Difference between revisions of "Running the Yerevan Test Stand DAQ"
From HallCWiki
Jump to navigationJump to search (Created page with "The Yerevan test stand for Aerogel detector is located in the room adjacent to Walter's office above the machine shop in the EEL building. Has a single VME crate connected t...") |
|||
Line 2: | Line 2: | ||
to Walter's office above the machine shop in the EEL building. | to Walter's office above the machine shop in the EEL building. | ||
− | + | It has a single VME crate connected to a PC called hcdaq2. | |
+ | The system runs CODA 3.10. The VME crate currently has a single CAEN V792 ADC. | ||
Login to hcdaq2 with the username "coda" and ask Bob Michaels for the password. | Login to hcdaq2 with the username "coda" and ask Bob Michaels for the password. |
Revision as of 15:13, 17 July 2022
The Yerevan test stand for Aerogel detector is located in the room adjacent to Walter's office above the machine shop in the EEL building.
It has a single VME crate connected to a PC called hcdaq2. The system runs CODA 3.10. The VME crate currently has a single CAEN V792 ADC.
Login to hcdaq2 with the username "coda" and ask Bob Michaels for the password.
To start the DAQ, on hcdaq2 type startcoda then login to the VME and start the ROC. The command "govme" logs into the VME crate and from there you type "startroc". [coda@hcdaq2 ~]$ govme [sbs-onl@amoll2 ~]$ startroc
Hopefully the control GUI rcgui is familiar. If not:
After doing startcoda and the gui pops up in a few seconds ... Click Control (upper left) Click Connect Click the Tools icon (far left of GUI, at the top) Click Download (2nd button from left) Click Start to start a run (the >> button) Click End to end a run