Difference between revisions of "1-pass calibration for A1n"
From HallCWiki
Jump to navigationJump to searchLine 58: | Line 58: | ||
− | == Polarized 3He Runs == | + | == Polarized 3He and Reference Cell Runs, Delta resonance setting == |
=== Polarized 3He Delta Transverse Asymmetry === | === Polarized 3He Delta Transverse Asymmetry === | ||
Line 96: | Line 96: | ||
- Expected asymmetries (after W and Ytg cuts): To be filled HMS Aphys=4.8% Ameas=2.1% and SHMS Aphys=2.5% Ameas=0.98% if Pt=50%, Pb=85%, N2 dilution 0.92. | - Expected asymmetries (after W and Ytg cuts): To be filled HMS Aphys=4.8% Ameas=2.1% and SHMS Aphys=2.5% Ameas=0.98% if Pt=50%, Pb=85%, N2 dilution 0.92. | ||
- Expected uncertainties better than dA/A=10% on each of HMS and SHMS. | - Expected uncertainties better than dA/A=10% on each of HMS and SHMS. | ||
+ | |||
+ | == Polarized 3He and Reference Cell Runs, elastic setting == | ||
+ | |||
+ | === Reference Cell Hydrogen (H2) Cross Section | ||
+ | ''' Goals ''' | ||
+ | - To measure hydrogen elastic cross section as a cross check of data quality | ||
+ | |||
+ | '''Setup''' | ||
+ | - Beam Current: up to 30uA and DAQ rate-limited, raster always ON (5mm diameter) | ||
+ | - HMS setting: 12.5 deg, -2.1435 GeV/c (electron) | ||
+ | - SHMS setting: 12.5 deg, -2.1435 GeV/c (electron) | ||
+ | - Target: Polarized 3He Cell or Reference Cell | ||
+ | - Detectors: All SC on to start with | ||
+ | |||
+ | |||
+ | === 3He Elastic Scattering Setup === | ||
+ | ''' Goals ''' | ||
+ | - To determine which two SC bars should be turned ON to focus on elastic (suppress quasielastic) | ||
+ | - Must be done Prior to Polarized 3He Longitudinal Asymmetry and all reference cell runs at the elastic setting | ||
+ | |||
+ | '''Setup''' | ||
+ | - Beam Current: 1-2uA or DAQ rate-limited, raster always ON (5mm diameter) | ||
+ | - HMS setting: 12.5 deg, -2.1435 GeV/c (electron) | ||
+ | - SHMS setting: 12.5 deg, -2.1435 GeV/c (electron) | ||
+ | - Target: Polarized 3He Cell or Reference Cell | ||
+ | - Detectors: All SC on to start with | ||
+ | |||
+ | '''Procedure''' | ||
+ | - Set spectrometers at the required settings, target at the Polarized 3He Cell position; | ||
+ | - Take a short run at 1 uA, check focal plane event distribution and determine which two scintillator bars should be kept on to accept the whole 3He elastic peak and to suppress quasielastic and inelastic events. (expert driven). | ||
+ | |||
+ | |||
=== 3He Reference Cell Runs === | === 3He Reference Cell Runs === | ||
Line 122: | Line 154: | ||
- Once scintillators are set, take another short run at 1uA to determine rate and prescale factor. Adjust beam current to reach close to 4.5kHz with PS=1 (but do not exceed 30uA), which means current is probably limited by SHMS rate. | - Once scintillators are set, take another short run at 1uA to determine rate and prescale factor. Adjust beam current to reach close to 4.5kHz with PS=1 (but do not exceed 30uA), which means current is probably limited by SHMS rate. | ||
- Take a five-minute run or until reaching 10,000 events for both HMS and SHMS | - Take a five-minute run or until reaching 10,000 events for both HMS and SHMS | ||
− | |||
=== Reference Cell Hydrogen (H2) Cross Section === | === Reference Cell Hydrogen (H2) Cross Section === |
Revision as of 13:33, 29 November 2019
Outline
For polarized 3He running in Fall 2019 and Spring 2020, the nominal raster for production is 5mm diameter circular. The nominal triggers used are 3/4 for both HMS and SHMS. Both HMS and SHMS should be taking data (for example if run plan specify one run, it means one run on HMS and one run on SHMS simultaneously). No coincidence trigger will be used/setup.
