Line 85: Line 85:
 
| Ask the target operator (TO).
 
| Ask the target operator (TO).
 
| Record which target is in the beam.
 
| Record which target is in the beam.
 +
|-
 +
| Drift Ch. Mixer Flows: Ar/Eth (~0.3/~0.3 slpm)
 +
| Type “go munin” and select “gas_shed” under Categories on the left side of the page. You can also find the needed charts here: [[https://hallcweb.jlab.org/munin/gas_shed-day.html Flow and Temperature Charts]]
 +
| Record “Argon Flow” and “Ethane Flow.”
 +
|-
 +
| Drift Chamber Gas Delivery Press (500 ± 5 Torr)
 +
| Same as above.
 +
| Record “Ar:Eth Pressure.”
 +
|-
 +
| Drift Ch. Gas Isopropanol Temp. (2 ± 2 C)
 +
| Same as above.
 +
| Record the “Isopropanol Fridge” temperature.
 +
|-
 +
|
 +
|
 +
|
 +
|-
 +
|
 +
|
 +
|
 +
|-
 +
|
 +
|
 +
|
 +
|-
 +
|
 +
|
 +
|
 +
|-
 +
|
 +
|
 +
|
 +
|-
 +
|
 +
|
 +
|
 +
|-
 +
|
 +
|
 +
|
 +
|-
 +
|
 +
|
 +
|
 +
|-
 +
|
 +
|
 +
|
 +
|-
 +
|
 +
|
 +
|
 
|-
 
|-
 
| SHMS Trigger Rate
 
| SHMS Trigger Rate

Revision as of 11:05, 2 October 2019

Direct link to the Hall C Shift Checklist

Shift Checklist Help

  • The How-to page is extremely helpful. Use Firefox's "Page search" to help find what you are looking for.
  • There is also an older Shift Checklist help file, which may help to locate items. (But it is a stale document.) Below is an updated table.
What Where to find it How to do it
DATE (day-month-year) As given by Run Control Enter as DD-MM-YY
Time (24-hour clock) As given by Run Control Write the time into the cell
Your Name As given by your parents Write legibly
Beam Conditions from Overview Screen, DAQ, MEDMs, and Scalers
Beam Energy (MeV) The Hall C Summary Screen should be displayed on a wall monitor. If not see this page to bring it up: EPICS/Accelerator_Screens. Write down the Hall C beam energy
Beam Current (uA) Same as above MPS and Hall C BCM readings should be similar. Record the

reading

Beam position lock on? Use jmenu and search for 'lock.' Then choose 'Orbit Locks Normal View

and look for 'HallCOrb.' It should be in green and show 'Running.'

Write 'yes' or 'no' in the cell
Beam energy lock on? Use jmenu and search for 'lock.' Then choose 'Energy Locks.' Write down the three colors: one for '0R, 1A, 2A.' (They should all be green. This indicates 'yes.')
IPM3H07A, B, and C BPM positions. Are they nominal? See the Hall C Summary Screen. Record the displayed beam position. Compare it with the nominal position posted on

the white board or sticky note.

Fast Raster Pattern Go to the electronics room and verify the raster pattern on the oscilloscope screen is circular. Record 'yes' if so (you should! especially if beam is on!) If not, notify the TO and MCC.
Fast Raster Setpoint for Width,

Height [x, y] (mm)

Use jmenu and search for 'fast raster.' The screen is "Hall C Raster Controls." Record the set sizes in the “Set Raster Size” box.
Fast Raster Measured Currents (IxO3H, Ix04H, Iy03V, IyO4V) [A] Use jmenu and search for 'fast raster.' The screen is "Hall C Raster Controls." Record the the four measured currents (Ix and Iy for two H and two V coils) from the

“Raster Measured Current” box.

Helicity Reporting Delay (No Delay, ___ Windows) Use jmenu and search for 'helicity,' then choose 'Helicity Control.' Look for 'Reporting Delay.' Record the time of delay (it should be 8 windows).
Alarm Handler Status (General) The Alarm Handler process must be running and always visible on one of the desktop machines. If it isn't, bring it up using this page: Hall_C_Alarm_Handler Verify that it is running, and that the alarm sound is NOT silenced. (N.B.: this is not the old legacy high-voltage alarm. )
HMS Collimator Setting Use jmenu → 'Standalone Menus' → 'Hall C.' On the Hall C Operations Menu under “Experiment Specific” is HMS Coll. Motion Control. Determine and record the collimator position. For physics runs, HMS should be either PION or LARGE.
SHMS Collimator Setting Use jmenu → 'Standalone Menus' → 'Hall C.' On the Hall C Operations Menu under “Experiment Specific” is SHMS Coll. Motion Control. Determine and record the collimator position. For physics runs, SHMS should be COLLIMATOR.
HMS Trigger Rate On the CODA GUI Screen, next to the appropriate prescaler value (ps1 thru ps6). Record the typical instantaneous rate
Target (material, length) Ask the target operator (TO). Record which target is in the beam.
Drift Ch. Mixer Flows: Ar/Eth (~0.3/~0.3 slpm) Type “go munin” and select “gas_shed” under Categories on the left side of the page. You can also find the needed charts here: [Flow and Temperature Charts] Record “Argon Flow” and “Ethane Flow.”
Drift Chamber Gas Delivery Press (500 ± 5 Torr) Same as above. Record “Ar:Eth Pressure.”
Drift Ch. Gas Isopropanol Temp. (2 ± 2 C) Same as above. Record the “Isopropanol Fridge” temperature.
SHMS Trigger Rate On the CODA GUI Screen, next to the appropriate prescaler value (ps1 thru ps6). Record the typical instantaneous rate
HMS Dead Time On the CODA GUI Screen, under the "computer live time" tab. Calculate and record the average value
SHMS Dead Time On the CODA GUI Screen, under the "computer live time" tab. Calculate and record the average value
Hall C General Checks