Log entry time 02:32:49 on February 23, 2011

Entry number 221617

keyword=Target: close 4K JT by 2%, 15K JT by 12%

We had a lot of reserve heating power (almost 600W). When I talked to Greg about the alarm limits, he pointed this out and said I could close the JT valves. I closed the 4K JT valve by 2%, and the 15K JT valve by 12%. 15K flow now 31.6 g/sec, before it was closer to 37 g/sec. We still have ~380 W of reserver heating power at 150 uA, but I don't feel comfortable closing the JT further.



A copy of this log entry has been emailed to: smithg, covrig



Figure 1