Difference between revisions of "Moller Polarimeter"
Line 10: | Line 10: | ||
− | The following is a plot showing a good tune of the apparatus. It was taken | + | The following is a plot showing a good tune of the apparatus. It was taken Jan 30, 2011 from run 700, with conditions: hallc:p is 1159 MeV, with a 1um target. |
− | [[Image: | + | [[Image:Moller_plot_jan30_2011.png]] |
+ | |||
+ | '''The following image shows outdated collimator values. New ones will be posted as soon as I get over to the counting house. (JM)''' | ||
Also, a screen shot of the moller medm screen. Note the Q1 and Q2 currents, and the location of the collimators. This data may change, but should be accurate for the near term. | Also, a screen shot of the moller medm screen. Note the Q1 and Q2 currents, and the location of the collimators. This data may change, but should be accurate for the near term. |
Revision as of 13:04, 31 January 2011
General information
This article will cover all things Hall-C Moller. currently, this includes all of the Qweak information. Information for past experiments can be found in the experiments section. This page is currently under construction, so please come back soon for more, or update me.
Documentation and references
The following are useful references:
- A High Precision Polarimeter (jlab document JLAB-PHY-99-29)
- Loppacher's thesis
- g0 polarization (draft 3)
The following is a plot showing a good tune of the apparatus. It was taken Jan 30, 2011 from run 700, with conditions: hallc:p is 1159 MeV, with a 1um target.
The following image shows outdated collimator values. New ones will be posted as soon as I get over to the counting house. (JM)
Also, a screen shot of the moller medm screen. Note the Q1 and Q2 currents, and the location of the collimators. This data may change, but should be accurate for the near term.
Hardware configuration & changes
NB: Up until last run in December (date tbd)we used BCM 2 with gain from Kat's Faraday cup measurements.
- (Jan 30, 2011) Increased the left and right detector high voltages, as seen below. This is a significant change in the voltages since tune-up, and is evidence of pmt radiation effects.
New HV | Old HV | |
---|---|---|
Left | 1630 | 1480 |
Right | 1655 | 1520 |
Experiments
- Qweak
- Qweak Moller How-To (pdf)
- Qweak Moller How-To (tex file)