How-to

From HallCWiki
Revision as of 08:26, 14 December 2017 by Hcf (Talk | contribs) (Magnets)

Jump to: navigation, search

NOTE: If you can't solve the problem in 1 hour you MUST then contact the Run Coordinator! (See also Experts on call)

How-to's

Magnets

  • Magnet screen is brought up on any hcdesk machine using
 % go_magnets

SHMS Magnets

  • For the Fall 2017 run, the SHMS magnets can be set by current from the desired momentum in [GeV]. From any computer, run:
go_magnet_SHMS_current P

where P is the desired momentum in [GeV]. This will return the current that each SHMS magnet should be set to. The formulas and program are in sync with the repository. For precision physics data-taking, the magnets should be put "on loop". Turn off the beam prior to ramping the magnet and putting on loop. The procedure to put the magnet on loop is described here:

Ensure beam is off.
1. Select polarity for electrons or positrons in the magnet gui.
2. Ramp each magnet to the maximum current (see table below for SHMS).
3. Ramp each magnet down to its desired current (the output of go_magnet_SHMS_current). 
SHMS Magnet Max current [A]
HB 3600
Q1 2400
Q2 3660
Q3 2480
Dipole 3440

If you need to change polarity, first ramp the magnet to 0 [A], and then you can change the polarity. After this change, you will need to ramp the magnet to maximum again.

HMS Magnets

  • For HMS magnets, the current limits are shown in the following table:
HMS Magnet Max current [A]
Q1 1000
Q2 600
Q3 600
Dipole 3000

Do not exceed these limits! HMS magnets are put "on loop" by going 200 A above the desired set point and then reducing current to desired (see pdf below).

The currents for the HMS quads for a desired momentum are obtained by running the field03 program located at FIELD/field03.

The HMS dipole magnet has new NMR hardware and NMR settings. The desired NMR setting for a given momentum can be obtained from the following command:

go_magnet_HMS_current P

where P is the momentum in GeV. This will return the current and the recommended NMR setting.

This is the process that Mike guided us through to change the HMS dipole to the desired NMR field value:

  1) On the NMR screen, push the "stop" button and wait for it to be stopped.
  2) Enter the desired field in the input box and hit enter.
  3) Push "start" and wait. The system will calculate the initial current to set the dipole, and you should see the current setting change. This ought to be close to what you see from "go_magnet_HMS_current".
  4) The system is now going to wait a certain time for the power supply to reach the initial current, then it will start trying to get NMR lock. You just have to wait for it... it was almost 10 minutes for the change from 1.6162 GeV setting to 1.3773 setting. Once it gets NMR lock, it will begin making adjustments in the current. You can watch the dipole current stripchart to see when it begins regulating.
  5) Once the current and NMR values stabilize you should be good to go.

SHMS Detectors

HMS Detectors

Hall C target

Hall C target configuration for Spring 17 running

Legacy Documentation