keyword=Log file to accompany low current BCM calibration
This is the log file to accompany the low current BCM calibration. This corresponds to run number 9382 in the Qweak DAQ (running in parity mode at 1 Hz fake MPS) and run number 696 for the Moller (the polarimeter, not the experiment!)
Low current beam calibration Green Monster Log
We requested slow orbit lock off, since it may not behave well at some of the low beam currents.
Before and after each current setting we will do a Faraday cup measurement. The list below shows the
Faraday cup reading before and after.
Interleaved between the Faraday cup readings will be the Green Monster Setpoint corresponding to that reading:
Partway through, I remembered that I should insert a beam off period (while they had the Faraday cup in) between
each setting.
5.13 uA
1
5.12
3.05
2
3.05
1.06
3
1.06
0.92
4
0.92
0.70
5
0.699
6 = beam off
0.490
7
0.493
8 = beam off
0.320
9
0.321
10=off
0.186
11
0.185
12=off
0.089
13
0.090
beam blocked: -0.0002 on Faraday cup