Spectrometer Rotation Reset
From HallCWiki
Jump to navigationJump to searchThis is how to reset the spectrometer rotation controls.
You *must* have explicit permission from Steve Lassiter to do this operation. No exceptions!
- Locate the HMS rotation control box on the beam-right side of the HMS carriage
- Disable the rotation control by turning the key from 'Normal' to 'Bypass'
- Locate the HMS breaker panel (to the left of the rotation control panel) on the HMS carriage
- Turn the breaker labeled 'HMS Drive' (22) OFF, then back ON
- Locate the SHMS rotation control box at the rear of the SHMS carriage between the wheels (be careful of the angle camera!)
- Disable the rotation control by turning the key from 'Normal' to 'Bypass'
- Locate the SHMS breaker panel (on the beam-left side of the carriage, under the upper platform, downstream of the stairway).
- Turn the breaker labeled 'DRIVE MOTOR/SHUNT TRIP' (39) OFF, then back ON
- Call the Counting House and ask the shift crew to reset the rotation controls for both spectrometers.
- Re-enable the remote rotation controls in both the SHMS and HMS by turning their keys back to 'Normal'
- Call the Counting House and ask for them to try rotating the problematic arm(s). If it works, go back upstairs.