Difference between revisions of "BPM calibration information"

From HallCWiki
Jump to navigationJump to search
Line 15: Line 15:
 
|-  
 
|-  
 
! 12/8/2019  !! Calibration !! 0.14 !! 0.22  !! -1.13 !! -0.06 !! -0.29 !! -0.78 !! -0.96 !! -1.09 !! -0.96 !! 0.95 !! 1.16 !! 0.85 !! Polarized 3He running
 
! 12/8/2019  !! Calibration !! 0.14 !! 0.22  !! -1.13 !! -0.06 !! -0.29 !! -0.78 !! -0.96 !! -1.09 !! -0.96 !! 0.95 !! 1.16 !! 0.85 !! Polarized 3He running
 +
|-
 +
! 6/9/2022  !! Harp scan !! 0.16 !! -0.22  !! -0.54 !! -0.18 !! 0.02 !! 0.39 !! - !! - !! - !! - !! - !! - !!
 
|-  
 
|-  
 
! 1/11/2023  !! Harp scan !! 0.06 !! -0.35  !! -0.69 !! -0.20 !! -0.03 !! 0.38 !! - !! - !! - !! - !! - !! - !!  
 
! 1/11/2023  !! Harp scan !! 0.06 !! -0.35  !! -0.69 !! -0.20 !! -0.03 !! 0.38 !! - !! - !! - !! - !! - !! - !!  

Revision as of 14:31, 5 May 2023

  • The beam position monitors (BPMs) on the diagnostic girder in Hall C (IPM3H07A, IPM3H07B, and IPM3H07C) are calibrated by comparison with the Hall C superharps (IHA3H07A and IHA3H07B). Ideally, a "bulls-eye" scan is doe which allows one to determine the BPMS offsets and slopes. However, a single harp scan is sufficient to check the BPM offsets.
  • If re-analyzing/revisiting BPM calibrations, make sure that the appropriate survey offsets are loaded into the harp analyzer tool
Date Method AXoff BXoff CXoff AYoff BYoff CYoff AXslope BXslope CXslope AYslope BYslope CYslope Notes
1/13/2018 Calibration -0.06 0.07 -1.06 -0.46 0.12 0.61 -0.98 -1.12 -0.96 0.97 1.17 0.87 1st calibration
8/29/2018 Harp scan 0.15 0.29 -1.08 -0.12 0.63 1.09 - - - - - - -
2/9/2019 Harp scan 0.25 0.46 -1.01 -0.08 -0.04 -0.11 - - - - - - Radiator installed during January 2019 SAD
6/23/2019 Calibration 0.11 0.17 -1.07 -0.40 -0.18 -0.40 -0.97 -1.11 -0.95 0.95 1.15 0.85 -
12/8/2019 Calibration 0.14 0.22 -1.13 -0.06 -0.29 -0.78 -0.96 -1.09 -0.96 0.95 1.16 0.85 Polarized 3He running
6/9/2022 Harp scan 0.16 -0.22 -0.54 -0.18 0.02 0.39 - - - - - -
1/11/2023 Harp scan 0.06 -0.35 -0.69 -0.20 -0.03 0.38 - - - - - -
  • Beamline surveys