1-pass calibration for A1n
From HallCWiki
Jump to navigationJump to searchOne-pass calibration/commissioning for Fall 2019
For polarized 3He running in Fall 2019 and Spring 2020, the nominal raster for production is 5mm diameter circular. The nominal triggers used are 3/4 for both HMS and SHMS. Both HMS and SHMS should be taking data (for example if run plan specify one run, it means one run on HMS and one run on SHMS simultaneously). No coincidence trigger will be used/setup.
Outline
We will need to do optics calibration, reference cell (cross section and pressure curves), 3He elastic longitudinal asymmetry, and 3He Delta transverse asymmetry measurements.
Beam and raster checkout
Please refer to 2-pass commissioning procedure, follow: - Establish beam in the hall - Initial Raster checks - Harp scans - Target/Beam centering, Raster calibration
Optics (angle and ytarg) Calibration
Goals - To calibration spectrometer optics for 40cm long target at 30 deg. Calibrate angle reconstruction.
Setup - Beam Current: 5 -- 20 uA CW, raster varies - HMS setting: 30 degrees, -1.2 GeV/c (electron) - SHMS setting: 30 degrees, -1.68 GeV/c (electron) - Target: Multi(7)-foil Carbon Foils
Procedure - Move target to multi-foil position; - Move Sieve Slit IN for both HMS and SHMS; - Take a short run (2min) at 20uA, verify sieves are both IN; - Ask MCC to increase current to 20uA (Max for raster off on this target). Take two runs, 45 minutes each. - Ask MCC to turn raster ON to 5mm diameter, take one run at 20 uA for 45 minutes. - Move Sieve Slit OUT for both HMS and SHMS; - Take one run at 20 uA for 15 minutes. - Ask MCC to turn raster OFF, take one run at 20 uA for 15 minutes.
Optics (Momentum) Calibration
Goals - To calibration momentum reconstruction of spectrometers.
Setup - Beam Current: 5--20 uA CW, raster always OFF - HMS setting: 11.7 deg, momentum varies (electron) - SHMS setting: 8.5 deg, momentum varies (electron) - Target: Double Carbon Foils
Procedure - Move target to Double carbon foil position.
- Ask the target operator to do one target polarimetry measurement. - Move target to polarized 3He cell position, take a 2 minute run at 5uA and confirm we see events from both 3He gas and two glass windows at z=+20 and -20cm. - Stop the beam, ask the target operator to do one target polarimetry measurement. - take a 10 (what's the time constant for beam depolarization to set in?) minute run at 30uA, watch the beam ramp rate (confirm it is no faster than 1uA/sec), and confirm we see events from both 3He gas and two glass windows at z=+20 and -20cm. - Stop the beam, ask the target operator to do one target polarimetry measurement.
Polarized 3He Delta Transverse Asymmetry
Goals - To measure/confirm sign of physics transverse asymmetry of Delta resonance
Setup - Beam Current: up to 5 uA CW, raster always ON (5mm diameter) - HMS setting: 11.7 deg, -2.068 GeV/c (electron) - SHMS setting: 8.5 deg, -2.083 GeV/c (electron) - Target: Polarized 3He Cell - Target spin direction: 0-deg, 180-deg - Detectors: We will need to keep only 2 SC bars ON to limit quasi-elastic scattering events. - Parity check: Inform Hall A prior to taking data to make sure they set the charge asymmetry feedback cutoff at 0.5uA.
Procedure - Set spectrometers at the required settings, target at Polarized 3He Cell position, spin polarization at either 0 or 180 degrees. - Perform one polarimetry measurement before taking data. - Take one 1-minute run at 1 uA, check focal plane event distribution and determine which two scintillator bars should be kept on. (expert driven). - Once scintillators are set, take one 1-minute run at 2uA, check rates and adjust prescale factor. Adjust current so HMS is at 4.5kHz PS=1. SHMS expected to require PS=15 to limit DAQ rate at 4.5kHz. - Take three 30-minute runs at the optimal current set above, perform one polarimetry measurement after each run. Verify data quality (see below) and verify target is still polarized. - Rotate target spin direction from 0 to 180 deg (or from 180 to 0 deg) - Take three 30-minute runs at the optimal current set above, perform one polarimetry measurement after each run. Verify data quality (see below) and verify target is still polarized.
