$ & 50 A& HMS Hut \\ DC1 $V_t$ & +5.5 & $\pm 0.1$ & 0 & CR Rack CH03B10\\ DC2 $V_t$ & +5.5 & $\pm 0.1$ & 0 & CR Rack CH03B10 \\ \hline \end{tabular} \caption{Nominal drift chamber low voltages and currents \label{tab:hms_drift_lv_settings}} \end{table} \begin{table}[hp] \centering \small \begin{tabular}{|l|l|l|l|l|} \hline Item & Nominal & Units &Tol. & Where to Look\\ \hline Ethane Bottle Pressure & \multicolumn{4}{|l|}{ Temperature Dependent -- {\it practically meaningless}}\\ Ethane Bottle Weight &\multicolumn{3}{|l|}{Full weight - (0 to 12) lbs} & above bottle\\ Ethane Supply Pressure & 20 & psig & -1 +5 & manifold outside\\ Argon Bottle Pressure & $> 200$ & psig & & manifold outside \\ Argon Supply Pressure & 30 & psig & -1 +5 & manifold outside \\ HMS Ethane Flow & 0.3 & slpm & $\pm 0.01$ & CRT channel 3 \\ HMS Argon Flow & 0.3 & slpm & $\pm 0.01$ & CRTchannel 4 \\ Alcohol Temperature & 1 & $ \degg C $& $\pm 3~C \degg$&Temp. Control Box \\ Alcohol Level & about 2 & cm &$ \pm 1cm$ & Glass Flask in Ref. \\ \hline \end{tabular} \caption{Nominal drift chamber gas supply readings. The equipment referenced is inside the gas shed and under the lean-to outside the gas shed. \label{tab:gas_shed_settings}} \end{table} \begin{table}[hp] \centering \small \begin{tabular}{|c|c||l|r||l|r|} \hline Plane & Paddle & Channel & Voltage & Channel & Voltage \\ \hline 1x & 1 & h1x01+ & 2214 & h1x01- & 2362 \\ 1x & 2 & h1x02+ & 2445 & h1x02- & 2585 \\ 1x & 3 & h1x03+ & 2569 & h1x03- & 2658 \\ 1x & 4 & h1x04+ & 2415 & h1x04- & 2368 \\ 1x & 5 & h1x05+ & 2664 & h1x05- & 2335 \\ 1x & 6 & h1x06+ & 2360 & h1x06- & 2385 \\ 1x & 7 & h1x07+ & 2368 & h1x07- & 2495 \\ 1x & 8 & h1x08+ & 2510 & h1x08- & 2434 \\ 1x & 9 & h1x09+ & 2639 & h1x09- & 2700 \\ 1x & 10 & h1x10+ & 2427 & h1x10- & 2570 \\ 1x & 11 & h1x11+ & 2668 & h1x11- & 2440 \\ 1x & 12 & h1x12+ & 2574 & h1x12- & 2920 \\ 1x & 13 & h1x13+ & 2460 & h1x13- & 2575 \\ 1x & 14 & h1x14+ & 2713 & h1x14- & 2655 \\ 1x & 15 & h1x15+ & 2500 & h1x15- & 2912 \\ 1x & 16 & h1x16+ & 2500 & h1x16- & 2665 \\ \hline 1y & 1 & h1y01+ & 2680 & h1y01- & 2550 \\ 1y & 2 & h1y02+ & 2575 & h1y02- & 2640 \\ 1y & 3 & h1y03+ & 2859 & h1y03- & 2680 \\ 1y & 4 & h1y04+ & 2684 & h1y04- & 2704 \\ 1y & 5 & h1y05+ & 2539 & h1y05- & 2400 \\ 1y & 6 & h1y06+ & 2646 & h1y06- & 2355 \\ 1y & 7 & h1y07+ & 2811 & h1y07- & 2630 \\ 1y & 8 & h1y08+ & 2460 & h1y08- & 2445 \\ 1y & 9 & h1y09+ & 2761 & h1y09- & 2732 \\ 1y & 10 & h1y10+ & 2700 & h1y10- & 2500 \\ \hline \end{tabular} \caption{Typical HMS