Difference between revisions of "Pol He-3 Target Motion Reset"

From HallCWiki
Jump to: navigation, search
 
(3 intermediate revisions by the same user not shown)
Line 1: Line 1:
 
To reset the hall C target motion, several devices need to be rebooted.
 
To reset the hall C target motion, several devices need to be rebooted.
  
'''WARNING: After iochc26 boots up, the target will home itself, so you need to mask the FSD first!!'''
+
'''WARNING: You must call MCC and mask the target before doing this procedure'''
  
First, use hcreboot5 to turn off the iochc26 VME crate (outlet 2).
+
# First, use [[hcreboot5]] to '''turn off''' the iochc26 VME crate and '''leave it off for now''' (outlet 2).
 
+
# Next, use [[hcreboot6]] to reboot
Next, use hcreboot6 to reboot portservhc26 (outlet 1) and the VXM-1 Motor controller (outlet 2). Wait a couple minutes for the portserver to boot up.
+
## portservhc26 (outlet 1) and  
 
+
## the VXM-1 Motor controller (outlet 2).  
Finally, use hcreboot5 to turn iochc26 (outlet 2 again) back on. The IOC will take a few minutes to boot up completely.
+
## Wait 60 seconds for the portserver to boot up.
 +
# Finally, use [[hcreboot5]] to turn iochc26 (outlet 2 again) back on. The IOC will take a few minutes to boot up completely. When it does, the white fields in the alarm handler and Target motion control GUI will go back to their regular colors.
 +
# Click the 'Initialize' button at the top of the Target Motion Control GUI and wait for the process to complete.
 +
# Move to your chosen target position and '''verify that the 'Encoder Readout (volts)' field matches the setting you expect listed next to the vertical target position bar!'''

Latest revision as of 09:15, 8 August 2020

To reset the hall C target motion, several devices need to be rebooted.

WARNING: You must call MCC and mask the target before doing this procedure

  1. First, use hcreboot5 to turn off the iochc26 VME crate and leave it off for now (outlet 2).
  2. Next, use hcreboot6 to reboot
    1. portservhc26 (outlet 1) and
    2. the VXM-1 Motor controller (outlet 2).
    3. Wait 60 seconds for the portserver to boot up.
  3. Finally, use hcreboot5 to turn iochc26 (outlet 2 again) back on. The IOC will take a few minutes to boot up completely. When it does, the white fields in the alarm handler and Target motion control GUI will go back to their regular colors.
  4. Click the 'Initialize' button at the top of the Target Motion Control GUI and wait for the process to complete.
  5. Move to your chosen target position and verify that the 'Encoder Readout (volts)' field matches the setting you expect listed next to the vertical target position bar!