Things to watch online - Ensure the correct SC bars are on (We need elastic events! One SC bar is in principle enough but we keep TWO on) - Apply y or z cuts to isolate events scattered from 3He gas (We do not want events from the glass windows!) - Apply W or P cuts to isolate elastic events from others. - Run online script to ensure we see physics asymmetry.
- Expected total rate: HMS at 1uA total 1.8KHz 3He gas elastic 240Hz; SHMS at 1uA total 27.7kHz Prescaled by 15 to 3He gas elastic 740Hz - Expected asymmetries (after W and Ytg cuts): HMS Aphys=4.8% Ameas=2.1% and SHMS Aphys=2.5% Ameas=0.98% if Pt=50%, Pb=85%, N2 dilution 0.92. - Expected uncertainties dA/A=3% on HMS.
Polarized 3He Elastic Longitudinal Asymmetry
Goals - To measure physics longitudinal asymmetry of 3He elastic scattering, cross check PbPt from Moller and Target Polarimetry
Setup - Beam Current: up to 5 uA CW, raster always ON (5mm diameter) - HMS setting: 11.7 deg, -2.068 GeV/c (electron) - SHMS setting: 8.5 deg, -2.083 GeV/c (electron) - Target: Polarized 3He Cell - Target spin direction: 0-deg, 180-deg - Detectors: We will need to keep only 2 SC bars ON to limit quasi-elastic scattering events. - Parity check: Inform Hall A prior to taking data to make sure they set the charge asymmetry feedback cutoff at 0.5uA.
Procedure - Set spectrometers at the required settings, target at Polarized 3He Cell position, spin polarization at either 0 or 180 degrees. - Perform one polarimetry measurement before taking data. - Take one 1-minute run at 1 uA, check focal plane event distribution and determine which two scintillator bars should be kept on. (expert driven). - Once scintillators are set, take one 1-minute run at 2uA, check rates and adjust prescale factor. Adjust current so HMS is at 4.5kHz PS=1. SHMS expected to require PS=15 to limit DAQ rate at 4.5kHz. - Take three 30-minute runs at the optimal current set above, perform one polarimetry measurement after each run. Verify data quality (see below) and verify target is still polarized. - Rotate target spin direction from 0 to 180 deg (or from 180 to 0 deg) - Take three 30-minute runs at the optimal current set above, perform one polarimetry measurement after each run. Verify data quality (see below) and verify target is still polarized.
Things to watch online - Ensure the correct SC bars are on (We need elastic events! One SC bar is in principle enough but we keep TWO on) - Apply y or z cuts to isolate events scattered from 3He gas (We do not want events from the glass windows!) - Apply W or P cuts to isolate elastic events from others. - Run online script to ensure we see physics asymmetry.
- Expected total rate: HMS at 1uA total 1.8KHz 3He gas elastic 240Hz; SHMS at 1uA total 27.7kHz Prescaled by 15 to 3He gas elastic 740Hz - Expected asymmetries (after W and Ytg cuts): HMS Aphys=4.8% Ameas=2.1% and SHMS Aphys=2.5% Ameas=0.98% if Pt=50%, Pb=85%, N2 dilution 0.92. - Expected uncertainties dA/A=3% on HMS.
Reference Cell Hydrogen (H2) Cross Section
Goals - To measure hydrogen elastic cross section, need 10,000 elastic events each for HMS and SHMS
Setup
Procedure
Reference N2 Pressure Curve
Goals - To measure yield and cross section for varying N2 pressure. This will be used to determine amount of N2 in the polarized 3He cell in the offline analysis.
Setup
Procedure
Reference 3He Pressure Curve
Goals - To measure yield and cross section for varying N2 pressure. This will be used to determine amount of 3He in the polarized 3He cell in the offline analysis. - Important: Among all reference cell runs (N2, H2 and 3He), run 3He last, and do NOT vent 3He gas after completing the run plan.
Setup
Procedure