We will need to do optics calibration, reference cell (cross section and pressure curves), 3He elastic longitudinal asymmetry, and 3He Delta transverse asymmetry measurements.
Beam and raster checkout
Please refer to 2-pass commissioning procedure, follow: - Establish beam in the hall - Initial Raster checks - Harp scans - Target/Beam centering, Raster calibration
Optics Calibration
Optics (angle and ytarg) Calibration
Goals - To calibration spectrometer optics for 40cm long target at 30 deg. Calibrate angle reconstruction.
Setup - Beam Current: 5-20 uA CW, raster varies - HMS setting: 30 degrees, -1.2 GeV/c (electron) - SHMS setting: 30 degrees, -1.68 GeV/c (electron) - Target: Multi(7)-foil Carbon Foils
Procedure - Move target to multi-foil position; - Move Sieve Slit IN for both HMS and SHMS; - Take a short run (2min) at 20uA, verify sieves are both IN; - Ask MCC to increase current to 20uA (Max for raster off on this target). Take two runs, 45 minutes each. - Ask MCC to turn raster ON to 5mm diameter, take one run at 20 uA for 45 minutes. - Move Sieve Slit OUT for both HMS and SHMS; - Take one run at 20 uA for 15 minutes. - Ask MCC to turn raster OFF, take one run at 20 uA for 15 minutes.
Things to watch for (Expert Analysis) - There should be optics expert(s) checking data quality as we go.
Optics (Momentum) Calibration
Goals - To calibration momentum reconstruction of spectrometers.
Setup - Beam Current: 5--20 uA CW, raster always OFF - HMS setting: 11.7 deg, momentum varies (electron) - SHMS setting: 8.5 deg, momentum varies (electron) - Target: Double Carbon Foils
Procedure - Move target to Double carbon foil position.
- To be filled
Things to watch for (Expert Analysis) - There should be optics expert(s) checking data quality as we go.
Polarized 3He and Reference Cell Runs, Delta resonance setting
Polarized 3He Delta Transverse Asymmetry
Goals - To measure physics transverse asymmetry of Delta resonance to (5-10)% relative on each spectrometer.
Setup - Beam Current: DAQ rate-limited and up to 30uA, raster always ON (5mm diameter) - Beam IHWP: should be switched half-way - HMS setting: 12.5 deg, -1.799 GeV/c (electron) - SHMS setting: 12.5 deg, -1.799 GeV/c (electron) - Target: Polarized 3He Cell, Empty Reference Cell - Target spin direction: 90 or 270-deg - Target coil setting: TBD - Detectors: All scintillator bars ON. - Parity check: Inform Hall A prior to taking data to make sure they set the charge asymmetry feedback cutoff at 0.5uA.
Procedure - Set spectrometers at the required settings, target at Reference Cell position, pump the reference cell to empty; - Check and make sure all scintillators are ON; - Take a short run at 2uA to determine rate and prescale factor. Adjust beam current to reach close to 4.5kHz with PS=1 (but do not exceed 30uA), which means current is probably limited by SHMS rate. - Take a five-minute run or until reaching 10,000 events for both HMS and SHMS
- Move target to Polarized 3He Cell position, spin polarization at either 90 or 270 degrees. - Perform one polarimetry measurement before taking data. - Take one 1-minute run at 2uA, check rates and adjust beam current and prescale factors. Adjust beam current (but do not exceed 30uA) to reach close to 4.5kHz with PS=1, which means current is probably limited by SHMS rate. - Take five 30-minute runs at the optimal current set above, perform one polarimetry measurement after each run. Verify data quality (see below) and verify target is still polarized. - Switch beam IHWP position. (may need to inform other halls in advance). For example from IN to OUT or from OUT to IN. - Take five 30-minute runs at the optimal current set above, perform one polarimetry measurement after each run. Verify data quality (see below) and verify target is still polarized.
