==Moller measurement summaries for A1n/d2n==
 
==Moller measurement summaries for A1n/d2n==
 
This is the "Cliff notes section on Moller polarimetry runs during the Polarized Helium-3 Running.
 
This is the "Cliff notes section on Moller polarimetry runs during the Polarized Helium-3 Running.
 +
 +
Note on sign of polarization [https://hallcweb.jlab.org/polwiki/index.php/Sign_of_the_beam_polarization]
 +
 +
Hall A spin dance Jan. 07 th, 2020 [https://logbooks.jlab.org/entry/3759691]
  
 
<b>December 12, 2019</b>
 
<b>December 12, 2019</b>
Line 11: Line 15:
 
Determined the optimum quad settings, set the detector HV and performed the collimator scan pretty quickly.  Before the collimator scan the measured polarization was 80%, after the scan it was 84.5% Unpolarized background?   
 
Determined the optimum quad settings, set the detector HV and performed the collimator scan pretty quickly.  Before the collimator scan the measured polarization was 80%, after the scan it was 84.5% Unpolarized background?   
 
Summary:[https://logbooks.jlab.org/entry/3756521], Result: [https://logbooks.jlab.org/entry/3756510]  
 
Summary:[https://logbooks.jlab.org/entry/3756521], Result: [https://logbooks.jlab.org/entry/3756510]  
 +
Epics:[https://logbooks.jlab.org/entry/3756380]
  
 
<b>January 18, 2020</b>   
 
<b>January 18, 2020</b>   
Line 28: Line 33:
 
Some troubles delivering beam.  Tried to take the other IHWP state but ended up backing out.
 
Some troubles delivering beam.  Tried to take the other IHWP state but ended up backing out.
 
summary: [https://logbooks.jlab.org/entry/3782578], result: [https://logbooks.jlab.org/entry/3782733]
 
summary: [https://logbooks.jlab.org/entry/3782578], result: [https://logbooks.jlab.org/entry/3782733]
 +
 +
<b>FLIP RIGHT (88.0)  to FLIP LEFT (-90.6) February 17, 2020  ~10am </b>
  
 
<b>February 19, 2020</b>
 
<b>February 19, 2020</b>
Line 86: Line 93:
 
  1377 -85.18   1.60         -29.64 right      10.3857   Production IHWP IN
 
  1377 -85.18   1.60         -29.64 right      10.3857   Production IHWP IN
 
  1378 -85.74   0.96         -29.64 right      10.3857   Production IHWP IN
 
  1378 -85.74   0.96         -29.64 right      10.3857   Production IHWP IN
 +
</pre>
 +
 +
 +
==Table of Bleed Thru Runs==
 +
 +
<pre>
 +
Date    Run    Comment
 +
1/18    1373    40 Hz Coin (5 kHz at 0.8 uA)
 +
</pre>
 +
 +
==Table of Tune Check Runs==
 +
 +
<pre>
 +
Date    Run    Comment
 +
1/18    1364    1294.6
 +
1/18    1365    1289.6
 +
1/18    1366    1284.6
 +
</pre>

Latest revision as of 13:02, 17 August 2021

Moller measurement summaries for A1n/d2n

This is the "Cliff notes section on Moller polarimetry runs during the Polarized Helium-3 Running.

Note on sign of polarization [1]

Hall A spin dance Jan. 07 th, 2020 [2]

December 12, 2019

Five pass commissioning. Used preliminary <azz>. Randoms were higher than expected. Need more shielding?. result: [3]

December 19, 2019

Determined the optimum quad settings, set the detector HV and performed the collimator scan pretty quickly. Before the collimator scan the measured polarization was 80%, after the scan it was 84.5% Unpolarized background? Summary:[4], Result: [5] Epics:[6]

January 18, 2020

Before this measurement, timed in randoms so that right arm comes exactly 16 ns after left. [7]. During measurement, replaced 8 ns with 6 ns delay on left arm. Reduced left/right HV by 25 V (1245/1225). Ran with collimator 5 in - this reduced backgrounds a little. Q2/Q3 current changed to 1284.604 A. Bleedthrough check, ~ 10 nA. [8] Settings: [9], BPMs: [10] summary: [11], result: [12]

January 27, 2020

Moller tune check and collimator settings[13].The beam current was 1 uA and IHWP IN. During the first three runs (1385-1387), Hall A was changing their slit and attenuator. During the last 3 runs(1388-1390),had the same slit settings. BPMs:[14], summary [15], result: [16]

February 3, 2020

Spin Dance, summary: [17], result: [18]

February 12, 2020

Some troubles delivering beam. Tried to take the other IHWP state but ended up backing out. summary: [19], result: [20]

FLIP RIGHT (88.0) to FLIP LEFT (-90.6) February 17, 2020 ~10am

February 19, 2020

Energy change in injector starting Feb 18th: [21] Measurements corrected for Hall A bleedthru (57 Hz) summary: [22], result: [23]

February 27, 2020

Changed HVs: [24] Noisy BPMS/readbacks: [25] , Large coin rates with no target, no beam, no collimators. Bleedthru/Scraping notes: [26] summary: [27], result: [28]

March 6th, 2020

Collimator 6 and 7 were scanned. Found 6 was impinging on coin rate. In and Out measured. Hall A Bleedthru (33Hz/4700Hz) summary: [29], results:[30]

March 12th, 2020 Checked collimator 6 (2.95 vs 2.65). summary: [31], result: [32]

August 3rd, 2020 Gain matched L/R. High rates with no target and no collimators to start but operator tuned beam and the rates came down. summary: [33], result: [34]

August 12th, 2020 Day shift operator had a lot of trouble tuning. Then no beam for several hours (storms/RF), came back at 7pm. summary: [35], result: [36]

August 25th, 2020

September 3rd, 2020 Measurement completed in 1/2 shift. :) Bleedthru@ 0.78% Summary:[37], result: [38]

September 11th, 2020

September 16th, 2020

Table of good Moller runs

This will be a log of all "good" Moller runs; ie, our "production" runs. It will include some detailed notes when necessary.

Date	  Run	 Pol (%)   Error (%)	Hwien	Flip       Energy(MeV)     Comment
01/18/20
	  1366	  	   	        -29.64	right      10.3857	   Detector Tune Check
 	  1368	  86.31	   0.98	        -29.64	right      10.3857	   Production IHWP OUT
 	  1369	  85.23	   0.98	        -29.64	right      10.3857	   Production IHWP OUT
	  1370	  86.53	   1.49	        -29.64	right      10.3857	   Production IHWP OUT
	  1371	 -86.05	   0.90	        -29.64	right      10.3857	   Production IHWP IN   
	  1372	 -85.69	   0.99	        -29.64	right      10.3857	   Production IHWP IN
	  1374	 -86.44	   1.06	        -29.64	right      10.3857	   Production IHWP IN
	  1375	 -84.96	   1.79	        -29.64	right      10.3857	   Production IHWP IN
	  1376	 -83.91	   1.55	        -29.64	right      10.3857	   Production IHWP IN
	  1377	 -85.18	   1.60	        -29.64	right      10.3857	   Production IHWP IN
	  1378	 -85.74	   0.96	        -29.64	right      10.3857	   Production IHWP IN


Table of Bleed Thru Runs

Date     Run     Comment
1/18     1373     40 Hz Coin (5 kHz at 0.8 uA)

Table of Tune Check Runs

Date     Run     Comment
1/18     1364     1294.6 
1/18     1365     1289.6
1/18     1366     1284.6