Standard models used in predicting vacuum
window performance do not predict the actual performance of Kevlar laminate
fabrics accurately [2,4]. Therefore, extensive tests of the
Mylar/Kevlar composite windows have been done and more are planned.
A summary of pressure
tests on windows composed of the custom laminated
fabric currently used in construction of all Hall C vacuum windows is
given in Tables 2.6 and 2.7, below. Both vacuum and hydrostatic test
tanks
are available for the HMS circular windows and for the SOS rectangular
window. Figure 2.21 shows a picture of the SOS vacuum
test tank. The large round
HMS vacuum test tank uses the inch vacuum extension piece shown
earlier in Figure 2.19 (but removed from the spectrometer) with a
inch thick aluminum blanking flange. For hydrostatic testing, windows (both
round and rectangular) are mounted directly on
blanking flanges with appropriate water fittings installed.
The tests show that the large HMS exit window does not begin to leak until
the pressure on it reaches over atmospheres. The small HMS window
did not begin to leak until over
atmospheres of pressure was applied.
This is consistent with the load being transferred to
the outer circumference, the circumference of the small circle being four
times smaller than that of the large circle. The large rectangular SOS windows
began to leak at around 10 atmospheres. It may be possible to construct thinner
SOS vacuum windows for improved spectrometer resolution if required.
In addition to determining the absolute failure pressure of the windows, a major goal of the testing effort was to observe the failure modes of the window material and flange design. The large windows failed by ripping around the opening perimeter. The flange ring was then rounded more and a better result was obtained. It may be possible to further round the flange if future tests indicate this to be necessary or advantageous. The small windows failed by ripping out at the bolt holes.
Another question the tests in Table 2.6 were designed to answer was if a biting aluminum clamping O-ring was necessary or advantageous. It was found that windows with aluminum clamping rings failed at lower pressures than windows without aluminum rings. The aluminum tended to tear the Mylar, allowing leaking and uneven stress. The only window which delaminated had water between the Mylar and Kevlar which leaked in through a tear in the Mylar caused by the aluminum ring.
In conjunction with outlined full-size window testing, small samples
were subjected to stress and tensile analysis using strain gauge
instrumentation. Through this work, it was found that the composite
window material maintained the Young's Modules of Kevlar, and had the
same ultinate failure load as Kevlar pieces of the same size.
Further, it was determined that the failure tears were along a
45 angle to the woven fibers, consistent with observations
of the rips in the destructively-tested full-size windows.
In addition to the above testing efforts, two large round HMS exit windows composed of the custom laminate material were ``knife" tested. The windows were installed on the vacuum test tank and placed under vacuum. They were then cut by a blade at the end of a long pole (so as to protect the personnel testing the window from the results of a catastrophic failure). The window did not rip any further than the cut and the material did not delaminate. No evidence for catastrophic failure was observed in either of these tests.
Although more long term reliability testing is necessary (see below),
results have so far been favorable.
Windows previously used on the HMS spectrometer are routinely hydro
tested to failure after biannual removeable and found to
hold up as well as freshly made windows (see, for example
Table 2.6). Windows which have undergone multiple cycles fail at nearly the same pressure as uncycled
windows (see, for example, Table 2.7). A small round window did not fail after
being left pressurized for
a week at psi. (this test is not entered in the tables).
Entrance and exit windows composed of the Mylar/Kevlar laminate
material here described have held vacuum on the Hall C
spectrometer effectively and safely for over three years.