==== Fiber RTD hardware trip points ====
 +
* Fire Safety RTDs
 +
** HcFiberTemp0, HcFiberTemp1
 +
** max measured about 32C
 +
** '''set trip to 36C'''
 +
 +
* Fiber Temperature RTDs
 +
** HcFiberTemp4+  (except for 12, 14 below)
 +
** Max measured of about 32C
 +
** '''set trip to 46C''' (as of Jan 14, 2020 -- '''will set to 36C''' during next downtime)
 +
 +
* HcFiberTemp12, 14
 +
** Max measured of 40C
 +
** '''set trip to 46C'''
 +
 +
<div id="collapse-pre-one" class="mw-collapsible mw-collapsed">
 +
<div class="mw-collapsible-toggle toccolours" style="float: none;">
 +
  <div class="mw-collapsible-toggle-row">
 +
  <div class="mw-collapsible-toggle-header"><big>'''Reset instructions &bull; Expert only (Brad or Ethan) [click]'''</big></div>
 +
  </div>
 +
</div>
 +
<div class="mw-collapsible-content">
 +
Address 0xF6
 +
  FiberTemp0-1:  (and the unused 2, 3)
 +
    Fire safety RTDs (max 36)
 +
Address 0xF5
 +
  FiberTemp4--7
 +
  * max 36
 +
Address 0xF3
 +
  FiberTemp8--11
 +
  * max 36
 +
Address 0xF4
 +
  FiberTemp12-15
 +
  * max 46 due to 12, 14
 +
 +
'''NOTE''': power cycling the acromag controller below will trip the laser interlock!
 +
 +
* Set the hardware trip limits
 +
% ssh pi@polhe3pi2  ('hccuser' password)
 +
% cd Documents
 +
% scls  # select 'acromag' entry, then type Ctrl-C to kill the program
 +
# Set the limits
 +
% for f in 0xf3 0xf5 0xF6; do ./setupAcro -l 36 -a $f; done
 +
% for f in 0xf4; do ./setupAcro -l 46 -a $f; done
 +
# Reboot the acromag controller using hcreboot6
 +
# Then restart the software by (for example) running bash again
 +
% bash
 +
 +
* Verify that the stripcharts are updating again and close the terminal window.
 +
</div>
  
 
== Additional Documentation ==
 
== Additional Documentation ==

Revision as of 11:36, 14 January 2020

Target Operator Instructions

FIXME: ADD LINKS/DOCS/WHATEVER HERE


System Description

Checklists

Expert Information

Pol He3 Field Mapping Data

Circular Raster Information

Reboot Switches

Reset Target Ladder Motion Control

Pol He-3 Target Motion Reset

Cable/RTD/Signal Maps

NOTE: If you update a map below
 Click on a the previous hclog entry and either add a comment 
 pointing to your new entry, or use the 'Follow-up' link on the
 old entry to make a new entry. That will make it easier to track
 when/why things change.

Fiber RTD hardware trip points

  • Fire Safety RTDs
    • HcFiberTemp0, HcFiberTemp1
    • max measured about 32C
    • set trip to 36C
  • Fiber Temperature RTDs
    • HcFiberTemp4+ (except for 12, 14 below)
    • Max measured of about 32C
    • set trip to 46C (as of Jan 14, 2020 -- will set to 36C during next downtime)
  • HcFiberTemp12, 14
    • Max measured of 40C
    • set trip to 46C
Reset instructions • Expert only (Brad or Ethan) [click]
Address 0xF6
 FiberTemp0-1:  (and the unused 2, 3)
   Fire safety RTDs (max 36)
Address 0xF5
 FiberTemp4--7
 * max 36
Address 0xF3
 FiberTemp8--11
 * max 36
Address 0xF4
 FiberTemp12-15
 * max 46 due to 12, 14
NOTE: power cycling the acromag controller below will trip the laser interlock!
  • Set the hardware trip limits
% ssh pi@polhe3pi2  ('hccuser' password)
% cd Documents
% scls  # select 'acromag' entry, then type Ctrl-C to kill the program
# Set the limits
% for f in 0xf3 0xf5 0xF6; do ./setupAcro -l 36 -a $f; done
% for f in 0xf4; do ./setupAcro -l 46 -a $f; done
# Reboot the acromag controller using hcreboot6
# Then restart the software by (for example) running bash again
% bash
  • Verify that the stripcharts are updating again and close the terminal window.