Obtaining a BCM calibration consists of two steps that cross calibrate the BCM's to the Unser Monitor. First the response of the Unser and the BCM's is measured at a series of beam currents that cover the range of beam currents to the current experiment. Second those measurements are analyzed to produce calibration constants for the BCM's.
The following is the recommended method of acquiring BCM calibration data. This procedure was designed for the Spring 2003 Hall C running. For other experiments the procedure may be modified, depending on the range of currents required by the experiment, but the procedure will be similar.
0, 10, 0, 20, 0, 30, 0, 40, 0, 50, 0, 60, 0, 70, 0, 80, 0, 90, 0, 100,
and should then be repeated. Each current setting should be 1.5-2 minutes duration.
If the green light is flashing on the scaler crate, you're probably taking data. The files are usually small enough to fit in the daq buffer, so the output .log file will be nearly empty until you finally stop the run.
via charge####.txt
.
The basic principle of the analysis is as follows. The Unser monitor has a well known calibration, but it has an offset that is both noisy and drifts with time. In taking the calibration data, each different beam current is bracketed by a period of beam off so that the drifting offset can be well determined for each different beam current used in the cabibration. With these offsets determined, the average current for each nominal current is well determined and the unser calibration can be trasnferred to the BCM's. The BCM's are not linearly, particularly at small currents, so the calibration fit is made using only currents over the range that is required by the experiment. Typically this means that zero current is excluded from the fit.