Things to watch online - Apply y or z cuts to isolate events scattered from 3He gas (We do not want events from the glass windows!) - Apply 1.1<W<1.35 GeV cut to isolate the Delta resonance. - Run online script to ensure we see physics asymmetry. HMS and SHMS should see opposite asymmetries.
- Expected total rate: To be filled - Expected asymmetries (after W and Ytg cuts): To be filled HMS Aphys=4.8% Ameas=2.1% and SHMS Aphys=2.5% Ameas=0.98% if Pt=50%, Pb=85%, N2 dilution 0.92. - Expected uncertainties better than dA/A=10% on each of HMS and SHMS.
Polarized 3He and Reference Cell Runs, elastic setting
=== Reference Cell Hydrogen (H2) Cross Section
Goals - To measure hydrogen elastic cross section as a cross check of data quality
Setup - Beam Current: up to 30uA and DAQ rate-limited, raster always ON (5mm diameter) - HMS setting: 12.5 deg, -2.1435 GeV/c (electron) - SHMS setting: 12.5 deg, -2.1435 GeV/c (electron) - Target: Polarized 3He Cell or Reference Cell - Detectors: All SC on to start with
3He Elastic Scattering Setup
Goals - To determine which two SC bars should be turned ON to focus on elastic (suppress quasielastic) - Must be done Prior to Polarized 3He Longitudinal Asymmetry and all reference cell runs at the elastic setting Setup - Beam Current: 1-2uA or DAQ rate-limited, raster always ON (5mm diameter) - HMS setting: 12.5 deg, -2.1435 GeV/c (electron) - SHMS setting: 12.5 deg, -2.1435 GeV/c (electron) - Target: Polarized 3He Cell or Reference Cell - Detectors: All SC on to start with
Procedure - Set spectrometers at the required settings, target at the Polarized 3He Cell position; - Take a short run at 1 uA, check focal plane event distribution and determine which two scintillator bars should be kept on to accept the whole 3He elastic peak and to suppress quasielastic and inelastic events. (expert driven).
3He Reference Cell Runs
Goals - To determine window background for elastic setting - To determine which two SC bars should be on to focus on elastic (suppress quasielastic) - To determine yield vs. pressure (pressure curve) - Must be done Prior to Polarized 3He Longitudinal Asymmetry - Must be done as the
Setup - Beam Current: DAQ rate-limited and up to 30uA, raster always ON (5mm diameter) - HMS setting: 12.5 deg, -2.1435 GeV/c (electron) - SHMS setting: 12.5 deg, -2.1435 GeV/c (electron) - Target: Reference Cell - Detectors: We will need to keep only 2 SC bars ON to limit quasi-elastic scattering events. - Parity check: Inform Hall A prior to taking data to make sure they set the charge asymmetry feedback cutoff at 0.5uA.
Procedure - Set spectrometers at the required settings, target at the Reference Cell position; - Fill the reference cell with 3He gas to 8 atm;
to empty (vacuum in the cell);
- Take one 1-minute run at 1 uA, check focal plane event distribution and determine which two scintillator bars should be kept on. (expert driven). - Once scintillators are set, take another short run at 1uA to determine rate and prescale factor. Adjust beam current to reach close to 4.5kHz with PS=1 (but do not exceed 30uA), which means current is probably limited by SHMS rate. - Take a five-minute run or until reaching 10,000 events for both HMS and SHMS
Reference Cell Hydrogen (H2) Cross Section
Goals - To measure hydrogen elastic cross section, need 10,000 elastic events each for HMS and SHMS
Setup
Procedure
Reference N2 Pressure Curve
Goals - To measure yield and cross section for varying N2 pressure. This will be used to determine amount of N2 in the polarized 3He cell in the offline analysis.
