BPM calibration information
From HallCWiki
Jump to navigationJump to search- The beam position monitors (BPMs) on the diagnostic girder in Hall C (IPM3H07A, IPM3H07B, and IPM3H07C) are calibrated by comparison with the Hall C superharps (IHA3H07A and IHA3H07B). Ideally, a "bulls-eye" scan is done which allows one to determine the BPM offsets and slopes. However, a single harp scan is sufficient to check the BPM offsets.
- If re-analyzing/revisiting BPM calibrations, make sure that the appropriate survey offsets are loaded into the harp analyzer tool
- Table of BPM calibration information. Offsets and slopes are from the "RAW" epics variables (no survey offsets applied).
Date | Method | AXoff | BXoff | CXoff | AYoff | BYoff | CYoff | AXslope | BXslope | CXslope | AYslope | BYslope | CYslope | Notes |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1/13/2018 | Calibration | -0.06 | 0.07 | -1.06 | -0.46 | 0.12 | 0.61 | -0.98 | -1.12 | -0.96 | 0.97 | 1.17 | 0.87 | 1st calibration |
8/29/2018 | Harp scan | 0.15 | 0.29 | -1.08 | -0.12 | 0.63 | 1.09 | - | - | - | - | - | - | - |
2/9/2019 | Harp scan | 0.25 | 0.46 | -1.01 | -0.08 | -0.04 | -0.11 | - | - | - | - | - | - | Radiator installed during January 2019 SAD |
6/23/2019 | Calibration | 0.11 | 0.17 | -1.07 | -0.40 | -0.18 | -0.40 | -0.97 | -1.11 | -0.95 | 0.95 | 1.15 | 0.85 | - |
12/8/2019 | Calibration | 0.14 | 0.22 | -1.13 | -0.06 | -0.29 | -0.78 | -0.96 | -1.09 | -0.96 | 0.95 | 1.16 | 0.85 | Polarized 3He running |
9/5/2021 | Harp scan | -0.35 | 0.17 | -0.83 | -0.35 | 0.19 | 0.47 | - | - | - | - | - | - | |
1/9/2022 | Harp scan | -0.08 | 0.32 | -0.67 | -0.40 | 0.20 | 0.43 | - | - | - | - | - | - | |
6/9/2022 | Harp scan | 0.16 | -0.22 | -0.54 | -0.18 | 0.02 | 0.39 | - | - | - | - | - | - | |
1/11/2023 | Harp scan | 0.06 | -0.35 | -0.69 | -0.20 | -0.03 | 0.38 | - | - | - | - | - | - |
- Beamline surveys