Difference between revisions of "BPM calibration information"

From HallCWiki
Jump to navigationJump to search
Line 39: Line 39:
  
 
* Summary of survey information
 
* Summary of survey information
 +
 
{|border="1" align="center"
 
{|border="1" align="center"
 
! Date  !!  BPM AXoff !! BPM BXoff  !! BPM CXoff !! BPM AYoff !! BPM BYoff !! BPM CYoff !! Harp AXoff !! Harp BXoff !! Harp AYoff !! Harp BYoff !! Notes
 
! Date  !!  BPM AXoff !! BPM BXoff  !! BPM CXoff !! BPM AYoff !! BPM BYoff !! BPM CYoff !! Harp AXoff !! Harp BXoff !! Harp AYoff !! Harp BYoff !! Notes

Revision as of 07:01, 8 May 2023

  • The beam position monitors (BPMs) on the diagnostic girder in Hall C (IPM3H07A, IPM3H07B, and IPM3H07C) are calibrated by comparison with the Hall C superharps (IHA3H07A and IHA3H07B). Ideally, a "bulls-eye" scan is done which allows one to determine the BPM offsets and slopes. However, a single harp scan is sufficient to check the BPM offsets.
  • If re-analyzing/revisiting BPM calibrations, make sure that the appropriate survey offsets are loaded into the harp analyzer tool
  • Table of BPM calibration information. Offsets and slopes are from the "RAW" epics variables (no survey offsets applied).
Date Method AXoff BXoff CXoff AYoff BYoff CYoff AXslope BXslope CXslope AYslope BYslope CYslope Notes
1/13/2018 Calibration -0.06 0.07 -1.06 -0.46 0.12 0.61 -0.98 -1.12 -0.96 0.97 1.17 0.87 1st calibration
8/29/2018 Harp scan 0.15 0.29 -1.08 -0.12 0.63 1.09 - - - - - - -
2/9/2019 Harp scan 0.25 0.46 -1.01 -0.08 -0.04 -0.11 - - - - - - Radiator installed during January 2019 SAD
6/23/2019 Calibration 0.11 0.17 -1.07 -0.40 -0.18 -0.40 -0.97 -1.11 -0.95 0.95 1.15 0.85 -
12/8/2019 Calibration 0.14 0.22 -1.13 -0.06 -0.29 -0.78 -0.96 -1.09 -0.96 0.95 1.16 0.85 Polarized 3He running
9/5/2021 Harp scan -0.35 0.17 -0.83 -0.35 0.19 0.47 - - - - - -
1/9/2022 Harp scan -0.08 0.32 -0.67 -0.40 0.20 0.43 - - - - - -
6/9/2022 Harp scan 0.16 -0.22 -0.54 -0.18 0.02 0.39 - - - - - -
1/11/2023 Harp scan 0.06 -0.35 -0.69 -0.20 -0.03 0.38 - - - - - -
  • Summary of survey information
3/7/2017 !! -2.84 !! -2.88 !! -3.27 !! -0.21 !! -0.20 !! 0.05 !! -2.88 !! -3.07 !! -0.13 !! -0.04 !! 12/1/2017 !! -0.01 !! -0.01 !! -0.17 !! -0.27 !! -0.22 !! -0.16 !! 0.01 !! -0.12 !! -0.12 !! 0.02 !! 1/10/2018 !! - !! - !! - !! - !! - !! - !! 0.01 !! 0.11 !! -0.12 !! 0.08 !! Harps only after wire repair
Date BPM AXoff BPM BXoff BPM CXoff BPM AYoff BPM BYoff BPM CYoff Harp AXoff Harp BXoff Harp AYoff Harp BYoff Notes