Setup
Procedure
Reference 3He Pressure Curve
Goals - To measure yield and cross section for varying N2 pressure. This will be used to determine amount of 3He in the polarized 3He cell in the offline analysis. - Important: Among all reference cell runs (N2, H2 and 3He), run 3He last, and do NOT vent 3He gas after completing the run plan.
Setup
Procedure
Polarized 3He Elastic Longitudinal Asymmetry
Goals - To measure physics longitudinal asymmetry of 3He elastic scattering to (2-3)% relative, cross check PbPt from Moller and Target Polarimetry
Setup - Beam Current: DAQ rate-limited and up to 30uA, raster always ON (5mm diameter) - HMS setting: 12.5 deg, -2.1435 GeV/c (electron) - SHMS setting: 12.5 deg, -2.1435 GeV/c (electron) - Target: Polarized 3He Cell, Reference Cell - Target spin direction: 0-deg, 180-deg - Target coil settings: TBD - Detectors: We will need to keep only 2 SC bars ON to limit quasi-elastic scattering events. - Parity check: Inform Hall A prior to taking data to make sure they set the charge asymmetry feedback cutoff at 0.5uA.
Procedure - Set spectrometers at the required settings, target at the Reference Cell position; - Pump the reference cell to empty (vacuum in the cell); - Take one 1-minute run at 1 uA, check focal plane event distribution and determine which two scintillator bars should be kept on. (expert driven). - Once scintillators are set, take another short run at 1uA to determine rate and prescale factor. Adjust beam current to reach close to 4.5kHz with PS=1 (but do not exceed 30uA), which means current is probably limited by SHMS rate. - Take a five-minute run or until reaching 10,000 events for both HMS and SHMS
- Move target to Polarized 3He Cell position, spin polarization at either 0 or 180 degrees. - Perform one polarimetry measurement before taking data. - Make sure the same two SC bars are on as the Reference Cell, - Take one 1-minute run at 1 uA, check focal plane event distribution and determine which two scintillator bars should be kept on. (expert driven). - Once scintillators are set, take one 1-minute run at 2uA, check rates and adjust prescale factor. Ideally we want 4.5kHz on each of HMS/SHMS with PS=1. - Take three 30-minute (TBD) runs at the optimal current set above, perform one polarimetry measurement after each run. Verify data quality (see below) and verify target is still polarized. - Rotate target spin direction from 0 to 180 deg (or from 180 to 0 deg) - Take three 30-minute runs at the optimal current set above, perform one polarimetry measurement after each run. Verify data quality (see below) and verify target is still polarized. - Call MCC and ask beam IHWP to be inserted (or taken out). Need to inform other halls prior to the switch. - Take xxx runs ... (repeat above)
Things to watch online - Ensure the correct SC bars are on (We need elastic events! One SC bar is in principle enough but we keep TWO on) - Apply y or z cuts to isolate events scattered from 3He gas (We do not want events from the glass windows!) - Apply W cuts to isolate elastic events from others. - Run online script to ensure we see physics asymmetry. HMS and SHMS should see the same asymmetry. Asymmetry should flip sign with beam IHWP switch.
- Expected total rate: (To be updated) HMS at 1uA total 1.8KHz 3He gas elastic 240Hz; SHMS at 1uA total 27.7kHz Prescaled by 15 to 3He gas elastic 740Hz - Expected asymmetries (after W and Ytg cuts): (To be updated) HMS Aphys=4.8% Ameas=2.1% and SHMS Aphys=2.5% Ameas=0.98% if Pt=50%, Pb=85%, N2 dilution 0.92. - Expected uncertainties dA/A=2% (goal) or 3% (minimum) on each of HMS and SHMS.