next up previous
Next: Monitor Up: Hall C Expert Howto Previous: Purpose and Optics

Setting the magnet

Since this magnet is not directly effect the beam, thus it will be controlled and operated by Hall C experimentalists. However, due to fringe field does have slight effect to beam and other magnets nearby, such as Splitter and HKS Q1, it is preferred to turn off the beam during the process of setting up the magnet. Since this magnet is sharing the same power supply with Moller magnet, existing control will be used in setting up the magnet by referring to steps of ``Getting started'' and ``Setting the magnets'' in ``How to set the SOS magnets'' in Hall C online Howto document. Since no polarity change in operating this magnet for hypernuclear experiment, step of ``Degaussing'' is not Needed.



Stephen A. Wood 2005-05-13