BPM calibration information
From HallCWiki
Jump to navigationJump to search- The beam position monitors (BPMs) on the diagnostic girder in Hall C (IPM3H07A, IPM3H07B, and IPM3H07C) are calibrated by comparison with the Hall C superharps (IHA3H07A and IHA3H07B). Ideally, a "bulls-eye" scan is doe which allows one to determine the BPMS offsets and slopes. However, a single harp scan is sufficient to check the BPM offsets.
Date | Method | AXoff | BXoff | CXoff | AYoff | BYoff | CYoff | AXslope | BXslope | CXslope | AYslope | BYslope | CYslope |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1/13/2018 | Calibration | -0.06 | 0.07 | -1.06 | -0.46 | 0.12 | 0.61 | -0.98 | -1.12 | -0.96 | 0.97 | 1.17 | 0.87 |
8/29/2018 | Harp scan | 0.15 | 0.29 | -1.08 | -0.12 | 0.63 | 1.09 | - | - | - | - | - | - |
2/9/2019 | Harp scan | 0.25 | 0.46 | -1.01 | -0.08 | -0.04 | -0.11 | - | - | - | - | - | - |
6/23/2019 | Calibration | 0.11 | 0.17 | -1.07 | -0.40 | -0.18 | -0.40 | -0.97 | -1.11 | -0.95 | 0.95 | 1.15 | 0.85 |
12/8/2019 | Calibration | 0.14 | 0.22 | -1.13 | -0.06 | -0.29 | -0.78 | -0.96 | -1.09 | -0.96 | 0.95 | 1.16 | 0.85 |