BPM calibration information

From HallCWiki
Revision as of 16:16, 5 May 2023 by Gaskelld (talk | contribs)
Jump to navigationJump to search
  • The beam position monitors (BPMs) on the diagnostic girder in Hall C (IPM3H07A, IPM3H07B, and IPM3H07C) are calibrated by comparison with the Hall C superharps (IHA3H07A and IHA3H07B). Ideally, a "bulls-eye" scan is done which allows one to determine the BPM offsets and slopes. However, a single harp scan is sufficient to check the BPM offsets.
  • If re-analyzing/revisiting BPM calibrations, make sure that the appropriate survey offsets are loaded into the harp analyzer tool
  • Table of BPM calibration information (still under construction)
Date Method AXoff BXoff CXoff AYoff BYoff CYoff AXslope BXslope CXslope AYslope BYslope CYslope Notes
1/13/2018 Calibration -0.06 0.07 -1.06 -0.46 0.12 0.61 -0.98 -1.12 -0.96 0.97 1.17 0.87 1st calibration
8/29/2018 Harp scan 0.15 0.29 -1.08 -0.12 0.63 1.09 - - - - - - -
2/9/2019 Harp scan 0.25 0.46 -1.01 -0.08 -0.04 -0.11 - - - - - - Radiator installed during January 2019 SAD
6/23/2019 Calibration 0.11 0.17 -1.07 -0.40 -0.18 -0.40 -0.97 -1.11 -0.95 0.95 1.15 0.85 -
12/8/2019 Calibration 0.14 0.22 -1.13 -0.06 -0.29 -0.78 -0.96 -1.09 -0.96 0.95 1.16 0.85 Polarized 3He running
9/5/2021 Harp scan -0.35 0.17 -0.83 -0.35 0.19 0.47 - - - - - -
1/9/2022 Harp scan -0.08 0.32 -0.67 -0.40 0.20 0.43 - - - - - -
6/9/2022 Harp scan 0.16 -0.22 -0.54 -0.18 0.02 0.39 - - - - - -
1/11/2023 Harp scan 0.06 -0.35 -0.69 -0.20 -0.03 0.38 - - - - - -