    User name I. Shin

    Log entry time 16:19:26 on January 28,1999

    Entry number 267

    keyword=Charge survey for BCM2.

    Here are Beam Charge survey plots for BCM2.
    In figure 1, there are two guys which are obviously strange. Those run number are
    17029 and 17360. We got the good yield for 17029 with Beam on current Q_tot in
    stead of BCM2 Q_tot. Run 17360 might have long pause of beam.
    Followings are from gen17360.txt. BCM Q_tot is quite different from Beam on current Q_tot.

             I_avg           Q_tot             Q_tot
          (microAmps)   (microCoulombs)   (microAmp-hours)
    BCM1      2.916          2118.689           0.589
    BCM2      2.964          2153.535           0.598
    BCM3      3.774          2742.709           0.762
     
    ***********
    * Beam On *
    ***********
    Threshold       =     5.0
    Beam on time    =    25.78 sec        (   3.5 % )
     
             I_avg           Q_tot             Q_tot
          (microAmps)   (microCoulombs)   (microAmp-hours)
    bocur     89.374          2303.823           0.640
    


    Figure 2 is more closer look of figure 1. It seems that every charge ratio are smaller than or equal to 1.

    Figure 3 was made to see whether correlation exist between time and charge or not.
    The charge ratio decreases as time ratio decreases. We need not care about the one
    run which is located at roughly (0.86,0.99) because it is not Physics run.
    Charge ratio not equal to zero is largely due to current monitor offset when beam is off.

    Conclusion : We should use gated charge.

    And we will use smaller threshold than 5.0 micro Ampere in replay_4.



    FIGURE 1


    FIGURE 2


    FIGURE 3