Log entry time 11:30:54 on January 21, 2008
Entry number 149107
This entry is a followup to: 149091
keyword=updated BigCal ADC thresholds from replay of run 68088
I copied the thresholds calculated from run 68088 into
gep07/coda25/roc11(12)_thresholds.dat. Let's keep an eye on this and make
log entries whenever this problem happens. Something is wrong with either
the HV/PMTs or hopefully with the ADC module (more easily replaced) on
rows 29 and 30, because the pedestals are very unstable and have been for
some time. The new ADC thresholds should take effect next before the next
coin run.