Difference between revisions of "Shift crew procedures"
From PolWiki
Jump to navigationJump to search (Formatting wiki-fied.) |
|||
Line 1: | Line 1: | ||
− | + | === Compton Procedures Sheet (short version for shift workers) === | |
− | + | ||
− | General Compton running: | + | General Compton running: |
− | + | # At least once every hour restart a new CODA run. CODA should be open on one of the desktops of ccomptl1. | |
− | + | # At least once every hour check the status of the HV. If a trip occurred, see below. | |
− | + | # Check whether the laser is still flipping (EPICS channel cComptCavityPower). | |
− | + | #:This should be obvious from the camera as well. If not flipping, call an expert. This EPICS PV should be visible on the desktop of Striptools on ccomptl1. | |
− | + | # Monitor the background scintillator rates. cComptScintRateNorm should stay below 200 Hz/uA. | |
− | + | ||
− | + | ||
After a photon detector HV trip:<br> | After a photon detector HV trip:<br> | ||
− | + | # Check that the normalized rate on the background scintillator (EPICS channel cComptScintRateNorm, or manually normalize hcrate14) is still below 150 Hz/uA (200 Hz/uA is tolerable, <150 Hz/uA is "good"). | |
− | + | #: This EPICS PV should be visible on the desktop of Striptools on ccomptl1. If not, call an expert.<br> | |
− | + | # Wait 5 minutes after the trip and turn on the HV on the photon detector to 1600 V (channel PhotonDet_6).<br> | |
− | + | #: The HV controls should be open on one of the ccomptl1 desktops. If not log on to cdaql4 as cvxwrks and cd to $EPSHV/tk then run /HV_compton.tcl and turn on PhotonDet_6.<br> | |
− | + | # If the photon detector HV trips again within 15 minutes, call an expert.<br> | |
− | + | ||
− | + | ||
− | + | === Experts on call === | |
− | + | ||
− | Experts: | + | Experts: |
− | Dave Gaskell, pager gaskelld-page@jlab.org, cell 719-5482 | + | * Dave Gaskell, pager gaskelld-page@jlab.org, cell 719-5482 |
− | Don Jones, cell 768-1148 | + | * Don Jones, cell 768-1148 |
− | Amrendra Narayan, cell 952-6126 | + | * Amrendra Narayan, cell 952-6126 |
− | Wouter Deconinck, pager 238-4235, cell 927-2413 | + | * Wouter Deconinck, pager 238-4235, cell 927-2413 |
− | + | ||
− | Schedule: | + | Schedule: |
− | January 12 - January 19: Don Jones | + | * January 12 - January 19: Don Jones |
− | January 20 - January 26: Dave Gaskell | + | * January 20 - January 26: Dave Gaskell |
− | January 27 - February 3: Amrendra Narayan | + | * January 27 - February 3: Amrendra Narayan |
Revision as of 19:29, 21 January 2011
Compton Procedures Sheet (short version for shift workers)
General Compton running:
- At least once every hour restart a new CODA run. CODA should be open on one of the desktops of ccomptl1.
- At least once every hour check the status of the HV. If a trip occurred, see below.
- Check whether the laser is still flipping (EPICS channel cComptCavityPower).
- This should be obvious from the camera as well. If not flipping, call an expert. This EPICS PV should be visible on the desktop of Striptools on ccomptl1.
- Monitor the background scintillator rates. cComptScintRateNorm should stay below 200 Hz/uA.
After a photon detector HV trip:
- Check that the normalized rate on the background scintillator (EPICS channel cComptScintRateNorm, or manually normalize hcrate14) is still below 150 Hz/uA (200 Hz/uA is tolerable, <150 Hz/uA is "good").
- This EPICS PV should be visible on the desktop of Striptools on ccomptl1. If not, call an expert.
- This EPICS PV should be visible on the desktop of Striptools on ccomptl1. If not, call an expert.
- Wait 5 minutes after the trip and turn on the HV on the photon detector to 1600 V (channel PhotonDet_6).
- The HV controls should be open on one of the ccomptl1 desktops. If not log on to cdaql4 as cvxwrks and cd to $EPSHV/tk then run /HV_compton.tcl and turn on PhotonDet_6.
- The HV controls should be open on one of the ccomptl1 desktops. If not log on to cdaql4 as cvxwrks and cd to $EPSHV/tk then run /HV_compton.tcl and turn on PhotonDet_6.
- If the photon detector HV trips again within 15 minutes, call an expert.
Experts on call
Experts:
- Dave Gaskell, pager gaskelld-page@jlab.org, cell 719-5482
- Don Jones, cell 768-1148
- Amrendra Narayan, cell 952-6126
- Wouter Deconinck, pager 238-4235, cell 927-2413
Schedule:
- January 12 - January 19: Don Jones
- January 20 - January 26: Dave Gaskell
- January 27 - February 3: Amrendra Narayan