Log entry time 13:50:18 on January 07, 2011

Entry number 212856

keyword=Target: manifold in B is off 1 degree

Since this thermometer is 1 degree less than all the others in the loop, we assume it is wrong. Even its partner in the same feedthru (QWT_miA) reados the same 20 K as the other thermometers. SO we assume QWT_miB is radiation damaged or something.

Note we are pidding on QWT_poA...






Figure 1