Log entry time 16:44:49 on April 13, 2007

Entry number 136455

This entry is a followup to: 136450

Followups:

keyword=HMS Fastbus slot 19 bad

I tried sending the gate for the 1877 TDC in HMS hut fastbus slot 19 to the front panel, but it still would not work.

I further ran self tests on the module. From the VXworks prompt, I ran

test_1877(19,2)

ends with the message "Sparse Data scan indicates no Conversion after 20 tries". Running the test on other slots does not give that error, so it looks like we need to replace that module. That's a job for next week. It should not be too difficult as it is towards the left side of the crate where the cables are not packed in so tight.