Main INDEX, Monthly INDEX, ,
Make New Entry, Make Followup Entry (Advanced)

User name leacock/duvall

Log entry time 16:25:32 on November 18,2011

Entry number 238897

Followups:

keyword=dslumi checkout OK, dslumi3 not making it to epics...

All dslumi event mode show reasonable analog PMT pulses. All except dslumi3 show counts on epics_scaler2.adl (the medm gui) at 950 V readback value.

We traced the dslumi3 to be a bad channel in the scaler readout. It is being discriminated properly right before the scaler.
dslumi1 20
dslumi2 1
dslumi3 0
dslumi4 35
dslumi5 5
dslumi6 50
dslumi7 1
dslumi8 40