! 64      || Aug 14th 11:28:27  || N1 && N14 ||    10mV      ||  15mV  ||  Aug 15th 09:30:34  ||  12698 ||  [https://hallcweb.jlab.org/wiki/index.php/Image:HV_preshower_N_run63.png Run 63 Preshower HV ] [https://hallcweb.jlab.org/wiki/index.php/Image:HV_scint_run52.png Run 52 Scint HV]      || TDC range 1024ns (250ps/ch) no change looking for a bug
 
! 64      || Aug 14th 11:28:27  || N1 && N14 ||    10mV      ||  15mV  ||  Aug 15th 09:30:34  ||  12698 ||  [https://hallcweb.jlab.org/wiki/index.php/Image:HV_preshower_N_run63.png Run 63 Preshower HV ] [https://hallcweb.jlab.org/wiki/index.php/Image:HV_scint_run52.png Run 52 Scint HV]      || TDC range 1024ns (250ps/ch) no change looking for a bug
 
|-
 
|-
! 65      || Aug 15th 09:35:43 || N1 && N2 ||    10mV      ||  15mV  ||  Aug 15th   ||   ||  [https://hallcweb.jlab.org/wiki/index.php/Image:HV_preshower_N_run65.png Run 65 Preshower HV ] [https://hallcweb.jlab.org/wiki/index.php/Image:HV_scint_run52.png Run 52 Scint HV]      || TDC range 1024ns (250ps/ch) changed votage on PMT 1 to 1650 and changed trigger to N1 and N2
+
! 65      || Aug 15th 09:35:43 || N1 && N2 ||    10mV      ||  15mV  ||  Aug 15th 13:28:56  || 112919  ||  [https://hallcweb.jlab.org/wiki/index.php/Image:HV_preshower_N_run65.png Run 65 Preshower HV ] [https://hallcweb.jlab.org/wiki/index.php/Image:HV_scint_run52.png Run 52 Scint HV]      || TDC range 1024ns (250ps/ch) changed votage on PMT 1 to 1650 and changed trigger to N1 and N2
 
|}
 
|}

Revision as of 12:36, 15 August 2012

Status

  • The SHMS Calorimeter has been cabled with signal and HV.
  • Have checked setting each channel to 1400V that signal is seen.
  • Altered voltages within SHMS Calorimeter to match around 50 mV per tube.
  • Added external PMT's across top of calorimeter, 6 in total,
  • 06/28 Taking data for number of events through calorimeter and external PMT's
  • 07/10 Added diagram of Modules and Live connections
  • 07/18 Added plots of SHMS Cal. data

Computer Setup

  • Computer is called cnu. The account is gep5.

HV Setup

  • Can control the CAEN 403 through terminal on cnu.
    • Open terminal
    • type "su" to login as superuser.
    • type "minicom"
  • Listing of the HV setting and cable numbering for the preshower and scintillators
  • Preshower HV for Neg and for Positive side on July 30th at 4:41

Event Data

  • Coincidence data between SHMS Cal and external PMTs Hits per hour
  • Diagram of modules used for collecting coincidence data Module Diagram
  • Voltage versus PMT# plots for the negative and positive sides of the calorimeter. plot

