NEXT
User name horn
Log entry time 21:50:18 on March 14, 2005
Entry number 539
keyword=Horizontal Beam Position
In order to determine the absolute horizontal position of the beam in the
target-beam system two effects have to be taken into account: The
horizontal position of the target itself and the horizontal position of
the beam relative to the target system.
Due to the circular geometry of the target cells used a horizontal offset
of the target (>2mm -> 1% difference in length, Fig 1) will result in a
significant reduction of the target length relative to the nominal cell
diameter. Combining the inital target survey result with the measured
values for target vacuum motion (1.20 mm left) and target cooldown motion
(3.22mm left) results in a horizontal target offset of 3.42mm beam left.
This value agrees well with the observed burn marks on the
target cell, which are located ~3-4mm beam left of the center of the target.
To determine the relative horizontal beam position in the target system a
combination of the horizontal BPM information projected to the target and
the Ytar reconstruction from optics data (sieve measurement, carbon z=0
target) is used. From the reconstructed optics data the beam location in
the target x-z plane can be depicted as the intersection of two lines
parallel to the line emanating from the origin of the x-z coordinate
system with a slope given by the central spectrometer angle. The parallel
distances are determined by the value of the reconstructed Ytar offset
corrected for the spectrometer mispointing . Further discussion and a visual representation of
this procedure by D. Gaskell for the HMS specific case is available here. Although it is possible to use the information from one
spectrometer only, this analysis will use the information from both HMS
and SOS due to the limited angle range of the available data.
The following two tables summarize the sieve data used in determining the
horizontal beam offset:
Distances of BPM to the target:
BPM A=327.15cm
BPM B=231.47cm
BPM C=135.38cm
HMS:
---
theta Run Ytar(mm) HMisptg Ycor(mm) BPMAx BPMBx xtar(proj)
12.00 46996 0.16 0.19d 0.35 0.93 0.81 0.52
12.00 47012 0.13 0.19d 0.32 0.97 0.86 0.59
19.99 47037 1.04 0.0 1.04 0.81 0.70 0.43
12.00 47336 0.31 0.19d 0.50 0.98 0.81 0.40
12.00 47525 -1.38 0.19d -1.19 1.05 1.29 1.87
12.00 47526 -1.42 0.19d -1.23 1.04 1.27 1.83
10.55 47556 -1.72 0.14d -1.58 1.06 1.30 1.88
12.55 47619 -1.64 0.19d -1.45 1.06 1.47 2.46
15.00 47631 -0.55 0.25d -0.30 1.49 1.46 1.39
15.00 47644 -0.45 0.25d -0.20 1.56 1.53 1.46
15.00 47645 -0.36 0.25d -0.11 1.36 1.33 1.26
SOS
---
theta Run Ytar(mm) HMisptg Ycor(mm)
30.0 46996 -0.71 0.95d -1.66
30.0 47012 -0.81 0.95d -1.76
30.0 47037 -0.59 0.95d -1.54
30.0 47336 -0.90 0.95d -1.85
25.0 47525 -1.43 1.08d -2.51
25.0 47526 -1.51 1.08d -2.59
25.0 47556 -1.52 1.08d -2.61
27.0 47619 -2.14 1.12d -3.26
30. 47631 -1.90 0.95d -2.85
23.0 47644 -1.21 1.05d -2.26
23.0 47645 -1.02 1.05d -2.07
Note that horizontal mispointing downstream implies a different sign of
the correction for HMS/SOS due to the definition of +y in the
spectrometer coordinate system.
Depicted in Figure 2 are the intersection points in the target x-z plane
for each beam energy considered. In this coordinate system the vertical
axis represents the target z-position (along the beam) and the horizontal
the horizontal beam position. The green lines indicate the values of the
initial and final survey values for the target position along the beam.
The data suggests that it is not possible to use a single value for the
horizontal beam position throughout the entire run. However, given the
stability of the beam positions within one kinematic setting (defined
here by the beam energy) one value can be assigned to each such setting.
E_e x(mm) z(mm)
4.2 +0.25 +3.15
5.2 +0.15 +3.15
3.8 +2.00 +3.51
4.7 +1.00 +3.15
Fig 1
target_len.eps
Fig 2

Fig 3

Fig 4
bpm_epics_ytar_test.eps