Difference between revisions of "Beam off: Compton shut down and restart"

From PolWiki
Jump to navigationJump to search
 
(3 intermediate revisions by the same user not shown)
Line 1: Line 1:
'''Beam off: Compton shut down and restart'''<br>
+
In general, the procedures below should be done by a Compton expert. If you can't reach one, follow the instructions below (with caution).
<br>
 
  
In general, the procdures below should be done by a Compton expert. If you can't reach one, follow the instructions below (with caution).
+
== For extended Beam Off period ==
 +
During a "long" beam off, the Compton should be put in a safe state
 +
# Retract the electron detector (GARAGE)
 +
# Turn off the photon detector
 +
# Leave the laser table scintillators ON
  
During a "long" beam off, the Compton should be put in a safe state:<br>
+
'''NOTE''': If a "long" beam off is, for example, a beam studies day; It is also best to "safe" the Compton if MCC indicates they will be doing significant beam steering or tuning
1. Retract the electron detector (GARAGE).<br>
 
2. Turn off the photon detector.<br>
 
3. Leave the laser table scintillators ON.<br>
 
  
NOTE: In this case, a "long" beam off is, for example, a beam studies day. It is also best to "safe" the Compton if MCC indicates they will be doing significant beam steering or tuning.<br>
+
==Restoring the Compton for running==
 
+
# Leave the electron detector in the GARAGE
 
+
# Leave the photon detector OFF
Restoring the Compton for running:<br>
+
# When beam at the nominal current (typically > 50 uA) has been restored, observe cComptScintRateNorm
1. Leave the electron detector in the GARAGE.<br>
+
##If it is below 50 Hz/uA, it is safe to turn on the photon detector and insert the electron detector
2. Leave the photon detector OFF.<br>
+
##If it is below 100 Hz/uA, but above 75 Hz/uA, do '''NOT''' turn on the photon detector. It is probably still ok to insert the electron detector
3. When beam at the nominal current (typically > 150 uA) has been restored, observe cComptScintRateNorm.<br>
+
##If the cComptScintRateNorm is greater than 500 Hz/uA, something is very wrong. DO NOT turn on the photon detector. DO NOT insert the electron detector. Inform MCC and/or the Run Coordinator to let them know the backgrounds are much too large
--> If it is below 150 Hz/uA, it is safe to turn on the photon detector and insert the electron detector.<br>
+
# If either the photon detector is on, or the electron detector is in, or both, ask MCC to turn on the '''Compton Vertical Orbit Lock'''. To check the status of the orbit lock:
--> If it is below 300 Hz/uA, but above 200 Hz/uA, do NOT turn on the photon detector. It is probably still ok to insert the electron detector.<br>
+
##from edmmonticello screen, get to General Tools.
-->If the cComptScintRateNorm is greater than 500 Hz/uA, something is very wrong. DO NOT turn on the photon detector. DO NOT insert the electron detector. Consult with MCC and/or the Run Coordinator to let them know the backgrounds are much too large.<br>
+
##in the lower left region (below Faraday Cup settings) select "Locks". This should show new options of choosing between Energy Lock, Current Locks and FSD reset (on the left most part of screen)
4. If the either the photon detector is on, or the electron detector is in, or both, ask MCC to turn on the Compton position lock. Lately it is about 0.92 mm on IPM3P02B.YPOS.<br>
+
##choose 'Current Locks', this opens another window with several tabs showing the current locks
5. If you don't see any collisions, it may be time for a vertical scan. If you're an expert - do it. If you're not, call an expert.<br>
+
##select the 4th Tab 'Other Locks'
 +
##the left bottom of this page shows the 'Hall C Compton Vertical Orbit Lock'
 +
##under the above heading, you should see "Lock On" and next to it "Enabled". If you see the status as 'Disabled' even if the 'Lock On' message is green; the lock is actually '''not''' On. Call MCC to get it turned on. If in any confusion, call an expert.
 +
# If you don't see any collisions, it may be time for a vertical scan. If you're an expert - do it. If you're not, call an expert

Latest revision as of 09:38, 21 March 2012

In general, the procedures below should be done by a Compton expert. If you can't reach one, follow the instructions below (with caution).

For extended Beam Off period

During a "long" beam off, the Compton should be put in a safe state

  1. Retract the electron detector (GARAGE)
  2. Turn off the photon detector
  3. Leave the laser table scintillators ON

NOTE: If a "long" beam off is, for example, a beam studies day; It is also best to "safe" the Compton if MCC indicates they will be doing significant beam steering or tuning

Restoring the Compton for running

  1. Leave the electron detector in the GARAGE
  2. Leave the photon detector OFF
  3. When beam at the nominal current (typically > 50 uA) has been restored, observe cComptScintRateNorm
    1. If it is below 50 Hz/uA, it is safe to turn on the photon detector and insert the electron detector
    2. If it is below 100 Hz/uA, but above 75 Hz/uA, do NOT turn on the photon detector. It is probably still ok to insert the electron detector
    3. If the cComptScintRateNorm is greater than 500 Hz/uA, something is very wrong. DO NOT turn on the photon detector. DO NOT insert the electron detector. Inform MCC and/or the Run Coordinator to let them know the backgrounds are much too large
  4. If either the photon detector is on, or the electron detector is in, or both, ask MCC to turn on the Compton Vertical Orbit Lock. To check the status of the orbit lock:
    1. from edmmonticello screen, get to General Tools.
    2. in the lower left region (below Faraday Cup settings) select "Locks". This should show new options of choosing between Energy Lock, Current Locks and FSD reset (on the left most part of screen)
    3. choose 'Current Locks', this opens another window with several tabs showing the current locks
    4. select the 4th Tab 'Other Locks'
    5. the left bottom of this page shows the 'Hall C Compton Vertical Orbit Lock'
    6. under the above heading, you should see "Lock On" and next to it "Enabled". If you see the status as 'Disabled' even if the 'Lock On' message is green; the lock is actually not On. Call MCC to get it turned on. If in any confusion, call an expert.
  5. If you don't see any collisions, it may be time for a vertical scan. If you're an expert - do it. If you're not, call an expert