Data taking Log

Run number Start Date/Time Trigger PreShower Threshold Scintillator Threshold End Date/Time Events HV figure Comments
34 July 30th 17:53:19 P1 && P14 20mV N/A July 31st 09:29:09 368 Pos HV] TDC range 1024ns (250ps/ch)
35 July 31st 09:37:27 P1 && P2 20mV N/A July 31st 10:29:10 2178 Pos HV] TDC range 1024ns (250ps/ch)
36 July 31st 14:29:14 P1 && P2 20mV N/A July 31st 14:35:00 144 Pos HV] TDC range 1024ns (250ps/ch power outage ending run)
37 July 31st 14:38:13 P1 && P2 20mV N/A July 31st 14:56:13 738 Pos HV] TDC range 1024ns (250ps/ch)
38 July 31st 15:26:35 P1 && P5 20mV N/A July 31st 15:47:02 119 Pos HV] TDC range 280ns (70ps/ch) power outage ending run
40 July 31st 19:43:43 P1 && P5 20mV N/A Aug 1st 10:24:51 6367 Pos HV] TDC range 280ns (70ps/ch)
41 Aug 1st 10:33:25 P1 && P2 20mV 50mV Aug 1st 11:04:16 1240 Pos HV Scint HV TDC range 280ns (70ps/ch) Added scintillators to the TDC with threshold = 50mV
42 Aug 1st 11:07:31 P1 && P2 20mV 50mV Aug 2nd 9:38:29 55941 Pos HV Scint HV TDC range 1024ns (250ps/ch)
43 Aug 2nd 9:52:34 P1 && P2 15mV 40mV Aug 2nd 11:46:39 9183 Pos HV] Scint HV TDC range 1024ns (250ps/ch)
44 Aug 2nd 11:55:03 P1 && P2 10mV 30mV Aug 2nd 15:02:54 28155 Pos HV] Scint HV TDC range 1024ns (250ps/ch)
45 Aug 2nd 15:45:12 P1 && P2 10mV 30mV Aug 2nd 17:26:06 26089 Updated Pos Preshower HV Scint HV TDC range 1024ns (250ps/ch) raised HV for bars 1,2,11 & 14
46 Aug 2nd 17:48:46 P1 && P6 10mV 15mV Aug 3rd 11:09:04 33524 Run 45 Preshower HV Scint HV TDC range 1024ns (250ps/ch)
47 Aug 3rd 11:10:29 P1 && P14 10mV 15mV Aug 3rd 14:43:08 1090 Run 45 Preshower HV Scint HV TDC range 1024ns (250ps/ch)
48 Aug 3rd 14:47:49 P2 && P14 10mV 15mV Aug 3rd 17:53:55 1243 Run 48 Preshower HV Run 48 Scint HV TDC range 1024ns (250ps/ch) Raised PSH1 HV, Scint Bot Mid HV
49 Aug 3rd 17:54:51 P2 && P14 10mV 15mV Aug 6th 10:24:59 25869 Run 49 Preshower HV Run 48 Scint HV TDC range 1024ns (250ps/ch)
50 Aug 6th 10:33:46 P2 && P14 20mV 15mV Aug 7th 09:31:34 2593 Run 49 Preshower HV Run 48 Scint HV TDC range 1024ns (250ps/ch)
51 Aug 7th 09:40:47 P2 && P14 10mV 15mV Aug 7th 12:27:43 1178 Run 51 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Raised HV on PSH1 & 2
52 Aug 7th 12:32:51 P2 && P14 10mV 15mV Aug 7th 16:33:46 1766 Run 52 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Raised HV on PSH 3-8, 10 & 11
53 Aug 7th 16:38:59 P2 && P14 10mV 15mV Aug 8th 09:50:15 7583 Run 53 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Raised HV on PSH 1,3-8,10 by 25V
54 Aug 8th 09:54:13 P2 && P14 10mV 15mV Aug 8th 13:54:02 1860 Run 54 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Raised all HV to 1600 except 2 (1550)
55 Aug 8th 14:01:04 P2 && P14 10mV 15mV Aug 8th 15:39:46 966 Run 55 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Raised all HV to 1650
56 Aug 8th 15:44:28 P2 && P14 10mV 15mV Aug 9th 09:45:36 12628 Run 56 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Raised all HV to 1700
57 Aug 9th 09:58:02 P1 && P2 10mV 15mV Aug 9th 13:21:32 99370 Run 56 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Changed trigger to p1 and p2
58 Aug 9th 13:42:16 N1 && N2 10mV 15mV Aug 9th 15:13:46 18024 Run 58 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Changed sides to N side
59 Aug 9th 15:17:31 N1 && N14 10mV 15mV Aug 10th 13:47:50 5633 Run 58 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Changed trigger to N1 && N14
60 Aug 10th 14:19:03 N1 && N14 10mV 15mV Aug 13th 10:03:31 16130 Run 60 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Changed HV on PMT# 3,5,6
61 Aug 13th 10:15:54 N1 && N14 10mV 15mV Aug 13th 13:38:57 1150 Run 61 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Changed HV on all PMT# up 50
62 Aug 13th 13:42:56 N1 && N14 10mV 15mV Aug 14th 09:33:02 9564 Run 62 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Changed HV on all PMT# to 1650
63 Aug 14th 09:43:11 N1 && N14 10mV 15mV Aug 14th 11:26:30 1047 Run 63 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) Changed HV on all PMT# to 1700
64 Aug 14th 11:28:27 N1 && N14 10mV 15mV Aug 15th 09:30:34 12698 Run 63 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) no change looking for a bug
65 Aug 15th 09:35:43 N1 && N2 10mV 15mV Aug 15th 13:28:56 112919 Run 65 Preshower HV Run 52 Scint HV TDC range 1024ns (250ps/ch) changed votage on PMT 1 to 1650 and changed trigger to N1 and N2