<b>December 12, 2019</b>
 
<b>December 12, 2019</b>
  
summary: [https://logbooks.jlab.org/entry/3752542]
 
  
Five pass commissioning.  Used preliminary <azz>.  Randoms were higher than expected.  Need more shielding?
+
Five pass commissioning.  Used preliminary <azz>.  Randoms were higher than expected.  Need more shielding?. result: [https://logbooks.jlab.org/entry/3752542]
  
 
<b>December 19, 2019</b>
 
<b>December 19, 2019</b>
  
 
Determined the optimum quad settings, set the detector HV and performed the collimator scan pretty quickly.  Before the collimator scan the measured polarization was 80%, after the scan it was 84.5% Unpolarized background?   
 
Determined the optimum quad settings, set the detector HV and performed the collimator scan pretty quickly.  Before the collimator scan the measured polarization was 80%, after the scan it was 84.5% Unpolarized background?   
Result: [https://logbooks.jlab.org/entry/3756510] Summary:[https://logbooks.jlab.org/entry/3756521]
+
Summary:[https://logbooks.jlab.org/entry/3756521], Result: [https://logbooks.jlab.org/entry/3756510]  
  
 
<b>January 18, 2020</b>   
 
<b>January 18, 2020</b>   
  
Before this measurement, timed in randoms so that right arm comes exactly 16 ns after left. [https://logbooks.jlab.org/entry/3764920].  During measurement, replaced 8 ns with 6 ns delay on left arm. Reduced left/right HV by 25 V (1245/1225).  Ran with collimator 5 in - this reduced backgrounds a little. Q2/Q3 current changed to 1284.604 A. Bleedthrough check, ~ 10 nA. [https://logbooks.jlab.org/entry/3766408] Settings: [https://logbooks.jlab.org/entry/3766400]: [https://logbooks.jlab.org/entry/3766434] BPMs: [https://logbooks.jlab.org/entry/3766441] summary: [https://logbooks.jlab.org/entry/3766402]<br>
+
Before this measurement, timed in randoms so that right arm comes exactly 16 ns after left. [https://logbooks.jlab.org/entry/3764920].  During measurement, replaced 8 ns with 6 ns delay on left arm. Reduced left/right HV by 25 V (1245/1225).  Ran with collimator 5 in - this reduced backgrounds a little. Q2/Q3 current changed to 1284.604 A. Bleedthrough check, ~ 10 nA. [https://logbooks.jlab.org/entry/3766408] Settings: [https://logbooks.jlab.org/entry/3766400],  BPMs: [https://logbooks.jlab.org/entry/3766441] summary: [https://logbooks.jlab.org/entry/3766402], result: [https://logbooks.jlab.org/entry/3766434]
 +
 
 +
<br>
  
 
<b>January 27, 2020</b>  
 
<b>January 27, 2020</b>  

Revision as of 14:15, 16 August 2020

Moller measurement summaries for A1n/d2n

This is the "Cliff notes section on Moller polarimetry runs during the Polarized Helium-3 Running.

December 12, 2019


Five pass commissioning. Used preliminary <azz>. Randoms were higher than expected. Need more shielding?. result: [1]

December 19, 2019

Determined the optimum quad settings, set the detector HV and performed the collimator scan pretty quickly. Before the collimator scan the measured polarization was 80%, after the scan it was 84.5% Unpolarized background? Summary:[2], Result: [3]

January 18, 2020

Before this measurement, timed in randoms so that right arm comes exactly 16 ns after left. [4]. During measurement, replaced 8 ns with 6 ns delay on left arm. Reduced left/right HV by 25 V (1245/1225). Ran with collimator 5 in - this reduced backgrounds a little. Q2/Q3 current changed to 1284.604 A. Bleedthrough check, ~ 10 nA. [5] Settings: [6], BPMs: [7] summary: [8], result: [9]


January 27, 2020

Moller tune check and collimator settings[10].The beam current was 1 uA and IHWP IN. During the first three runs (1385-1387), Hall A was changing their slit and attenuator. During the last 3 runs(1388-1390),had the same slit settings. BPMs:[11], summary [12], result: [13]

February 3, 2020

Spin Dance, summary: [14], result: [15]

February 12, 2020

Some troubles delivering beam. Tried to take the other IHWP state but ended up backing out. summary: [16], result: [17]

February 19, 2020

Energy change in injector starting Feb 18th: [18] Measurements corrected for Hall A bleedthru (57 Hz) summary: [19], result: [20]

February 27, 2020

Changed HVs: [21] Noisy BPMS/readbacks: [22] , Large coin rates with no target, no beam, no collimators. Bleedthru/Scraping notes: [23] summary: [24], result: [25]

March 6th, 2020

Collimator 6 and 7 were scanned. Found 6 was impinging on coin rate. In and Out measured. Hall A Bleedthru (33Hz/4700Hz) summary: [26], results:[27]

March 12th, 2020

summary: [28]

August 3rd, 2020

summary: [29]

August 12th, 2020

summary: [30]

Table of good Moller runs

This will be a log of all "good" Moller runs; ie, our "production" runs. It will include some detailed notes when necessary.

Date	  Run	 Pol (%)   Error (%)	Hwien	Flip       Energy(MeV)     Comment
01/18/20
	  1366	  	   	        -29.64	right      10.3857	   Detector Tune Check
 	  1368	  86.31	   0.98	        -29.64	right      10.3857	   Production IHWP OUT
 	  1369	  85.23	   0.98	        -29.64	right      10.3857	   Production IHWP OUT
	  1370	  86.53	   1.49	        -29.64	right      10.3857	   Production IHWP OUT
	  1371	 -86.05	   0.90	        -29.64	right      10.3857	   Production IHWP IN   
	  1372	 -85.69	   0.99	        -29.64	right      10.3857	   Production IHWP IN
	  1374	 -86.44	   1.06	        -29.64	right      10.3857	   Production IHWP IN
	  1375	 -84.96	   1.79	        -29.64	right      10.3857	   Production IHWP IN
	  1376	 -83.91	   1.55	        -29.64	right      10.3857	   Production IHWP IN
	  1377	 -85.18	   1.60	        -29.64	right      10.3857	   Production IHWP IN
	  1378	 -85.74	   0.96	        -29.64	right      10.3857	   Production IHWP IN