Trigger Scintillator $1$ High Voltages. All bases draw roughly 2.0-3.0~mA (loaded at $1M\Omega$). \label{tab:hms_scint1_hv}} \end{table} \begin{table}[hp] \centering \small \begin{tabular}{|c|c||l|r||l|r|} \hline Plane & Paddle & Channel & Voltage & Channel & Voltage \\ \hline 2x & 1 & h2x01+ & 1800 & h2x01- & 2950 \\ 2x & 2 & h2x02+ & 2950 & h2x02- & 2460 \\ 2x & 3 & h2x03+ & 2955 & h2x03- & 2715 \\ 2x & 4 & h2x04+ & 2860 & h2x04- & 2570 \\ 2x & 5 & h2x05+ & 2520 & h2x05- & 2445 \\ 2x & 6 & h2x06+ & 2358 & h2x06- & 2618 \\ 2x & 7 & h2x07+ & 2605 & h2x07- & 2645 \\ 2x & 8 & h2x08+ & 2660 & h2x08- & 2655 \\ 2x & 9 & h2x09+ & 2442 & h2x09- & 2660 \\ 2x & 10 & h2x10+ & 2700 & h2x10- & 2575 \\ 2x & 11 & h2x11+ & 2573 & h2x11- & 2610 \\ 2x & 12 & h2x12+ & 2750 & h2x12- & 2900 \\ 2x & 13 & h2x13+ & 2750 & h2x13- & 2342 \\ 2x & 14 & h2x14+ & 2575 & h2x14- & 2670 \\ 2x & 15 & h2x15+ & 2597 & h2x15- & 2615 \\ 2x & 16 & h2x16+ & 2550 & h2x16- & 2451 \\ \hline 2y & 1 & h2y01+ & 2600 & h2y01- & 2750 \\ 2y & 2 & h2y02+ & 2735 & h2y02- & 2895 \\ 2y & 3 & h2y03+ & 2750 & h2y03- & 2755 \\ 2y & 4 & h2y04+ & 2350 & h2y04- & 2900 \\ 2y & 5 & h2y05+ & 2880 & h2y05- & 2775 \\ 2y & 6 & h2y06+ & 2890 & h2y06- & 2950 \\ 2y & 7 & h2y07+ & 2900 & h2y07- & 2770 \\ 2y & 8 & h2y08+ & 2800 & h2y08- & 2780 \\ 2y & 9 & h2y09+ & 2780 & h2y09- & 2860 \\ 2y & 10 & h2y10+ & 2910 & h2y10- & 2900 \\ \hline \end{tabular} \caption{Typical HMS Trigger Scintillator $2$ High Voltages. All bases draw roughly 2.0-3.0~mA (loaded at $1M\Omega$). \label{tab:hms_scint2_hv}} \end{table} \begin{table}[hp] \centering \small \begin{tabular}{|c||c|c||c|c|} \hline Section & Channel & High Voltage & Channel & High Voltage \\ \hline 1 & hAero1+ & 1570 & hAero1- & 1640 \\ 2 & hAero2+ & 1690 & hAero2- & 1590 \\ 3 & hAero3+ & 1580 & hAero3- & 1750 \\ 4 & hAero4+ & 1680 & hAero4- & 2040 \\ 5 & hAero5+ & 1690 & hAero5- & 1910 \\ 6 & hAero6+ & 1620 & hAero6- & 1900 \\ 7 & hAero7+ & 1700 & hAero7- & 1840 \\ 8 & hAero8+ & 1660 & hAero8- & 1750 \\ \hline \end{tabular} \caption{Nominal HMS Aerogel High Voltage Setpoints (Volts) [Voltages as found 3/28/03]. \label{tab:hms_aero_hv}} \end{table} \begin{table}[hp] \centering \small \begin{tabular}{|c|l|r|r|r|} \hline Channel & High Voltage \\ \hline 1 & 2535 \\ 2 & 2550 \\ \hline Temperature & $35\degg C$ \\ Pressure & varies by experiment \\ \hline \end{tabular} \caption{Nominal HMS Gas \v{C}erenkov Conditions. Both photomultiplier tubes draw approximately 1~mA. \label{tab:hms_gas_cer_hv}} \end{table} \begin{table}[hp] \centering \small \begin{tabular}{|c|c|l|r||c|c|l|r|} \hline Layer & Block & Channel & Voltage & Layer & Block & Channel & Voltage \\ \hline A & 1 & hA01 & 1750 & A & 1 & hA01- & 1750\\ A & 2 & hA02 & 1500 & A & 2 & hA02- & 1600\\ A & 3 & hA03 & 1400 & A & 3 & hA03- & 1550\\ A & 4 & hA04 & 1550 & A & 4 & hA04- & 1550\\ A & 5 & hA05 & 1550 & A & 5 & hA05- & 1510\\ A & 6 & hA06 & 1500 & A & 6 & hA06- & 1650\\ A & 7 & hA07 & 1500 & A & 7 & hA07- & 1750\\ A & 8 & hA08 & 1550 & A & 8 & hA08- & 1750\\ A & 9 & hA09 & 1450 & A & 9 & hA09- & 1600\\ A & 10 & hA10 & 1550 & A & 10 & hA10- & 1700\\ A & 11 & hA11 & 1550 & A & 11 & hA11- & 1550\\ A & 12 & hA12 & 1500 & A & 12 & hA12- & 1520\\ A & 13 & hA13 & 1650 & A & 13 & hA13- & 1550\\ \hline B & 1 & hB01 & 1600 & B & 1 & hB01- & 1610\\ B & 2 & hB02 & 1500 & B & 2 & hB02- & 1750\\ B & 3 & hB03 & 1470 & B & 3 & hB03- & 1600\\ B & 4 & hB04 & 1500 & B & 4 & hB04- & 1530\\ B & 5 & hB05 & 1600 & B & 5 & hB05- & 1550\\ B & 6 & hB06 & 1500 & B & 6 & hB06- & 1600\\ B & 7 & hB07 & 1500 & B & 7 & hB07- & 1600\\ B & 8 & hB08 & 1550 & B & 8 & hB08- & 1520\\ B & 9 & hB09 & 1600 & B & 9 & hB09- & 1550\\ B & 10 & hB10 & 1600 & B & 10 & hB10- & 1550\\ B & 11 & hB11 & 1630 & B & 11 & hB11- & 1650\\ B & 12 & hB12 & 1550 & B & 12 & hB12- & 1630\\ B & 13 & hB13 & 1700 & B & 13 & hB13- & 1520\\ \hline C & 1 & hC01 & 1700 & D & 1 & hD01 & 1760 \\ C & 2 & hC02 & 1450 & D & 2 & hD02 & 1450 \\ C & 3 & hC03 & 1450 & D & 3 & hD03 & 1500 \\ C & 4 & hC04 & 1500 & D & 4 & hD04 & 1450 \\ C & 5 & hC05 & 1450 & D & 5 & hD05 & 1370 \\ C & 6 & hC06 & 1480 & D & 6 & hD06 & 1350 \\ C & 7 & hC07 & 1530 & D & 7 & hD07 & 1500 \\ C & 8 & hC08 & 1500 & D & 8 & hD08 & 1340 \\ C & 9 & hC09 & 1420 & D & 9 & hD09 & 1380 \\ C & 10 & hC10 & 1520 & D & 10 & hD10 & 1510 \\ C & 11 & hC11 & 1500 & D & 11 & hD11 & 1450 \\ C & 12 & hC12 & 1540 & D & 12 & hD12 & 1450 \\ C & 13 & hC13 & 1450 & D & 13 & hD13 & 1600\\ \hline \end{tabular} \caption{Typical HMS Shower Counter High Voltage Setpoints (Volts). All tube bases draw about 2.5~mA. \label{tab:hms_lg_hv}} \end{table} \end{document} % Revision history: % $Log: hms_nominals.tex,v $ % Revision 1.3 2003/04/04 15:08:27 saw % Fix some citations, add nominal Aerogel voltages % % Revision 1.2 2003/03/12 16:36:51 saw % Add some related howtos, fix rack numbers for low voltage % % Revision 1.1 2003/03/08 03:56:25 saw % *** empty log message *** %
Analyzer/Replay: Mark Jones, Documents: Stephen Wood |
Powered by ViewCVS 0.9.2-cvsgraph-1.4.0 |