==Moller measurement summaries for A1n/d2n==
 
==Moller measurement summaries for A1n/d2n==
 
This is the "Cliff notes section on Moller polarimetry runs during the Polarized Helium-3 Running.
 
This is the "Cliff notes section on Moller polarimetry runs during the Polarized Helium-3 Running.
 +
 +
 +
'''January 27, 2020'''
  
 
<b>January 18, 2020</b> Before this measurement, timed in randoms so that right arm comes exactly 16 ns after left. [https://logbooks.jlab.org/entry/3764920].  During measurement, replaced 8 ns with 6 ns delay on left arm. Reduced left/right HV by 25 V (1245/1225).  Ran with collimator 5 in - this reduced backgrounds a little. Q2/Q3 current changed to 1284.604 A. Bleedthrough check, ~ 10 nA. [https://logbooks.jlab.org/entry/3766408] Settings: [https://logbooks.jlab.org/entry/3766400] Summary: [https://logbooks.jlab.org/entry/3766402] Measurements: [https://logbooks.jlab.org/entry/3766434] BPMs: [https://logbooks.jlab.org/entry/3766441]<br>
 
<b>January 18, 2020</b> Before this measurement, timed in randoms so that right arm comes exactly 16 ns after left. [https://logbooks.jlab.org/entry/3764920].  During measurement, replaced 8 ns with 6 ns delay on left arm. Reduced left/right HV by 25 V (1245/1225).  Ran with collimator 5 in - this reduced backgrounds a little. Q2/Q3 current changed to 1284.604 A. Bleedthrough check, ~ 10 nA. [https://logbooks.jlab.org/entry/3766408] Settings: [https://logbooks.jlab.org/entry/3766400] Summary: [https://logbooks.jlab.org/entry/3766402] Measurements: [https://logbooks.jlab.org/entry/3766434] BPMs: [https://logbooks.jlab.org/entry/3766441]<br>

Revision as of 10:34, 30 January 2020

Moller measurement summaries for A1n/d2n

This is the "Cliff notes section on Moller polarimetry runs during the Polarized Helium-3 Running.


January 27, 2020

January 18, 2020 Before this measurement, timed in randoms so that right arm comes exactly 16 ns after left. [1]. During measurement, replaced 8 ns with 6 ns delay on left arm. Reduced left/right HV by 25 V (1245/1225). Ran with collimator 5 in - this reduced backgrounds a little. Q2/Q3 current changed to 1284.604 A. Bleedthrough check, ~ 10 nA. [2] Settings: [3] Summary: [4] Measurements: [5] BPMs: [6]

December 19, 2019

Table of good Moller runs

This will be a log of all "good" Moller runs; ie, our "production" runs. It will include some detailed notes when necessary.

Date	  Run	 Pol (%)   Error (%)	Hwien	Flip       Energy(MeV)     Comment
01/18/20
	  1378	 -85.76	   0.96	        -29.64	right      10.3857	   Production IHWP OUT