Line 7: Line 7:
 
|-
 
|-
 
! 8/29/2018  !! Harp scan !! 0.15 !! 0.29  !! -1.08 !! -0.12 !! 0.63 !! 1.09 !! - !!  - !! - !! - !! - !! -
 
! 8/29/2018  !! Harp scan !! 0.15 !! 0.29  !! -1.08 !! -0.12 !! 0.63 !! 1.09 !! - !!  - !! - !! - !! - !! -
 +
|-
 +
! 2/9/2019  !! Harp scan !! 0.25 !! 0.46  !! -1.01 !! -0.08 !! -0.04 !! -0.11 !! - !!  - !! - !! - !! - !! -
 +
|-
 +
! 6/23/2019  !! Calibration !! 0.11 !! 0.17  !! -1.07 !! -0.40 !! -0.18 !! -0.40 !! -0.97 !! -1.11 !! -0.95 !! 0.95 !! 1.15 !! 0.85
 
|-  
 
|-  
 
|}
 
|}

Revision as of 14:23, 5 May 2023

  • The beam position monitors (BPMs) on the diagnostic girder in Hall C (IPM3H07A, IPM3H07B, and IPM3H07C) are calibrated by comparison with the Hall C superharps (IHA3H07A and IHA3H07B). Ideally, a "bulls-eye" scan is doe which allows one to determine the BPMS offsets and slopes. However, a single harp scan is sufficient to check the BPM offsets.
Date Method AXoff BXoff CXoff AYoff BYoff CYoff AXslope BXslope CXslope AYslope BYslope CYslope
1/13/2018 Calibration -0.06 0.07 -1.06 -0.46 0.12 0.61 -0.98 -1.12 -0.96 0.97 1.17 0.87
8/29/2018 Harp scan 0.15 0.29 -1.08 -0.12 0.63 1.09 - - - - - -
2/9/2019 Harp scan 0.25 0.46 -1.01 -0.08 -0.04 -0.11 - - - - - -
6/23/2019 Calibration 0.11 0.17 -1.07 -0.40 -0.18 -0.40 -0.97 -1.11 -0.95 0.95 1.15 0.85