Login any PC of cdaql1 - cdaql6 with 'cvxwrks' and go to /home/cvxwrks/MEDM/HKS.
Type ./sst-control.tcl and see the window of Sieve slits / Target
control for E01-011. You can operate sieve slits and the target ladder from
this window. The direction is moving to left on the camera (from the upstream).
The top column is for the HKS sieve slit and the middle is for the Enge sieve slit.
> `STOP!' stop program,
> `Go home' move to HOME switch,
> `Collimator' move to collimeter,
> `Sieve' move to sieve slit,
> `Empty' move the box out from the beamline,
> `Go up' move up by +10 (+1) for HKS (Enge) S.S,
> `Go dw' move down by -10 (-1) for HKS (Enge) S.S,
To move,
- Push the button of `Go home' for a zero-position adjustment at first, before moving.
- See a message 'Moving..' on a right-white window after a few seconds.
- After a message of `Stopping...' is shown up, Check the current position on the `Position' window.
- Confirm `Position' is `0'.
- Push the button of `Collimator'/`Sieve'/'Empty' to move the box to the `Collimator'/`Sieve'/'Empty' position.
- After a message of `Stopping...' is shown up, Check the current position on the `Position' window.
- Start run.
In a case of emergency,
- Push the button of `STOP!' to stop running program.