Study of SHMS resolution
From HallCWiki
Revision as of 10:53, 29 February 2012 by Jones (talk | contribs) (→Comparing resolutions with vaccuum pipe versus helium bag)
Using the following setup for SHMS single arm MC
- 2 GeV , 20 deg scattered electrons
- Reconstruction of tracks in the focal plane assumed 200 microns resolution.
- LH2 target with 20cm length. Assume beer can with radius of 3.37cm and wall thickness in 0.005in Al.
- 16 mil Al scattering chamber window (X0=8.89cm) followed by 15 cm air (X0=30420cm)
- Spectrometer entrance window 15 mil Kevlar (X0=74.6cm) and 5 mil Mylar (X0=28.7cm)
- Two options for spectrometer exit window,
- 15 mil Kevlar, 5 mil Mylar (use 20 mil,X0=53.3)
- 0.02inches of Aluminum
- Two options in Hut when AR/Neon Cerenkov is removed
- Vacuum pipe so the the spectrometer exit window is moved to location of Cerenkov exit window
- Helium bag. The spectrometer exit window is located in front of Cerenkov entrance, but the Cerenkov is replaced by a helium bag. Assume the entrance and exit of helium bag is aluminized mylar. Each window is 17 micron mylar and 3 micron aluminum. Filled with helium ( Radlen = 754560 cm).
- Drift through air until drift chamber 1
- Chamber 1 entrance foil, 1 mil of Mylar
- Scattering in drift chamber, for each plane
- Effective cathode plane, Be/Cu with Rad Len = 28.7cm , 0.0005 in
- Drift through Chamber gas, 50/50 ethane/argon with Rad Len = 16700.0cm
- Effective wire plane, 25 micron W+Be/Cu
- Fill hit position for plane
- Drift through Chamber gas, 50/50 ethane/argon with Rad Len = 16700.0cm
- Chamber 1 exit foil, 1 mil of Mylar
- Drift through air until drift chamber 2
- Chamber 2 entrance foil, 1 mil of Mylar
- Scattering in drift chamber, for each plane
- Effective cathode plane, Be/Cu with Rad Len = 28.7cm , 0.0005 in
- Drift through Chamber gas, 50/50 ethane/argon with Rad Len = 16700.0cm
- Effective wire plane, 25 micron W+Be/Cu
- Fill hit position for plane
- Drift through Chamber gas, 50/50 ethane/argon with Rad Len = 16700.0cm
- Chamber 2 exit foil, 1 mil of Mylar
Comparing resolutions with vaccuum pipe versus helium bag
- Monte Carlo run for 2 GeV electrons for 20cm LH2 target with spectrometer at 20 degs.
- Table with the resolution for xptar, yptar, ytar and delta for all events
Single arm Monte Carlo Flags | Xptar (mr) | Yptar (mr) | Delta % | Ytar (cm) | |
---|---|---|---|---|---|
1 | no wire chamber smearing, no multiple scattering in material | 0.08 | 0.01 | .014 | .002 |
2 | wire chamber smearing, no multiple scattering in material | 0.36 | 0.26 | .038 | .084 |
3 | wire chamber smearing, multiple scattering in all materials with Kevlar exit window, vacuum pipe | 1.00 | 0.94 | .072 | .12 |
4 | wire chamber smearing, multiple scattering in all materials with Kevlar exit window,helium bag | 1.04 | 0.96 | .092 | .14 |
5 | wire chamber smearing, multiple scattering in all materials with Al exit window, vacuum pipe | 1.15 | 1.05 | .102 | .14 |
6 | wire chamber smearing, multiple scattering in all materials with Al exit window,helium bag | 1.10 | 1.02 | .134 | .16 |
- Plots of [xptar] , [yptar], [ytar] and [delta] resolutions versus the thrown delta, xptar and yptar comparing the first four different sets of MC flags listed in the above table.
- Plots of [xptar] , [yptar], [ytar] and [delta] resolutions versus the thrown delta, xptar and yptar comparing the sets 1,2,5 and 6 listed in the above table.
- Plot of [delta] versus the thrown delta, xptar and yptar comparing sets 3,4,5 and 6 listed in the above table.