Difference between revisions of "Moller Polarimeter"

From PolWiki
Jump to navigationJump to search
Line 1: Line 1:
 
== General information ==
 
== General information ==
This article will cover all things Hall-C Moller. currently, this includes all of the Qweak information. Information for past experiments can be found in the experiments section. This page is currently under construction, so please come back soon for more, or update me.
+
This article will cover all things Hall-C Moller. Currently, this includes all of the Qweak information. Information for past experiments can be found in the experiments section. This page is currently under construction, so please come back soon for more, or update me.
  
 
== Documentation and references ==
 
== Documentation and references ==
Line 14: Line 14:
 
[[Image:Moller_plot_jan30_2011.png]]
 
[[Image:Moller_plot_jan30_2011.png]]
  
'''The following image shows outdated collimator values. New ones will be posted as soon as I get over to the counting house. (JM)'''
+
As of Jan 30, 2011 the following quad current and collimator positions are as follows:
  
Also, a screen shot of the moller medm screen. Note the Q1 and Q2 currents, and the location of the collimators. This data may change, but should be accurate for the near term.
+
<pre>
 
+
Quad 1: 93 A
[[Image:Moller_setup_Nov_14_2010.png|600px]]
+
Quad 3: 129 A
 +
Col # Position (cm)
 +
1 -1.55
 +
2 -1.55
 +
3 -4.05
 +
4 -4.15
 +
5 Out of beam
 +
6 -3.20
 +
7 -3.20
 +
</pre>
  
 
==Hardware configuration & changes==
 
==Hardware configuration & changes==

Revision as of 15:38, 31 January 2011

General information

This article will cover all things Hall-C Moller. Currently, this includes all of the Qweak information. Information for past experiments can be found in the experiments section. This page is currently under construction, so please come back soon for more, or update me.

Documentation and references

The following are useful references:


The following is a plot showing a good tune of the apparatus. It was taken Jan 30, 2011 from run 700, with conditions: hallc:p is 1159 MeV, with a 1um target.

Moller plot jan30 2011.png

As of Jan 30, 2011 the following quad current and collimator positions are as follows:

Quad 1: 93 A
Quad 3: 129 A
Col #		Position (cm)
1		-1.55
2		-1.55
3		-4.05
4		-4.15
5		Out of beam
6		-3.20
7		-3.20

Hardware configuration & changes

NB: Up until last run in December (date tbd)we used BCM 2 with gain from Kat's Faraday cup measurements.

  • (Jan 30, 2011) Increased the left and right detector high voltages, as seen below. This is a significant change in the voltages since tune-up, and is evidence of pmt radiation effects.
New HV Old HV
Left 1630 1480
Right 1655 1520



Experiments